FDA product code
SCA: Multi-Analyte Respiratory Virus Antigen Detection Test (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 866.3987
- Review panel
- Microbiology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 134 in the last 12 months
- Median review time
- 93 daysthe latest 13 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 70last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code SCA per year, by decision date, 13 since 2024. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 10 |
| 2026 (so far) | 2 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K262432 | CorDx Tyfast Flu A/B & COVID-19 At Home Multiplex Rapid Test; CorDx Tyfast Flu A/B & COVID-19 Multiplex Rapid TestCorDx, Inc. | Oct 14, 202690 days |
| K261212 | Flowflex Plus RSV + Flu A/B + COVID Rapid TestACON Laboratories, Inc. | Jul 10, 202687 days |
| K251085 | iHealth Flu A&B/COVID-19/RSV Rapid Test; iHealth Flu A&B/COVID-19/RSV Rapid Test ProIhealth Labs, Inc. | Dec 12, 2025247 days |
| K251092 | iHealth Flu A&B/COVID-19 Rapid Test; iHealth Flu A&B/COVID-19 Rapid Test ProIhealth Labs, Inc. | Dec 12, 2025246 days |
| K251538 | Status™ COVID-19/Flu A&BPrinceton BioMeditech Corp. | Nov 14, 2025179 days |
| K251749 | Flowflex Plus RSV + Flu A/B + COVID Home TestACON Laboratories, Inc. | Oct 22, 2025138 days |
| K251604 | CareSuperb COVID-19/Flu A&B Antigen Combo Home TestAccess Bio, Inc. | Aug 22, 202587 days |
| K251563 | WELLlife Flu A&B Home Test; WELLlife Influenza A&B TestWondfo USA Co, Ltd. | Aug 20, 202590 days |
| K243561 | Nano-Check Influenza+COVID-19 Dual TestNano-Ditech Corporation | Jun 17, 2025211 days |
| K250377 | Flowflex Plus COVID-19 + Flu A/B Home TestACON Laboratories, Inc. | May 10, 202589 days |
Recalls
Recall records for product code SCA per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code SCA. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 4 |
| 2026 (so far) | 69 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions69
- Other or not stated1
Most reported device problems, last 12 months
- False Negative Result37
- Unable to Obtain Readings29
- False Positive Result3
- Defective Device1
- Imprecise or Erratic Results1
In plain English: what the newest reports describe Written by AI
Many reports say the test came back negative for flu, RSV or COVID, but the person was later found positive by a doctor, urgent care or another test. In one family, three home tests were negative while a doctor's flu test had been positive the day before.
9 reports: 2531491-2026-00145, 2531491-2026-00111, 2531491-2026-00106, 2531491-2026-00099, 2531491-2026-00102, 2531491-2026-00095, 2531491-2026-00090, 2531491-2026-00086, 2531491-2026-00094
In several reports, the control line, which shows the test ran properly, did not appear on one or both halves of the strip, so no clear result could be read. Reporters said they followed the directions.
5 reports: 2531491-2026-00148, 2531491-2026-00128, 2531491-2026-00130, 2531491-2026-00100, 2531491-2026-00101
Some tests showed no result at all, with no readable lines or only a faint pinkish background. One report described a test as defective, and one reporter threw the tests away before sending photos.
3 reports: 2531491-2026-00098, 2531491-2026-00130, 2531491-2026-00115
Some people got conflicting results. One home test read positive but a doctor's test was negative, and another report describes one test showing flu A and a different test the next day showing flu B.
2 reports: 3008573045-2026-00006, 2531491-2026-00126
One buyer said a pharmacist confirmed the first test was faulty, but the store would not refund it and the distributor did not answer emails. The buyer said a CDC referral sent them to the FDA.
Report: MW6001405
Written by AI from the narratives of the 19 newest reports that have one, received Mar 31, 2026 to Aug 24, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 84in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 44registered establishments that list this product code
Watch SCA in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get SCA every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in SCA. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:SCARetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:SCARetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:SCA · count decision_dateRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:SCARetrieved Nov 29, 2026, 4:49 PM UTC.
- device/recall · search product_code:SCA · count event_date_initiatedRetrieved Nov 29, 2026, 4:49 PM UTC.
- device/recall · search product_code:SCA · count res_event_numberRetrieved Nov 29, 2026, 4:49 PM UTC.
- device/event · search device.device_report_product_code:SCA AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:SCA AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:SCA AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:SCARetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:SCARetrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:866.3987Retrieved Nov 29, 2026, 4:49 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil