Product codes

FDA product code

PKT: Laser For Disruption Of Adipocyte Cells For Aesthetic Use (FDA product code)

Follow PKT: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 878.5400
Review panel
General and Plastic Surgery
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
Yes
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
190 in the last 12 months
Median review time
97 daysthe latest 19 decisions
Recall records
00 in the last 12 months
Adverse event reports
2last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code PKT per year, by decision date, 19 since 2015. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20172
20180
20193
20202
20212
20222
20233
20242
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K240658 1060nm laser body slimming machineHebei JT Medical Co., Ltd. May 13, 202466 days
K233962 ReBorn (1050nm)Lightfective , Ltd. Apr 16, 2024123 days
K231971 DEKA PHYSIQ 360El.En S.P.A. Sep 26, 202385 days
K231131 Diode Laser Body Sculpture SystemsShanghai Bele Medical Technology Co.,Ltd Jun 22, 202362 days
K222226 EONDominion Aesthetic Technologies, Inc. Feb 17, 2023207 days
K222265 DiosculptShanghai Wonderful Opto-Electrics Tech.Co.,Ltd Sep 23, 202257 days
K211681 EONDominion Aesthetic Technologies, Inc. Mar 30, 2022302 days
K212331 BodySculpRohrer Aesthetics, LLC Nov 9, 2021105 days
K211402 Powersculp laser lipolysis systemLotuxs Medtech (Suzhou) Co., Ltd. Jul 30, 202185 days
K192970 SLIMUSHironic Co., Ltd. Dec 1, 2020405 days

Recalls

Recall records for product code PKT per year, by the date the recall began. Each recalled product is one record.

All of them in the FDA’s recall database

No recalls on record for this product code.

Adverse event reports (MAUDE)

Medical device reports received per year that name product code PKT. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201710
201816
201913
20205
20216
20224
20230
20244
20255
2026 (so far)2

Last 12 months by event type

  • Deaths0
  • Injuries2
  • Malfunctions0
  • Other or not stated0

Most reported device problems, last 12 months

  1. Adverse Event Without Identified Device or Use Problem1
  2. Use of Device Problem1

In plain English: what the newest reports describe Written by AI

  • Several reports describe burns on the skin during laser treatment, in areas including the flank, abdomen and skin folds. One patient needed over-the-counter burn ointment and was left with a scar. The reports do not say what caused the burns.

    8 reports: 3006515340-2025-00002, MW5176753, 3006515340-2026-00004, 3031944951-2025-00001, MW5164178, 1222993-2024-00001, 1222993-2022-00033, 1222993-2021-00026

  • Several reports describe firm lumps, or nodules, under the skin after laser treatment, on the calves, abdomen and arm. In some cases the lump lasted many months, and one patient received steroid injections for it. One patient also had arm swelling and was hospitalized for cellulitis, a skin infection.

    6 reports: 1222993-2024-00002, 1222993-2022-00004, 1222993-2021-00046, 1222993-2021-00027, 1222993-2021-00021, 1222993-2021-00034

  • One report describes two open skin wounds after the applicator lost contact with the skin during a treatment. A photo showed crusting and redness, and the provider documented an open wound reaching deeper skin layers that was healing slowly.

    Report: 3031944951-2026-00008

  • One report describes smoke and a small hole with melting on the cable linking a laser applicator to the main unit. The manufacturer found the cable had been twisted and kinked in a way that goes against the instructions for use, and said no patients were harmed.

    Report: MW5172904

  • One report describes air trapped in the chest and under the skin two days after a treatment, causing chest tightness. The patient was hospitalized for two days and given oxygen. The report does not say what caused the air to become trapped.

    Report: MW5150226

Written by AI from the narratives of the 20 newest reports that have one, received Jun 10, 2021 to Aug 31, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
19in the FDA’s Global Unique Device Identification Database
Establishment listings
28registered establishments that list this product code

Product codes classified under the same regulation.

Watch PKT in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get PKT every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PKT. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.