FDA device news · Product code FYA
FDA 510(k) clearance
Henan Yadu Level 4 Surgical Gowns (Reinforced) (K252301): Yadu Medical (Henan) Co., Ltd. 510(k) clearance
He Nan, CN
- Decision
- Substantially Equivalent
- Decision date
- Feb 4, 2026
- Received
- Jul 24, 2025
- 510(k) type
- Traditional
- Product code
- FYAGown, Surgical
- Review panel
- General, Plastic Surgery
- Device class
- Class II
- Regulation
- 21 CFR 878.4040
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 195 daysreceipt to decision, calendar days
- Median in FYA
- 191 daysthe latest 24 other decisions
- FYA decisions
- 10in the last 12 months
- Applicant’s 510(k)s
- 3in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The Henan Yadu Level 4 Surgical Gown (Reinforced) is a sterile, single-use surgical gown worn by healthcare professionals during surgical procedures. The summary says it is made of multi-layer nonwoven fabric with a reinforced design, including knit cuffs and a poly-reinforcement film, and it comes in sizes from Small to XXXL.
- What it’s cleared for
- The gown is cleared to provide a liquid and microbial barrier for patients and medical staff. It is intended for operating room personnel during surgical operations and other invasive examinations, and it meets the Level 4 liquid barrier requirements of ANSI/AAMI PB70:2022.
- Compared with (the predicate)
- K182172, Surgical Apparel
- How it differs
- The summary says the device uses a multi-layer nonwoven fabric with an extra meltblown layer and double spunbond reinforcement, while the predicate uses three layers of nonwoven fabric. It also lists differences in materials, a white knitted cuff, and a size range that goes up to XXXL, compared with XXXX-Large for the predicate.
- Testing the summary lists
- Barrier performance (ANSI/AAMI PB70:2022, AATCC TM42, AATCC TM127, ASTM F1671)
- Physical properties (ASTM F2407, tensile, tear and seam strength, lint, flammability)
- Sterility and endotoxin (ISO 11737-2, ISO 11737-3)
- Ethylene oxide residue (ISO 10993-7)
- Biocompatibility (ISO 10993-5, ISO 10993-10, ISO 10993-23)
- Packaging performance and shelf life (ASTM F88, ASTM F1929, ISO 11737-1), planned under a predetermined change control plan
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K252301 took 195 calendar days from receipt to decision. Of the latest 24 other decisions in product code FYA, 12 took longer, 12 were faster; their median was 191 days.
Latest decisions in FYA
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260407 | Principle & Will ANSI/AAMI PB 70 (Sterile/Non-sterile) Level 3 Surgical Gowns (Reinforced) and Principle & Will ANSI/AMMI PB70 (Sterile/Non-sterile) Level 4 Surgical Gowns/Isolation GownsPrinciple & Will Co., Ltd. · Traditional | Aug 26, 2026198 days |
| K261028 | Surgical Gown Level2 (S, M, L, XL, XXL, XXXL)Yadu Medical (Henan) Co., Ltd. · Traditional | Aug 7, 2026130 days |
| K261949 | NYES Tuff Cuff L3 Disposable GownsNew York Embroidery Studio · Special | Jul 10, 202629 days |
| K252575 | Tuff Cuff™ Disposable Surgical Gowns AAMI L4 Sterile-Large (ML-L4-STL-DC-L)New York Embroidery Studio (NYES) · Traditional | Jun 4, 2026294 days |
| K260725 | BAÜMER Surgical Gowns (COP01, COP04, COP05, COP10)Hindernis DE Mexico, S.A.P.I. DE C.V. · Traditional | May 12, 202668 days |
| K253243 | Cardinal Health™ Poly Reinforced Surgical GownCardinal Health 200, LLC · Abbreviated | Apr 28, 2026211 days |
| K253003 | Non-sterile Level 3 Surgical Gown (4025); Sterile Level 3 Surgical Gown (M4025); Non-sterile Reinforced Level 4 Surgical Gown (G4025); Reinforced Sterile Level 4 Surgical Gown (GM4025)Hubei Xinxin Non-Woven Co., Ltd. · Traditional | Apr 20, 2026213 days |
| K251860 | Copioumed AAMI 3 Surgical GownCopioumed International, Inc. · Traditional | Feb 26, 2026254 days |
| K252301 | Henan Yadu Level 4 Surgical Gowns (Reinforced)Yadu Medical (Henan) Co., Ltd. · Traditional | Feb 4, 2026195 days |
| K250564 | Surgical Gown (Level 3: S, M, L, XL, XXL, XXXL)Yadu Medical (Henan) Co., Ltd. · Traditional | Dec 3, 2025280 days |
Other 510(k)s from Yadu Medical (Henan) Co., Ltd.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
Watch FYA in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get FYA every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in FYA. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K252301Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:FYARetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:FYARetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:FYA AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Yadu Medical Henan Co., Ltd."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil