FDA product code
PQP: Next Generation Sequencing Oncology Panel, Somatic Or Germline Variant Detection System (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class III
- Regulation
- None listed
- Review panel
- Pathology
- Premarket route
- PMA
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- PMA decisions
- 24518 original PMAs, the rest supplements
- Median review time
- 180 daysthe latest 25 decisions
- Recall records
- 183 in the last 12 months
- Adverse event reports
- 7last 12 months, 0 deaths
PMA decisions
Original PMAs and supplements for product code PQP per year, by decision date, 245 since 2016. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 6 |
| 2018 | 15 |
| 2019 | 21 |
| 2020 | 25 |
| 2021 | 17 |
| 2022 | 33 |
| 2023 | 34 |
| 2024 | 34 |
| 2025 | 21 |
| 2026 (so far) | 38 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| P240040 S008 | Oncomine Dx Express TestLife Technologies Corporation | Nov 4, 202675 days |
| P210011 S006 | xT CDxTempus AI, Inc. | Oct 23, 202630 days |
| P240010 S007 | MI Cancer SeekCaris Life Sciences | Oct 16, 2026324 days |
| P190032 S032 | FoundationOne Liquid CDx (F1LCDx)Foundation Medicine, Inc. | Sep 18, 202630 days |
| P200010 S029 | Guardant360 CDxGuardant Health, Inc. | Sep 18, 2026190 days |
| P170019 S069 | FoundationOne CDxFoundation Medicine, Inc. | Sep 16, 2026180 days |
| P240040 S005 | Oncomine™ Dx Express TestLife Technologies Corporation | Sep 9, 2026180 days |
| P230011 S008 | TruSight Oncology ComprehensiveIllumina, Inc. | Sep 4, 202616 days |
| P200010 S024 | Guardant360® CDxGuardant Health, Inc. | Sep 4, 2026463 days |
| P210011 S002 | xT CDxTempus AI, Inc. | Sep 2, 2026572 days |
Recalls
Recall records for product code PQP per year, by the date the recall began. Each recalled product is one record; these 18 records belong to 14 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 1 |
| 2018 | 2 |
| 2019 | 1 |
| 2020 | 4 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 2 |
| 2024 | 1 |
| 2025 | 2 |
| 2026 (so far) | 3 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2951-2026Class II | Foundation Medicine, Inc.Software issue that can cause an increase in spontaneous software closures. | Jun 10, 2026 |
| Z-1937-2026Class II | Foundation Medicine, Inc.Potential for spontaneous closures of Hamilton HxRunControl software during execution of Autolys protocols after deployment of the PCT software for AutoLys Ham… | Mar 24, 2026 |
| Z-1927-2026Class II | Foundation Medicine, Inc.Possible increase in phasing, potentially resulting in false-positive BARD1 splice site indel artifact variants | Mar 20, 2026 |
| Z-2521-2025Class II | CARIS LIFE SCIENCESDue to an incorrect test results provided that indicated the incorrect drug therapy recommendation | Jun 17, 2025 |
| Z-1747-2025Class II | Foundation Medicine, Inc.Four (4) total reports were sent to customers with clinically relevant variants requiring amendment, one (1) of which with an incorrect negative CDx claim. The… | Apr 9, 2025 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code PQP. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 1 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 1 |
| 2022 | 1 |
| 2023 | 2 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 7 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions7
- Other or not stated0
Most reported device problems, last 12 months
- Product Quality Problem6
- False Negative Result1
In plain English: what the newest reports describe Written by AI
A computer analysis step reported false BARD1 gene variants in stretches of repeated DNA letters. The manufacturer linked this to one supplied reagent lot, and the affected reports were amended. No patient harm was reported.
4 reports: 3010679023-2026-00004, 3010679023-2026-00002, 3010679023-2026-00006, 3010679023-2026-00005
The test may miss an FGFR3 gene change when it is present at very low levels. The manufacturer says DNA folds up in a region rich in guanine and cytosine building blocks, which blocks copying of the mutant DNA. The manufacturer reports no death or serious injury.
Report: 3003335080-2026-00001
A lung cancer test falsely showed an EGFR L858R gene change as positive. The manufacturer issued a corrected report, but the treating doctor had already chosen therapy based on the first result. It is unclear whether harm resulted.
Report: 3012090384-2023-00001
Low room temperature in the lab, near the bottom of the instrument's operating range, pushed microsatellite instability scores below the cutoff, so samples were misclassified. No patient samples were run, so no patient was affected.
Report: 3003335080-2022-00001
A manufacturing error left too little oligonucleotide (short DNA strands) in one RNA panel component. This could make the test fail or give false negatives for ROS1 gene changes in lung cancer, which could exclude a patient from a targeted lung cancer drug. The manufacturer found the problem internally.
Report: 3003335080-2018-00001
Written by AI from the narratives of the 12 newest reports that have one, received Dec 15, 2017 to May 13, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 30in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 18registered establishments that list this product code
Watch PQP in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get PQP every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PQP. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:PQPRetrieved Nov 29, 2026, 1:17 AM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:PQPRetrieved Nov 29, 2026, 9:37 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:PQP · count decision_dateRetrieved Nov 29, 2026, 1:17 AM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:PQP · count pma_numberRetrieved Nov 29, 2026, 1:17 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:PQPRetrieved Nov 29, 2026, 9:37 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PQP · count event_date_initiatedRetrieved Nov 29, 2026, 1:17 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PQP · count res_event_numberRetrieved Nov 29, 2026, 1:17 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2951-2026" "Z-1937-2026" "Z-1927-2026" "Z-2521-2025" "Z-1747-2025")Retrieved Nov 29, 2026, 9:37 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PQP AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 9:37 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PQP AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 9:37 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PQP AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 1:17 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:PQPRetrieved Nov 29, 2026, 1:17 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:PQPRetrieved Nov 29, 2026, 1:17 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil