Product codes

FDA product code

QWR: Simple Point-Of-Care Device To Detect Sar-Cov-2 Nucleic Acid Targets From Clinical Specimens In Near-Patient Settings (FDA product code)

Follow QWR: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 866.3982
Review panel
Microbiology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
510(k) and De Novo decisions
30 in the last 12 months
Median review time
353 daysthe latest 3 decisions
Recall records
20 in the last 12 months
Adverse event reports
42last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code QWR per year, by decision date, 3 since 2023. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20232
20240
20251
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K243346 cobas liat SARS-CoV-2 v2 nucleic acid testRoche Molecular Systems, Inc. Apr 11, 2025165 days
K223783 cobas SARS-CoV-2 Nucleic acid test for use on the cobas Liat SystemRoche Molecular Systems, Inc. Dec 4, 2023353 days
K221925 ID NOW COVID-19 2.0Abbott Diagnostics Scarborough, Inc. Aug 10, 2023405 days

Recalls

Recall records for product code QWR per year, by the date the recall began. Each recalled product is one record; these 2 records belong to 2 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20220
20230
20240
20252
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1571-2025Class II Abbott Diagnostics Scarborough, Inc.The impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use. Mar 4, 2025
Z-1574-2025Class II Abbott Diagnostics Scarborough, Inc.the impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use. Mar 4, 2025

Adverse event reports (MAUDE)

Medical device reports received per year that name product code QWR. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
20180
20190
20200
20210
20220
202312
202490
202525
2026 (so far)39

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions42
  • Other or not stated0

Most reported device problems, last 12 months

  1. False Positive Result36
  2. Non Reproducible Results5
  3. False Negative Result1

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
5in the FDA’s Global Unique Device Identification Database
Establishment listings
6registered establishments that list this product code

Product codes classified under the same regulation.

Watch QWR in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get QWR every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in QWR. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.