FDA product code
PSY: Whole Slide Imaging System (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 864.3700
- Review panel
- Pathology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 234 in the last 12 months
- Median review time
- 237 daysthe latest 23 decisions
- Recall records
- 21 in the last 12 months
- Adverse event reports
- 5last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code PSY per year, by decision date, 23 since 2017. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 2 |
| 2018 | 0 |
| 2019 | 2 |
| 2020 | 0 |
| 2021 | 2 |
| 2022 | 1 |
| 2023 | 1 |
| 2024 | 9 |
| 2025 | 2 |
| 2026 (so far) | 4 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K253554 | Aperio GT 450 DX; Aperio GT 180 DXLeica Biosystems Imaging, Inc. | Aug 7, 2026266 days |
| K260964 | NanoZoomer S540MD Slide Scanner SystemHamamatsu Photonics K.K. | Jul 22, 2026121 days |
| K251952 | VirasightVirasoft Corporation | Jul 17, 2026387 days |
| K253606 | NanoZoomer S20MD Slide scanner system; NanoZoomer S360MD Slide scanner systemHamamatsu Photonics K.K. | Jul 15, 2026239 days |
| K243871 | Philips IntelliSite Pathology Solution 5.1Philips Medical Systems Nederland B.V. | Mar 6, 202579 days |
| K241717 | E1000 Dx Digital Pathology SolutionShandon Diagnostics Limited | Feb 28, 2025259 days |
| K242783 | Roche Digital Pathology DxVentana Medical Systems, Inc. | Dec 17, 202492 days |
| K242848 | Philips IntelliSite Pathology Solution 5.1Philips Medical Systems Nederland B.V. | Dec 10, 202481 days |
| K241871 | Philips IntelliSite Pathology SolutionPhilips Medical Systems Nederland B.V. | Dec 2, 2024158 days |
| K233204 | Philips IntelliSite Pathology Solution 5.1Philips Medical Systems Nederland B.V. | Jun 24, 2024270 days |
Recalls
Recall records for product code PSY per year, by the date the recall began. Each recalled product is one record; these 2 records belong to 2 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2913-2026Class II | Ventana Medical Systems, Inc.Due to complaints received for the Image Management System regarding image/metadata discrepancy with the Viewer window | Apr 20, 2026 |
| Z-2280-2023Class II | Hamamatsu CorporationIf the wrong barcode information is recorded in the NDPi file, a pathologist may refer to the whole slide image (WSI) of a different patient. | Jun 6, 2023 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code PSY. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 5 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions5
- Other or not stated0
Most reported device problems, last 12 months
- Application Program Problem3
- Application Program Freezes or Fails to Launch1
- Computer Operating System Problem1
- Insufficient Device Problem Information1
- Program or Algorithm Execution Failure1
In plain English: what the newest reports describe Written by AI
Two reports describe a viewer problem in NavifyPath Digital Pathology where a slide image is shown with the wrong label or from the wrong specimen. One report says the cause is specimen names with leading zeros mixed with names of 100 or higher, which shifts the slide position in the viewer. The manufacturer says the images match correctly in the data management section.
2 reports: 2028492-2026-01688, 2028492-2026-01689
One report describes a software problem in the NavifyPath digital pathology system where the viewer shows the wrong patient information: surnames, dates of birth, ages and gender are assigned incorrectly, or display as placeholder text like XXXX, which forces staff to check the lab information system to match patients to cases. The first report says this causes delays.
Report: 2028492-2026-02791
One report describes a problem where the line measurement annotation on the image does not match the image's visual scale, sometimes. Pathologists use these measurements for things like melanoma thickness and margins, which are part of a primary diagnosis. The customer says no patient diagnosis has been affected.
Report: 2028492-2026-02306
One report describes a scanner that crashed while a slide was being processed, breaking the slide. The manufacturer's technical specialist reported that only one slide broke, that it affected one patient, and that enough spare slides were made from the same sample, so testing was not hindered.
Report: 1419341-2026-00008
One report describes a scanner in the United Kingdom where incorrectly printed barcode labels caused two slides to be saved with only a date and time as the file name. The manufacturer says the scan's reference data showed information from the previous slide, and it chose to file a report out of caution because a similar US scanner exists.
Report: 0009614683-2023-00001
Written by AI from the narratives of the 8 newest reports that have one, received May 9, 2019 to Jul 10, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 21in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 25registered establishments that list this product code
Other codes under 21 CFR 864.3700
Product codes classified under the same regulation.
Watch PSY in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get PSY every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PSY. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:PSYRetrieved Nov 29, 2026, 7:42 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PSYRetrieved Nov 29, 2026, 7:42 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PSY · count decision_dateRetrieved Nov 29, 2026, 7:42 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:PSYRetrieved Nov 29, 2026, 7:42 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PSY · count event_date_initiatedRetrieved Nov 29, 2026, 7:42 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PSY · count res_event_numberRetrieved Nov 29, 2026, 7:42 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2913-2026" "Z-2280-2023")Retrieved Nov 29, 2026, 7:42 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PSY AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:42 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PSY AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:42 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PSY AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:42 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:PSYRetrieved Nov 29, 2026, 7:42 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:PSYRetrieved Nov 29, 2026, 7:42 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:864.3700Retrieved Nov 28, 2026, 8:53 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil