Product codes

FDA product code

PQF: Sensor, Glucose, Invasive, Non-Adjunctive (FDA product code)

Follow PQF: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class III
Regulation
None listed
Review panel
Clinical Chemistry
Premarket route
PMA
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
PMA decisions
92 original PMAs, the rest supplements
Median review time
123 daysthe latest 9 decisions
Recall records
40 in the last 12 months
Adverse event reports
22,049last 12 months, 0 deaths

PMA decisions

Original PMAs and supplements for product code PQF per year, by decision date, 9 since 2016. Review time is the days from receipt to decision.

All of them in the FDA’s PMA database

PMA decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20173
20182
20191
20200
20210
20220
20230
20242
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
P160007 S052 Simplera (TM) SystemMedtronic Minimed Sep 17, 202427 days
P160007 S047 Simplera™ SystemMedtronic Minimed Jul 24, 2024525 days
P120005 S083 DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEMDexcom, Inc. Nov 1, 2019127 days
P120005 S077 G5 Mobile Continuous Glucose Monitoring SystemDexcom, Inc. Oct 11, 201864 days
P120005 S073 G5 Mobile transmitter Hardware and Firmware, G5 Mobile Continuous Glucose Monitoring SystemDexcom, Inc. Jul 3, 201890 days
P120005 S065 Dexcon G5 Mobile Continuous Glucose Monitoring SystemDexcom, Inc. Sep 21, 201773 days
P120005 S059 DEXCOM G4 PLATINUM CONTIUOUS GLUCOSE MONITORING SYSTEMDexcom, Inc. May 22, 2017123 days
P120005 S050 Dexcom G5 Mobile Continuous Glucose Monitoring SystemDexcom, Inc. Mar 8, 2017259 days
P120005 S041 DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEMDexcom, Inc. Dec 20, 2016452 days

Recalls

Recall records for product code PQF per year, by the date the recall began. Each recalled product is one record; these 4 records belong to 3 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20202
20210
20222
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1811-2022Class II Tandem Diabetes Care IncInsulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On,… May 24, 2022
Z-1812-2022Class II Tandem Diabetes Care IncInsulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On,… May 24, 2022
Z-1258-2022Class II Tandem Diabetes Care IncInsulin pumps may display unexpected fluctuation of the remaining battery life so ensure pump battery life does not fall below 25% remaining power. If the batt… Sep 10, 2020
Z-1561-2020Class II Dexcom IncThe company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous… Feb 3, 2020

Adverse event reports (MAUDE)

Medical device reports received per year that name product code PQF. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201765,182
2018116,245
201944,510
202010,579
20214,943
20221,378
20236,373
202422,392
202538,645
2026 (so far)18,325

Last 12 months by event type

  • Deaths0
  • Injuries60
  • Malfunctions21,989
  • Other or not stated0

Most reported device problems, last 12 months

  1. Wireless Communication Problem21,539
  2. Display or Visual Feedback Problem727
  3. Incorrect, Inadequate or Imprecise Result or Readings574
  4. Calibration Problem439
  5. Battery Problem404
  6. Inappropriate or Unexpected Reset59

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
17in the FDA’s Global Unique Device Identification Database
Establishment listings
9registered establishments that list this product code

Watch PQF in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get PQF every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PQF. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.