FDA product code
QQK: Normothermic Machine Perfusion System For The Preservation Of Donor Livers Prior To Transplantation (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class III
- Regulation
- None listed
- Review panel
- Gastroenterology and Urology
- Premarket route
- PMA
- Implant
- No
- Life-sustaining or supporting
- Yes
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- PMA decisions
- 542 original PMAs, the rest supplements
- Median review time
- 76 daysthe latest 25 decisions
- Recall records
- 10 in the last 12 months
- Adverse event reports
- 282last 12 months, 1 death
PMA decisions
Original PMAs and supplements for product code QQK per year, by decision date, 54 since 2021. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 2 |
| 2022 | 4 |
| 2023 | 10 |
| 2024 | 10 |
| 2025 | 17 |
| 2026 (so far) | 11 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| P200035 S038 | OrganOx metra® SystemOrganox Limited | Oct 9, 2026168 days |
| P200035 S041 | OrganOx metra SystemOrganox Limited | Aug 7, 202622 days |
| P200035 S040 | OrganOx metra SystemOrganox Limited | Jul 8, 202630 days |
| P200035 S037 | OrganOx metra® SystemOrganox Limited | Jul 7, 202689 days |
| P200035 S039 | OrganOx metra® SystemOrganox Limited | Jun 5, 202628 days |
| P200035 S033 | OrganOx metra® SystemOrganox Limited | May 21, 2026142 days |
| P200031 S013 | Organ Care System (OCS™) LiverTransMedics, Inc. | May 13, 2026155 days |
| P200035 S036 | OrganOx metra® SystemOrganox Limited | Apr 10, 202628 days |
| P200035 S035 | OrganOx metra® SystemOrganox Limited | Mar 30, 202689 days |
| P200035 S032 | OrganOx metra® SystemOrganox Limited | Mar 16, 202687 days |
Recalls
Recall records for product code QQK per year, by the date the recall began. Each recalled product is one record; these 1 records belong to 1 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0501-2025Class II | OrganOx LtdBattery movement led to damage of one of the battery cells causing it to short to the battery retention bracket | Oct 15, 2024 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code QQK. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 139 |
| 2025 | 168 |
| 2026 (so far) | 259 |
Last 12 months by event type
- Deaths1
- Injuries66
- Malfunctions213
- Other or not stated2
Most reported device problems, last 12 months
- Insufficient Flow or Under Infusion39
- Physical Resistance/Sticking35
- No Flow32
- Restricted Flow rate29
- Pressure Problem17
- Loose or Intermittent Connection14
In plain English: what the newest reports describe Written by AI
Four reports describe a pump problem that left the blood reservoir low or empty. One pump ran continuously, another stopped, and one recirculation pump was fixed by a stop and restart. Three livers were discarded or declined and one was transplanted.
4 reports: 3011560054-2026-00231, 3011560054-2026-00221, 3011560054-2026-00219, 3011560054-2026-00207
Four reports describe low flow or pressure through the liver, which looked congested or dark. In one, all flows read zero and the liver was discarded. Another found significant air entering the caval cannula, and a sensor was removed to raise pump speed.
4 reports: 3011560054-2026-00223, 3011560054-2026-00233, 3011560054-2026-00199, 3011560054-2026-00196
Three reports describe the syringe driver, which delivers medications, not advancing when its black knob was turned. One user found the syringes had not moved past a marked line, and the teams gave medications by hand instead. One liver was still transplanted.
3 reports: 3011560054-2026-00228, 3011560054-2026-00209, 3011560054-2026-00208
Three reports describe damaged parts: a leak near a pressure sensor, a bowl cracked possibly by bumps in transport, and a plastic connector that broke off the module. The leaking set was used anyway, and the cracked bowl contributed to the liver being declined.
3 reports: 3011560054-2026-00236, 3011560054-2026-00231, 3003152463-2026-00013
One report describes a recipient whose graft failed and who died 13 days after transplant, with a fungal infection suggestive of Candida found on examination. The manufacturer found no evidence the device caused it, but the possibility that the infection began during machine perfusion has not been ruled out.
Report: 50730997
Written by AI from the narratives of the 17 newest reports that have one, received Sep 4, 2026 to Oct 28, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 16in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 10registered establishments that list this product code
Watch QQK in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get QQK every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in QQK. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:QQKRetrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:QQKRetrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:QQK · count decision_dateRetrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:QQK · count pma_numberRetrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:QQKRetrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:QQK · count event_date_initiatedRetrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:QQK · count res_event_numberRetrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0501-2025")Retrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:QQK AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:QQK AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:QQK AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:QQKRetrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:QQKRetrieved Nov 29, 2026, 11:17 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil