FDA product code
PNG: Peritoneal, Drainage Catheter For Refractory Ascites, Long-Term Indwelling (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 876.5630
- Review panel
- Gastroenterology and Urology
- Premarket route
- 510(k)
- Implant
- Yes
- Life-sustaining or supporting
- Yes
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 150 in the last 12 months
- Median review time
- 141 daysthe latest 15 decisions
- Recall records
- 130 in the last 12 months
- Adverse event reports
- 47last 12 months, 1 death
510(k) and De Novo decisions
Decisions for product code PNG per year, by decision date, 15 since 2005. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 2 |
| 2024 | 1 |
| 2025 | 1 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K241278 | ASEPT® Glide Peritoneal Drainage SystemPfm Medical, Inc. | Jan 14, 2025253 days |
| K241946 | PeritX™ 1L Drainage Kit (90-1010); PeritX™ 2L Drainage Kit (90-2010); PeritX™ Peritoneal Catheter Kit (50-9000C); PeritX™ Peritoneal Catheter and Starter Kit (50-9900C); PeritX™ Peritoneal Catheter Mini Kit (50-9050A); PleurX™ Low Profile Catheter Mini Kit (50-8050); Valve Cap (50-7235A); PeritX™ Valve Kit (50-7272); Catheter Access Kit (50-7280A); Lockable Drainage Line (50-7245A); Lockable Drainage Line Kit (50-7265A); PleurX™ Supplemental Insertion Kit (50-7262); Procedure Pack (50-Bard Peripheral Vascular, Inc. | Dec 6, 2024156 days |
| K230319 | Passio Pump Drainage SystemBearpac Medical | Nov 24, 2023291 days |
| K221779 | ASEPT Peritoneal Drainage SystemPfm Medical, Inc. | May 2, 2023315 days |
| K212675 | Aspira Peritoneal Drainage SystemMerit Medical Systems, Inc. | Dec 14, 2021112 days |
| K201155 | PleurX Peritoneal Catheter SystemCare Fusion | Oct 21, 2020174 days |
| K162457 | Rocket Indwelling Peritoneal Catheter (IPC) Insertion KitRocket Medical Plc | Nov 14, 201673 days |
| K160437 | PleurX Peritoneal Catheter SystemCare Fusion | Nov 3, 2016260 days |
| K152105 | Rocket IPC Insertion Set, Rocket IPC Mini Insertion SetRocket Medical Plc | Mar 29, 2016244 days |
| K122422 | PLEURX PLEURAL CATHETER SYSTEM PLEURX CATHETER ACCESS KIT PLEURX SLIDE VALVED PEELABLE INTRODUCERCare Fusion | Oct 24, 201276 days |
Recalls
Recall records for product code PNG per year, by the date the recall began. Each recalled product is one record; these 13 records belong to 1 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 13 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1908-2023Class II | Merit Medical Systems, Inc.There is a potential that valve assemblies will not open. preventing fluid from draining. | Apr 22, 2023 |
| Z-1909-2023Class II | Merit Medical Systems, Inc.There is a potential that valve assemblies will not open. preventing fluid from draining. | Apr 22, 2023 |
| Z-1910-2023Class II | Merit Medical Systems, Inc.There is a potential that valve assemblies will not open. preventing fluid from draining. | Apr 22, 2023 |
| Z-1911-2023Class II | Merit Medical Systems, Inc.There is a potential that valve assemblies will not open. preventing fluid from draining. | Apr 22, 2023 |
| Z-1912-2023Class II | Merit Medical Systems, Inc.There is a potential that valve assemblies will not open. preventing fluid from draining. | Apr 22, 2023 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code PNG. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 3 |
| 2018 | 6 |
| 2019 | 9 |
| 2020 | 15 |
| 2021 | 15 |
| 2022 | 18 |
| 2023 | 21 |
| 2024 | 18 |
| 2025 | 31 |
| 2026 (so far) | 44 |
Last 12 months by event type
- Deaths1
- Injuries2
- Malfunctions44
- Other or not stated0
Most reported device problems, last 12 months
- Incomplete or Inadequate Connection22
- Device Dislodged or Dislocated14
- Break12
- Fluid/Blood Leak5
- Adverse Event Without Identified Device or Use Problem3
- Leak/Splash3
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 49in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 41registered establishments that list this product code
Other codes under 21 CFR 876.5630
Product codes classified under the same regulation.
Watch PNG in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get PNG every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PNG. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:PNGRetrieved Nov 29, 2026, 1:10 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PNGRetrieved Nov 29, 2026, 4:33 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PNG · count decision_dateRetrieved Nov 29, 2026, 1:10 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:PNGRetrieved Nov 29, 2026, 4:33 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PNG · count event_date_initiatedRetrieved Nov 29, 2026, 1:10 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PNG · count res_event_numberRetrieved Nov 29, 2026, 1:10 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1908-2023" "Z-1909-2023" "Z-1910-2023" "Z-1911-2023" "Z-1912-2023")Retrieved Nov 29, 2026, 4:33 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PNG AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:33 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PNG AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:33 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PNG AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 1:10 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:PNGRetrieved Nov 29, 2026, 1:10 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:PNGRetrieved Nov 29, 2026, 1:10 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:876.5630Retrieved Nov 28, 2026, 9:52 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.