FDA device news · Product code SHJ

FDA 510(k) clearance

Helo™ Return (K253717): Endovascular Engineering, Inc. 510(k) clearance

Menlo Park, CA

Follow SHJ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Decision
Substantially Equivalent
Decision date
Oct 11, 2026
Received
Nov 24, 2025
510(k) type
Traditional
Product code
SHJThrombectomy Whole Blood Return System
Review panel
Anesthesiology
Device class
Class II
Regulation
21 CFR 868.5830
Third-party review
No
Summary or statement
Summary
Review time
321 daysreceipt to decision, calendar days
Median in SHJ
Not availablethe latest 0 other decisions
SHJ decisions
1in the last 12 months
Applicant’s 510(k)s
2in openFDA, this one included

In plain English Written by AI

What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.

What it is
The Hēlo Return is a blood return system used with the Hēlo Thrombectomy System, a mechanical thrombectomy device that removes clots from blood vessels. It is a single-use, sterile canister mounted on a base with two suction tubes. Vacuum suction draws aspirated blood and clot into the canister, where a clot catcher with a coarse filter and two 40-micron polyester mesh filters remove particles.
What it’s cleared for
The Hēlo Return is cleared for use with the Hēlo Thrombectomy System to collect and filter aspirated whole blood from a thrombectomy procedure before the blood is returned to the patient. The summary does not describe a specific patient population beyond thrombectomy patients.
Compared with (the predicate)
  • K231782, Inari Medical, Inc. - FlowSaver Blood Return System
How it differs
The Hēlo Return uses two 40-micron polyester mesh filters, while both predicates use a single 40-micron and a single 200-micron filter assembly. The summary also notes that the Hēlo Return's canister allows clot visualization and that the Clot Catcher can be removed and emptied, as with the predicates.
Testing the summary lists
  • Sterilization (ISO 11135, ethylene oxide; sterility assurance level 10^-6)
  • Biocompatibility (ISO 10993-1)
  • Electrical safety and electromagnetic compatibility (IEC 60601-1, IEC 60601-2)
  • Shelf life and packaging (ASTM F1980 accelerated aging; transportation testing)

Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.

How the review time compares

Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.

K253717 took 321 calendar days from receipt to decision. openFDA has no other recent decisions in product code SHJ to compare it with.

Latest decisions in SHJ

NumberDevice and applicantDecision and review time
K253717 Helo™ ReturnEndovascular Engineering, Inc. · Traditional Oct 11, 2026321 days

Other 510(k)s from Endovascular Engineering, Inc.

The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.

NumberDevice and applicantDecision and review time
K252956 Helo Thrombectomy SystemEndovascular Engineering, Inc. · Traditional Dec 18, 202593 days

Watch SHJ in DeviceVigil

DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.

Add to Chrome — free Add to Edge

Get SHJ every Monday

One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in SHJ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.