Product codes

FDA product code

SAZ: Stimulator, Nerve Terminus, Totally Implanted For Chronic Intractable Post-Amputation Pain Of The Lower Limb (FDA product code)

Follow SAZ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class III
Regulation
None listed
Review panel
Neurology
Premarket route
PMA
Implant
Yes
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
PMA decisions
91 original PMA, the rest supplements
Median review time
81 daysthe latest 9 decisions
Recall records
00 in the last 12 months
Adverse event reports
6last 12 months, 0 deaths

PMA decisions

Original PMAs and supplements for product code SAZ per year, by decision date, 9 since 2024. Review time is the days from receipt to decision.

All of them in the FDA’s PMA database

PMA decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20242
20254
2026 (so far)3

Latest decisions

NumberDevice and applicantDecision and review time
P230020 S011 Altius® Direct Electrical Nerve Stimulation SystemNeuros Medical, Inc. Oct 16, 202629 days
P230020 S006 Altius® Direct Electrical Nerve Stimulation SystemNeuros Medical, Inc. May 5, 2026161 days
P230020 S007 Altius® Direct Electrical Nerve Stimulation SystemNeuros Medical, Inc. Feb 28, 202629 days
P230020 S005 Altius® Direct Electrical Nerve Stimulation SystemNeuros Medical, Inc. Dec 3, 2025149 days
P230020 S003 Altius Direct Electrical Nerve Stimulation SystemNeuros Medical, Inc. Jun 25, 202585 days
P230020 S004 Altius® Direct Electrical Nerve Stimulation SystemNeuros Medical, Inc. Jun 5, 202527 days
P230020 S002 Altius Direct Electrical Nerve Stimulation SystemNeuros Medical, Inc. May 6, 202581 days
P230020 S001 Altius® Direct Electrical Nerve Stimulation SystemNeuros Medical, Inc. Dec 19, 202424 days
P230020 Altius® Direct Electrical Nerve Stimulation SystemNeuros Medical, Inc. Aug 26, 2024423 days

Recalls

Recall records for product code SAZ per year, by the date the recall began. Each recalled product is one record.

All of them in the FDA’s recall database

No recalls on record for this product code.

Adverse event reports (MAUDE)

Medical device reports received per year that name product code SAZ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
20180
20190
20200
20210
20220
20230
20240
20253
2026 (so far)4

Last 12 months by event type

  • Deaths0
  • Injuries6
  • Malfunctions0
  • Other or not stated0

Most reported device problems, last 12 months

  1. Adverse Event Without Identified Device or Use Problem6

In plain English: what the newest reports describe Written by AI

  • Reports describe infection around the implanted pulse generator (the battery unit under the skin) and the stimulating cuff, leading to fever, swelling, hospital care and removal of the device. One report simply says a patient had the device removed due to infection.

    4 reports: 3026283936-2026-10014, 3026283936-2026-20014, 3026283936-2026-30014, 3026283936-2025-20021

  • In one report, the incision over the generator reopened the evening after the device was switched on. The patient had no infection, went to the emergency room and was admitted, and the generator was replaced in a second procedure.

    Report: 3026283936-2026-10022

  • In one report, a patient with a below-knee amputation felt strong shocks during home therapy sessions, and said they once fell from a chair. The therapy strength was lowered to reduce the discomfort. The patient had been in a clinical trial.

    Report: 3026283936-2025-00006

Written by AI from the narratives of the 6 newest reports that have one, received Mar 31, 2025 to Jul 10, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
9in the FDA’s Global Unique Device Identification Database
Establishment listings
3registered establishments that list this product code

Watch SAZ in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get SAZ every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in SAZ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.