Product codes

FDA product code

QQK: Normothermic Machine Perfusion System For The Preservation Of Donor Livers Prior To Transplantation (FDA product code)

Follow QQK: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class III
Regulation
None listed
Review panel
Gastroenterology and Urology
Premarket route
PMA
Implant
No
Life-sustaining or supporting
Yes
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
PMA decisions
542 original PMAs, the rest supplements
Median review time
76 daysthe latest 25 decisions
Recall records
10 in the last 12 months
Adverse event reports
282last 12 months, 1 death

PMA decisions

Original PMAs and supplements for product code QQK per year, by decision date, 54 since 2021. Review time is the days from receipt to decision.

All of them in the FDA’s PMA database

PMA decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20212
20224
202310
202410
202517
2026 (so far)11

Latest decisions

NumberDevice and applicantDecision and review time
P200035 S038 OrganOx metra® SystemOrganox Limited Oct 9, 2026168 days
P200035 S041 OrganOx metra SystemOrganox Limited Aug 7, 202622 days
P200035 S040 OrganOx metra SystemOrganox Limited Jul 8, 202630 days
P200035 S037 OrganOx metra® SystemOrganox Limited Jul 7, 202689 days
P200035 S039 OrganOx metra® SystemOrganox Limited Jun 5, 202628 days
P200035 S033 OrganOx metra® SystemOrganox Limited May 21, 2026142 days
P200031 S013 Organ Care System (OCS™) LiverTransMedics, Inc. May 13, 2026155 days
P200035 S036 OrganOx metra® SystemOrganox Limited Apr 10, 202628 days
P200035 S035 OrganOx metra® SystemOrganox Limited Mar 30, 202689 days
P200035 S032 OrganOx metra® SystemOrganox Limited Mar 16, 202687 days

Recalls

Recall records for product code QQK per year, by the date the recall began. Each recalled product is one record; these 1 records belong to 1 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20220
20230
20241
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0501-2025Class II OrganOx LtdBattery movement led to damage of one of the battery cells causing it to short to the battery retention bracket Oct 15, 2024

Adverse event reports (MAUDE)

Medical device reports received per year that name product code QQK. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
20180
20190
20200
20210
20220
20230
2024139
2025168
2026 (so far)259

Last 12 months by event type

  • Deaths1
  • Injuries66
  • Malfunctions213
  • Other or not stated2

Most reported device problems, last 12 months

  1. Insufficient Flow or Under Infusion39
  2. Physical Resistance/Sticking35
  3. No Flow32
  4. Restricted Flow rate29
  5. Pressure Problem17
  6. Loose or Intermittent Connection14

In plain English: what the newest reports describe Written by AI

  • Four reports describe a pump problem that left the blood reservoir low or empty. One pump ran continuously, another stopped, and one recirculation pump was fixed by a stop and restart. Three livers were discarded or declined and one was transplanted.

    4 reports: 3011560054-2026-00231, 3011560054-2026-00221, 3011560054-2026-00219, 3011560054-2026-00207

  • Four reports describe low flow or pressure through the liver, which looked congested or dark. In one, all flows read zero and the liver was discarded. Another found significant air entering the caval cannula, and a sensor was removed to raise pump speed.

    4 reports: 3011560054-2026-00223, 3011560054-2026-00233, 3011560054-2026-00199, 3011560054-2026-00196

  • Three reports describe the syringe driver, which delivers medications, not advancing when its black knob was turned. One user found the syringes had not moved past a marked line, and the teams gave medications by hand instead. One liver was still transplanted.

    3 reports: 3011560054-2026-00228, 3011560054-2026-00209, 3011560054-2026-00208

  • Three reports describe damaged parts: a leak near a pressure sensor, a bowl cracked possibly by bumps in transport, and a plastic connector that broke off the module. The leaking set was used anyway, and the cracked bowl contributed to the liver being declined.

    3 reports: 3011560054-2026-00236, 3011560054-2026-00231, 3003152463-2026-00013

  • One report describes a recipient whose graft failed and who died 13 days after transplant, with a fungal infection suggestive of Candida found on examination. The manufacturer found no evidence the device caused it, but the possibility that the infection began during machine perfusion has not been ruled out.

    Report: 50730997

Written by AI from the narratives of the 17 newest reports that have one, received Sep 4, 2026 to Oct 28, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
16in the FDA’s Global Unique Device Identification Database
Establishment listings
10registered establishments that list this product code

Watch QQK in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get QQK every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in QQK. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.