FDA product code
SEG: Leadless Cardiac Resynchronization Therapy (Crt) (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class III
- Regulation
- None listed
- Review panel
- Cardiovascular
- Premarket route
- PMA
- Implant
- Yes
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- PMA decisions
- 61 original PMA, the rest supplements
- Median review time
- 29 daysthe latest 6 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 78last 12 months, 6 deaths
PMA decisions
Original PMAs and supplements for product code SEG per year, by decision date, 6 since 2025. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 5 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| P240028 S006 | WiSE® (Wireless Stimulation of the Endocardium Technology) CRT SystemEBR Systems, Inc. | Mar 27, 202630 days |
| P240028 S004 | WiSE® (Wireless Stimulation of the Endocardium Technology) CRT SystemEBR Systems, Inc. | Dec 15, 202528 days |
| P240028 S003 | WiSE (Wireless Stimulation of the Endocardium Technology) CRT SystemEBR Systems, Inc. | Oct 3, 202523 days |
| P240028 S002 | WiSE® (Wireless Stimulation of the Endocardium Technology) CRT SystemEBR Systems, Inc. | Sep 3, 202528 days |
| P240028 S001 | WiSE® (Wireless Stimulation of the Endocardium Technology) CRT SystemEBR Systems, Inc. | Jun 18, 202530 days |
| P240028 | WiSE® (Wireless Stimulation of the Endocardium Technology) CRT SystemEBR Systems, Inc. | Apr 11, 2025225 days |
Recalls
Recall records for product code SEG per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code SEG. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 43 |
| 2026 (so far) | 70 |
Last 12 months by event type
- Deaths6
- Injuries22
- Malfunctions50
- Other or not stated0
Most reported device problems, last 12 months
- Human-Device Interface Problem18
- Communication or Transmission Problem14
- Adverse Event Without Identified Device or Use Problem12
- No Apparent Adverse Event6
- Insufficient Device Problem Information5
- Intermittent Capture3
In plain English: what the newest reports describe Written by AI
Patients developed ventricular fibrillation, a dangerous heart rhythm, during routine echocardiograms shortly after implant. One report describes a patient shocked back by an implanted defibrillator and another describes a death. The manufacturer paused implant procedures and noted a possible link to the ultrasound probe near the device.
4 reports: 3013596742-2026-00053, 3013596742-2026-00054, 25531810, MW5189421
After the electrode was released or repositioned, it came loose and moved (dislodged) in several reports. In one, it was retrieved with a snare; in another, retrieval failed. The manufacturer linked one dislodgement to the delivery sheath's pressure during release.
3 reports: 3013596742-2026-00063, 3013596742-2026-00056, 3013596742-2026-00052
Some patients had signal and pacing problems. In one report the transmitter was found flipped, and the device could not track the electrode. In another, the device could not be read by the programmer, pacing in both ventricles was lost, and the patient's condition worsened. The manufacturer's findings on cause were not stated.
2 reports: 3013596742-2026-00050, 3013596742-2026-00062
Parts of the implanted battery and transmitter caused problems under the skin. In one report, the skin over the battery eroded and the device was exposed, so the battery and transmitter were removed. In another, a patient with a flank hematoma needed a blood transfusion. The report says no device malfunction was found, and it does not state a link to the device.
2 reports: 3013596742-2026-00068, 3013596742-2026-00060
During one implant attempt, fluid around the heart (pericardial effusion) was seen before the electrode was placed and kept growing. About 600 cc of blood was drained, and the procedure was stopped. The report describes very high clotting time and low blood pressure during the case.
Report: 3013596742-2026-00064
Written by AI from the narratives of the 14 newest reports that have one, received Jun 15, 2026 to Sep 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 5in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 2registered establishments that list this product code
Watch SEG in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get SEG every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in SEG. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:SEGRetrieved Nov 29, 2026, 10:43 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:SEGRetrieved Nov 29, 2026, 10:43 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:SEG · count decision_dateRetrieved Nov 29, 2026, 10:43 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:SEG · count pma_numberRetrieved Nov 29, 2026, 10:43 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:SEGRetrieved Nov 29, 2026, 10:43 PM UTC.
- device/recall · search product_code:SEG · count event_date_initiatedRetrieved Nov 29, 2026, 10:43 PM UTC.
- device/recall · search product_code:SEG · count res_event_numberRetrieved Nov 29, 2026, 10:43 PM UTC.
- device/event · search device.device_report_product_code:SEG AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 10:43 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:SEG AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 10:43 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:SEG AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 10:43 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:SEGRetrieved Nov 29, 2026, 10:43 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:SEGRetrieved Nov 29, 2026, 10:43 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil