Product codes

FDA product code

PUM: Temporary Catheter, Embolic Protection, Transcatheter Intracardiac Procedures (FDA product code)

Follow PUM: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 870.1251
Review panel
Cardiovascular
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
510(k) and De Novo decisions
31 in the last 12 months
Median review time
163 daysthe latest 3 decisions
Recall records
00 in the last 12 months
Adverse event reports
34last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code PUM per year, by decision date, 3 since 2017. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20171
20180
20190
20201
20210
20220
20230
20240
20250
2026 (so far)1

Latest decisions

NumberDevice and applicantDecision and review time
K261715 Emboliner Embolic Protection System (EPC-03)Emboline, Inc. Sep 8, 2026109 days
K192460 Sentinel Cerebral Protection SystemBoston Scientific Corporation Feb 19, 2020163 days
DEN160043 Sentinel Cerebral Protection SystemClaret Medical, Inc. Jun 1, 2017254 days

Recalls

Recall records for product code PUM per year, by the date the recall began. Each recalled product is one record.

All of them in the FDA’s recall database

No recalls on record for this product code.

Adverse event reports (MAUDE)

Medical device reports received per year that name product code PUM. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
201819
201948
202073
202143
202240
202324
202445
202550
2026 (so far)29

Last 12 months by event type

  • Deaths0
  • Injuries6
  • Malfunctions28
  • Other or not stated0

Most reported device problems, last 12 months

  1. Nonstandard Device18
  2. Difficult to Remove16
  3. Material Integrity Problem6
  4. Adverse Event Without Identified Device or Use Problem4
  5. Failure to Advance4
  6. Mechanical Jam4

In plain English: what the newest reports describe Written by AI

  • The embolic protection filter failed to fully retract or recapture back into its sheath, so the filter could not be closed for removal. In several of these reports the physician removed the device in pieces or exchanged it for a new one to finish the procedure.

    5 reports: 2124215-2026-53733, 2124215-2026-41212, 2124215-2026-29689, 2124215-2026-23299, 2124215-2026-19070

  • The filter material or its frame came apart: in some reports the filter fabric tore or detached from its hoop, the frame that holds the filter open. In one case the detached fabric was folded inside the catheter and caused the resistance the team felt.

    4 reports: 2124215-2026-41212, 2124215-2026-31786, 2124215-2026-27005, 2124215-2026-27008

  • The device got stuck or would not withdraw through the introducer sheath in the arm or wrist artery. In these cases the sheath accordioned or wrinkled, and the team removed the device and sheath together as one unit.

    3 reports: 2124215-2026-31821, 2124215-2026-20439, 2124215-2026-19394

  • The device would not advance or could not be placed in the patient's blood vessels, often because the vessel was tortuous, meaning twisted. In a poster presentation on 102 cases, difficulty placing the device was linked to twisted or calcified arteries, and two patients had a stroke within 30 days.

    3 reports: 2124215-2026-44725, 2124215-2026-36902, 2124215-2026-31277

  • A patient had a stroke, called a cerebral vascular accident, one day after a valve replacement procedure in which the device was used. The patient was hospitalized and symptoms were improving. The physician judged the event related to the device procedure.

    Report: 2124215-2026-48756

Written by AI from the narratives of the 20 newest reports that have one, received Apr 8, 2026 to Oct 9, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
1in the FDA’s Global Unique Device Identification Database
Establishment listings
8registered establishments that list this product code

Watch PUM in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get PUM every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PUM. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.