FDA product code

PNE: Reprocessed Catheter Introducer (FDA product code)

Follow PNE: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 870.1340
Review panel
Cardiovascular
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
72 in the last 12 months
Median review time
241 daysthe latest 7 decisions
Recall records
00 in the last 12 months
Adverse event reports
17last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code PNE per year, by decision date, 7 since 2016. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20171
20180
20190
20200
20210
20221
20231
20241
20251
2026 (so far)1

Latest decisions

NumberDevice and applicantDecision and review time
K250305 Reprocessed Agilis NxT Steerable IntroducerInnovative Health Jan 7, 2026338 days
K250314 Medline ReNewal Reprocessed Abbott Agilis Nxt Steerable IntroducerSurgical Instrument Service and Savings, Inc. Dec 19, 2025318 days
K232037 Reprocessed VersaCross Steerable Sheath; Reprocessed SureFlex Steerable Guiding SheathInnovative Health, LLC Apr 4, 2024269 days
K230376 Reprocessed Agilis NxT Steerable IntroducerInnovative Health, LLC Aug 7, 2023175 days
K212165 Reprocessed Carto Vizigo Bi-Directional Guiding SheathInnovative Health, LLC Mar 10, 2022241 days
K170311 Reprocessed Agilis NxT Steerable IntroducerInnovative Health, LLC Jun 2, 2017121 days
K152090 Reprocessed Steerable IntroducerSterilmed, Inc. Mar 23, 2016240 days

Recalls

Recall records for product code PNE per year, by the date the recall began. Each recalled product is one record.

All of them in the FDA’s recall database

No recalls on record for this product code.

Adverse event reports (MAUDE)

Medical device reports received per year that name product code PNE. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20172
20180
20190
20200
20211
20222
20235
20244
20250
2026 (so far)17

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions17
  • Other or not stated0

Most reported device problems, last 12 months

  1. Fluid/Blood Leak12
  2. Material Puncture/Hole7
  3. Gas/Air Leak4
  4. Mechanical Problem2
  5. Break1
  6. Detachment of Device or Device Component1

In plain English: what the newest reports describe Written by AI

  • Several reports describe blood or fluid leaking from the valves or ports of the introducer sheath, including the bleed-back valve (the valve at the end of the handle that stops blood flowing out) and the hemostasis valve (the seal that keeps blood in when tools pass through). Some reports say the valve was leaking when first placed into the body.

    8 reports: 3011610434-2026-00010, 3011610434-2026-00009, 3011610434-2026-00006, 3011610434-2026-00012, 3011610434-2023-00011, 3011610434-2023-00010, 3011610434-2024-00007, 3011610434-2026-00013

  • Several older reports describe leaks or faulty valves in the sheath, with no injuries reported. In one, a reprocessed sheath was used and the new sheath crossed the septum (the wall between the heart's upper chambers) after balloon use.

    5 reports: 3011610434-2024-00011, 3011610434-2024-00004, 3011610434-2023-00012, 3011610434-2023-00007, 19070555

  • Several reports describe the side port or hub connection leaking blood or fluid. In one, fluid flushed out of the side port instead of into the body, and in another blood leaked where the tubing meets the hub.

    4 reports: 3011610434-2026-00002, 3011610434-2026-00008, 3011610434-2026-00007, 3011610434-2026-00013

  • Some reports describe bubbles in the line and poor maneuverability of the sheath, meaning it was hard to steer. One report also describes a small cut in the hemostasis tubing near the hub, after which the device was exchanged.

    2 reports: 3011610434-2026-00003, 3011610434-2026-00007

  • A report describes air being drawn into the line. The physician saw 10 mL of air when flushing after the needle and dilator were removed. The facility said a possible transient ischemic attack (a temporary stroke-like event) was suspected but was ruled out after imaging.

    Report: 3011610434-2026-00001

Written by AI from the narratives of the 20 newest reports that have one, received Apr 6, 2023 to Jul 10, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
24in the FDA’s Global Unique Device Identification Database
Establishment listings
11registered establishments that list this product code

Watch PNE in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get PNE every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PNE. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.