FDA device news · Product code QEW
FDA 510(k) clearance
Helo Thrombectomy System (K252956): Endovascular Engineering, Inc. 510(k) clearance
Menlo Park, CA
- Decision
- Substantially Equivalent
- Decision date
- Dec 18, 2025
- Received
- Sep 16, 2025
- 510(k) type
- Traditional
- Product code
- QEWPeripheral Mechanical Thrombectomy With Aspiration
- Review panel
- Cardiovascular
- Device class
- Class II
- Regulation
- 21 CFR 870.5150
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 93 daysreceipt to decision, calendar days
- Median in QEW
- 59 daysthe latest 24 other decisions
- QEW decisions
- 11in the last 12 months
- Applicant’s 510(k)s
- 2in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The Hēlo Thrombectomy System is a minimally invasive aspiration system for removing thromboembolic material (clots and emboli) from the pulmonary arteries. It is a single-use catheter, introduced through a 16 Fr introducer sheath that expands to 8 mm at its distal end once inside the vasculature.
- What it’s cleared for
- The system is cleared for the non-surgical removal of emboli and thrombi from pulmonary arteries and venous vasculature, and for injecting, infusing, or aspirating contrast media and other fluids into or from blood vessels. The summary says it is intended for use in the peripheral veins and for the treatment of pulmonary embolism. The Audible Flow Indicator is intended only for use with this system.
- Compared with (the predicate)
- How it differs
- The summary lists differences in several features compared with the predicates. For example, the Hēlo System includes a separate 9F dilator packaged with the device, has a 95 cm catheter working length, uses a stainless steel agitator, and has a 12 month shelf life. The summary says the Hēlo System is substantially equivalent to both predicates.
- Testing the summary lists
- Sterilization validation (ISO 11135), with a stated sterility assurance level of 10 -6
- Electrical safety and EMC (IEC 60601-1, IEC 60601-2)
- Shelf-life and packaging (ASTM F1980 accelerated aging; ASTM F1886/1886M-16, ASTM F2096-11, ASTM F88-21 packaging tests)
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K252956 took 93 calendar days from receipt to decision. Of the latest 24 other decisions in product code QEW, 7 took longer, 17 were faster; their median was 59 days.
Latest decisions in QEW
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262904 | INDIGO® Aspiration SystemPenumbra, Inc. · Traditional | Oct 15, 202658 days |
| K253419 | Triever Duo CatheterInari Medical, Inc. · Traditional | Sep 18, 2026352 days |
| K261800 | Helo Max Thrombectomy System (HEL2090)Endovascular Engineering · Traditional | Sep 3, 202694 days |
| K262538 | Artix™ MT; Artix™ Thin-Walled SheathInari Medical, Inc. · Special | Aug 21, 202629 days |
| K261523 | Vertex™ Thrombectomy SystemJupiter Endovascular · Traditional | Aug 4, 202689 days |
| K260599 | INDIGO® Aspiration System – INDIGO LinkPenumbra, Inc. · Traditional | Apr 24, 202659 days |
| K251949 | INDIGO® Aspiration System - Lightning Flash Aspiration Tubing with LINKPenumbra, Inc. · Traditional | Feb 5, 2026225 days |
| K260028 | CLEANER™ Vac Thrombectomy System, CLEANER™ Vac Aspiration Catheter with Handpiece, CLEANER™ Vac Aspiration CanisterArgon Medical Devices, Inc. · Special | Feb 5, 202631 days |
| K253730 | RoVo Mechanical Thrombectomy SystemVerge Medical, Inc. · Special | Jan 21, 202658 days |
| K251207 | Sangria™ Thrombectomy SystemAvantec Vascular Corporation · Traditional | Jan 7, 2026264 days |
Other 510(k)s from Endovascular Engineering, Inc.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K253717 | Helo™ ReturnEndovascular Engineering, Inc. · Traditional | Oct 11, 2026321 days |
Watch QEW in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get QEW every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in QEW. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K252956Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:QEWRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:QEWRetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:QEW AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Endovascular Engineering, Inc."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.