FDA product code
PNB: Biliary Stent System For Benign Strictures (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 876.5011
- Review panel
- Gastroenterology and Urology
- Premarket route
- 510(k)
- Implant
- Yes
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 10 in the last 12 months
- Median review time
- 280 daysthe latest 1 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 27last 12 months, 5 deaths
510(k) and De Novo decisions
Decisions for product code PNB per year, by decision date, 1 since 2016. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| DEN150040 | WallFlex Biliary RX Fully Covered Stent System RMVBoston Scientific | Jun 3, 2016280 days |
Recalls
Recall records for product code PNB per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code PNB. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 22 |
| 2018 | 48 |
| 2019 | 46 |
| 2020 | 48 |
| 2021 | 29 |
| 2022 | 16 |
| 2023 | 27 |
| 2024 | 39 |
| 2025 | 37 |
| 2026 (so far) | 25 |
Last 12 months by event type
- Deaths5
- Injuries16
- Malfunctions6
- Other or not stated0
Most reported device problems, last 12 months
- Device Stenosis11
- Migration7
- Break6
- Difficult to Remove6
- Adverse Event Without Identified Device or Use Problem4
- Material Deformation3
In plain English: what the newest reports describe Written by AI
In three reports the stent moved out of its intended position in the bile duct. In two, the patient needed a repeat procedure and a new stent, and one had pain, jaundice (yellowing of the skin) and raised liver tests.
3 reports: 3005099803-2026-02760, 3005099803-2026-02619, 3005099803-2026-02585
In three reports the stent broke, sheared, or failed to open fully during placement. In two of them another stent was used to finish the procedure, and in the third the outcome was not reported.
3 reports: 3005099803-2026-01907, 3005099803-2026-02399, 3005099803-2026-01526
In two reports the stent became stuck to the bile duct wall. In one, tissue grew into it and was bitten away with forceps to remove it; in the other, the mesh broke down, the stent became embedded and blocked, and repeated removal attempts failed.
2 reports: 3005099803-2026-04112, 3005099803-2026-03393
In two reports the stent was damaged during removal. One says the wires at one end were mangled and the stent was broken and deformed, and the other says retrieval loops broke and the stent shrank unevenly before a snare was used.
2 reports: 3005099803-2026-02582, 3005099803-2026-03874
In one report, a stent was found already opened inside its sealed package, even though the packaging looked undamaged. A new stent was opened and used, and no patient complications were reported.
Report: 3005099803-2026-03853
Written by AI from the narratives of the 16 newest reports that have one, received Apr 27, 2026 to Oct 13, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 5in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 5registered establishments that list this product code
Watch PNB in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get PNB every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PNB. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:PNBRetrieved Nov 29, 2026, 11:13 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PNBRetrieved Nov 29, 2026, 11:13 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PNB · count decision_dateRetrieved Nov 29, 2026, 11:13 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:PNBRetrieved Nov 29, 2026, 11:13 PM UTC.
- device/recall · search product_code:PNB · count event_date_initiatedRetrieved Nov 29, 2026, 11:13 PM UTC.
- device/recall · search product_code:PNB · count res_event_numberRetrieved Nov 29, 2026, 11:13 PM UTC.
- device/event · search device.device_report_product_code:PNB AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 11:13 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PNB AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 11:13 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PNB AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 11:13 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:PNBRetrieved Nov 29, 2026, 11:13 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:PNBRetrieved Nov 29, 2026, 11:13 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:876.5011Retrieved Nov 29, 2026, 11:13 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.