Product codes

FDA product code

QLQ: Manual Instruments Designed For Use With Total Disc Replacement Devices (FDA product code)

Follow QLQ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 888.4515
Review panel
Orthopedic
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
510(k) and De Novo decisions
103 in the last 12 months
Median review time
59 daysthe latest 10 decisions
Recall records
00 in the last 12 months
Adverse event reports
1last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code QLQ per year, by decision date, 10 since 2021. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20211
20222
20231
20242
20251
2026 (so far)3

Latest decisions

NumberDevice and applicantDecision and review time
K263159 prodisc® C; prodisc® C SK; prodisc® C Nova; prodisc® C Vivo InstrumentsCentinel Spine, LLC Oct 7, 202628 days
K262639 BAGUERA® C Cervical Disc Surgical InstrumentsSpineart SA Aug 14, 202616 days
K253392 Synergy Disc InstrumentsSynergy Spine Solutions, Inc. Mar 27, 2026178 days
K250554 prodisc® L InstrumentsCentinel Spine, LLC Apr 25, 202559 days
K242869 prodisc® L InstrumentsCentinel Spine, LLC Dec 17, 202488 days
K241117 M6-C™ Single Use, Disposable InstrumentationSpinal Kinetics / Orthofix / Seaspine Jun 21, 202459 days
K231769 activL® Next Generation InstrumentationAesculap Implants Systems, LLC Jul 13, 202327 days
K221848 prodisc C SK, prodisc C Nova, and prodisc C Vivo InstrumentsCentinel Spine, LLC Aug 19, 202256 days
K220861 M6-C Artificial Cervical Disc Instruments ASSpinal Kinetics, LLC Jun 2, 202270 days
K211757 Simplify DiscNuvasive, Inc. Aug 20, 202174 days

Recalls

Recall records for product code QLQ per year, by the date the recall began. Each recalled product is one record.

All of them in the FDA’s recall database

No recalls on record for this product code.

Adverse event reports (MAUDE)

Medical device reports received per year that name product code QLQ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
20180
20190
20200
20211
20222
20230
20240
20250
2026 (so far)1

Last 12 months by event type

  • Deaths0
  • Injuries1
  • Malfunctions0
  • Other or not stated0

Most reported device problems, last 12 months

  1. Fracture1

In plain English: what the newest reports describe Written by AI

  • In one report, parts of the surgical tools broke off inside the spine during surgery: broach tips (tools used to shape bone) and keel cutter blades (tools that cut a groove in bone). The broken pieces were removed, the tear in the membrane around the spinal cord was repaired, and the patient was admitted to intensive care because of a leak of spinal fluid. The report says the patient had no spinal cord injury.

    Report: 3004142400-2026-00052

  • In one report, a surgeon's glove got caught in the blue-ringed handle of an insertion instrument while the surgeon tapped the implant into place with a mallet. The surgeon used X-ray imaging to confirm the implant was correctly positioned, and the patient needed an additional, unexpected diagnostic procedure.

    Report: 2916714-2022-00012

  • In one report, a disc shaver (a tool that removes disc tissue) came apart between its proximal and distal ends, meaning the two sections separated during use.

    Report: 3009051471-2017-00008

  • In one report, a trial implant sizing tool and a keel chisel (a tool that cuts a groove in the bone) were pushed too far into the disc space during preparation. The report says the patient had a serious adverse effect after the trial was removed, but does not describe what that effect was.

    Report: 3004142400-2021-00008

Written by AI from the narratives of the 4 newest reports that have one, received Jan 14, 2021 to Mar 2, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
169in the FDA’s Global Unique Device Identification Database
Establishment listings
18registered establishments that list this product code

Watch QLQ in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get QLQ every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in QLQ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.