FDA device news · Product code DQA
FDA 510(k) clearance
Unimed Reusable SpO2 Sensors (-48 Series and -29 Series) (U410-48, U410-29) (K251696): Unimed Medical Supplies, Inc. 510(k) clearance
Shenzhen, CN
- Decision
- Substantially Equivalent
- Decision date
- Dec 1, 2025
- Received
- Jun 2, 2025
- 510(k) type
- Traditional
- Product code
- DQAOximeter
- Review panel
- Cardiovascular
- Device class
- Class II
- Regulation
- 21 CFR 870.2700
- Third-party review
- No
- Summary or statement
- Summary
- Review time
- 182 daysreceipt to decision, calendar days
- Median in DQA
- 176 daysthe latest 24 other decisions
- DQA decisions
- 12in the last 12 months
- Applicant’s 510(k)s
- 20in openFDA, this one included
In plain English Written by AI
What the company’s 510(k) summary says about the device, its predicate and its testing, in plain words.
- What it is
- The Unimed Reusable SpO2 Sensors (models U410-48 and U410-29) are finger clip sensors for measuring blood oxygen. Each sensor has a connector, a cable, and a reusable part that holds a light-emitting diode (LED) and a photodetector, which shine light through the finger and measure how much is absorbed. The sensors are supplied non-sterile and are designed to work with a Mindray M12 monitor.
- What it’s cleared for
- The sensors are cleared for continuous, non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate in adult patients weighing more than 30 kg who are well or poorly perfused, applied to the finger. They are for no-motion use in professional healthcare settings and are prescription only.
- Compared with (the predicate)
- K142832, Unimed Disposable and Reusable SPO2 Sensors
- How it differs
- The summary says the device is limited to adult patients over 30 kg, while the predicate is also cleared for pediatric patients weighing 10 to 50 kg and neonatal patients weighing at least 3 kg.
- Testing the summary lists
- Electrical safety (IEC 60601-1, including amendments)
- Electromagnetic disturbances (IEC 60601-1-2)
- Pulse oximeter particular requirements (ISO 80601-2-61)
- Biocompatibility: cytotoxicity (ISO 10993-5), skin sensitization (ISO 10993-10), skin irritation (ISO 10993-23)
- Performance tests: saturation accuracy of ±3% (70-100%) and pulse rate accuracy of ±3 bpm (30-250 bpm), under no motion and low perfusion
Written by AI from the 510(k) summary the company filed with the FDA (the PDF), and checked against it: predicate numbers and figures must appear in the summary. A summary is the company’s own account, not the FDA’s review, and the AI can still get it wrong: read the PDF before you rely on it.
How the review time compares
Calendar days from the day FDA received the 510(k) to its decision. They include time the submission waited on the company for answers, so they run longer than FDA’s 90-day review goal.
K251696 took 182 calendar days from receipt to decision. Of the latest 24 other decisions in product code DQA, 11 took longer, 12 were faster, 1 took the same time; their median was 176 days.
Latest decisions in DQA
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K261870 | ANNE LimbSibel Health · Traditional | Nov 5, 2026154 days |
| K260490 | Ninenovo Blood Oxygen Ring (BORMB-01-6,BORMB-01-7,BORMB-01-8,BORMB-01-9,BORMB-01-10,BORMB-01-11,BORMB-01-12,BORMB-01-13,BORMB-01-14)Shenzhen Ninenovo Technology Limited · Traditional | Oct 2, 2026231 days |
| K252821 | Reusable Adult SpO2 Clip Sensor (3m) (M1196A); Reusable Adult SpO2 Clip Sensor (0.9m) (M1196T)Philips Medizin Systeme Böblingen GmbH · Traditional | May 29, 2026267 days |
| K252655 | Pulse Oximeter (PO-A2AO, PO-A2AT, PO-A3AO, PO-H1AO, PO-B1AO, PO-C5AO, PO-C5AT, PO-C6AO, PO-C6AT)Shenzhen Imdk Medical Technology Co., Ltd. · Traditional | May 11, 2026262 days |
| K253887 | Nasal Alar SpO2 Sensor (989803205381, 989803205391); Nasal Alar Sp02 Sensor (989803205401)Philips Medizin Systeme · Traditional | May 1, 2026148 days |
| K260931 | Unimed Reusable Finger Clip SpO2 Sensors (U403-49R and U103-49R)Unimed Medical Supplies, Inc. · Special | Apr 14, 202625 days |
| K252448 | AViTA Pulse Oximeter (SP61)Avita Corporation · Traditional | Feb 27, 2026207 days |
| K252805 | YUWELL® Finger Pulse Oximeter (YX105, YX106, YX110, YX310)Jiangsu Yuyue Medical Equipment & Supply Co., Ltd. · Traditional | Feb 17, 2026167 days |
| K253109 | Reusable SPO2 Sensor (BSA307-47; BSA109-50; BSA109-31O)Shenzhen Best Electronics Co., Ltd. · Traditional | Jan 7, 2026105 days |
| K251751 | Spo2 Sensor CSS032DShenzhen Ykd Technology Co., Ltd. · Traditional | Dec 19, 2025193 days |
Other 510(k)s from Unimed Medical Supplies, Inc.
The applicant’s other clearances in openFDA, newest first. Company names are matched as the FDA recorded them, so a renamed or differently spelled company may not appear.
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K253162 | Temperature probes (T2252-PG, TCMA-AG, TCMA-PG, TMQ-DAG, THP-DAG)Unimed Medical Supplies, Inc. · Traditional | Jun 5, 2026252 days |
| K260931 | Unimed Reusable Finger Clip SpO2 Sensors (U403-49R and U103-49R)Unimed Medical Supplies, Inc. · Special | Apr 14, 202625 days |
| K253964 | Disposable Neonatal NIBP Cuff (U1681S-C51N)Unimed Medical Supplies, Inc. · Special | Jan 9, 202629 days |
| K251691 | Unimed Reusable SpO2 Sensors (-08 Series) (U403S-08, U203S-08, U103S-08, U103-08)Unimed Medical Supplies, Inc. · Traditional | Dec 1, 2025182 days |
| K251693 | Unimed Reusable SpO2 Sensors (-75 Series) (U410-75, U110-75, U410S-75, U110S-75, U210S-75)Unimed Medical Supplies, Inc. · Traditional | Nov 18, 2025169 days |
| K251695 | Unimed Reusable SpO2 Sensors (-06 Series) (U403S-06, U203S-06, U103S-06)Unimed Medical Supplies, Inc. · Traditional | Nov 18, 2025169 days |
| K250931 | Unimed Reusable SpO2 Sensors (-117 and -121 Series) (U403-117, U103-117, U410-121, U110-121)Unimed Medical Supplies, Inc. · Traditional | Aug 13, 2025138 days |
| K251045 | Disposable Neonatal NIBP Cuff (U1682S-C51N, U1683S-C51N, U1684S-C51N, U1685S-C51N)Unimed Medical Supplies, Inc. · Traditional | Jul 15, 2025103 days |
| K243086 | Unimed Reusable Finger Clip SpO2 Sensors (UXXX-91 Series) (U403-91, U410-91)Unimed Medical Supplies, Inc. · Traditional | Feb 20, 2025143 days |
| K242580 | Unimed Reusable Finger Clip SpO2 Sensors (U103-125, U403-254, U103-254, U403-125)Unimed Medical Supplies, Inc. · Traditional | Feb 6, 2025160 days |
| K222213 | Pulse Oximeter (UC-200, UC-201, UC-202, UC-203, UC-204, UC-100, UC-101, UC-102, UC-103, UC-104)Unimed Medical Supplies, Inc. · Traditional | Dec 19, 2022147 days |
| K210019 | Unimed CO2 Mainstream Module Model Capno-M, Unimed CO2 Mainstream Module Model Capno-M+Unimed Medical Supplies, Inc. · Traditional | Mar 4, 202159 days |
Watch DQA in DeviceVigil
DeviceVigil is a free Chrome and Edge extension. Watch the product codes and companies you follow, and it tells you about new clearances, recalls and MAUDE reports for them, right beside the FDA’s own pages.
Get DQA every Monday
One short email after openFDA’s weekly update with that week’s new 510(k)s, De Novos, PMAs and recalls in DQA. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/510k · search k_number:K251696Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:DQARetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/classification · search product_code:DQARetrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search product_code:DQA AND decision_date:[20251129 TO 20261129]Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
- device/510k · search applicant:"Unimed Medical Supplies, Inc."Retrieved Nov 22, 2026, 9:07 PM UTC. FDA data updated Nov 18, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.