FDA product code
PFF: High Throughput Dna Sequence Analyzer (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 862.2265
- Review panel
- Clinical Toxicology
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 20 in the last 12 months
- Median review time
- 100 daysthe latest 2 decisions
- Recall records
- 181 in the last 12 months
- Adverse event reports
- 1last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code PFF per year, by decision date, 2 since 2013. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
Recalls
Recall records for product code PFF per year, by the date the recall began. Each recalled product is one record; these 18 records belong to 9 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 4 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 3 |
| 2023 | 2 |
| 2024 | 7 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0211-2027Class II | Life Technologies Holdings Pte LtdDNA or RNA sequencing system software has vulnerabilities. | Aug 14, 2026 |
| Z-0068-2027Class II | Illumina, Inc.Retrospectively reported correction for an instrument intended for targeted sequencing of DNA libraries from human genomic DNA used with in vitro diagnostic as… | Sep 17, 2024 |
| Z-0320-2027Class II | Illumina, Inc.(Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, serv… | May 8, 2024 |
| Z-0321-2027Class II | Illumina, Inc.(Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, serv… | May 8, 2024 |
| Z-0322-2027Class II | Illumina, Inc.(Retrospectively reported) Cybersecurity vulnerability identified in communication protocol between the customer-provided infrastructure (network storage, serv… | May 8, 2024 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code PFF. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 1 |
| 2022 | 1 |
| 2023 | 1 |
| 2024 | 1 |
| 2025 | 2 |
| 2026 (so far) | 1 |
Last 12 months by event type
- Deaths0
- Injuries1
- Malfunctions0
- Other or not stated0
Most reported device problems, last 12 months
- Device Fell1
- Leak/Splash1
In plain English: what the newest reports describe Written by AI
Fluid leaks from the sequencer's reagent and waste parts appear in four reports. They include a cartridge dropped during delivery, a waste bottle not placed correctly, and a waste spout leak that ran about 10 hours before it was found.
4 reports: 3007102730-2022-00001, 3007102730-2023-00001, 3007102730-2021-00001, 3007102730-2026-00001
A run was started with no waste cartridge loaded, and the instrument did not detect the missing cartridge. The narrative is cut off before describing what followed.
Report: 3007102730-2024-00001
During a service visit to install a new solid state drive (a data storage part), an engineer saw smoke and discoloration on a flow sensor cable and a burnt hole in a black cloth. The narrative traces the source to that cable.
Report: 3005095756-2025-00001
A clinical genetics program reported a patient who received a consumer genetic report, generated by an AI system, claiming several serious neurological conditions. The report was based only on raw genomic data, without the expert variant review normally required for clinical diagnoses.
Report: MW5176790
A patient received a positive result for Turner syndrome from a prenatal screening test, but amniocentesis later showed normal cells. The patient said the false result caused four weeks of severe distress.
Report: MW5041521
Written by AI from the narratives of the 8 newest reports that have one, received Mar 13, 2015 to Jun 15, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 30in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 15registered establishments that list this product code
Watch PFF in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get PFF every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PFF. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:PFFRetrieved Nov 29, 2026, 3:18 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PFFRetrieved Nov 29, 2026, 3:36 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PFF · count decision_dateRetrieved Nov 29, 2026, 3:18 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:PFFRetrieved Nov 29, 2026, 3:36 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PFF · count event_date_initiatedRetrieved Nov 29, 2026, 3:18 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PFF · count res_event_numberRetrieved Nov 29, 2026, 3:18 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0211-2027" "Z-0068-2027" "Z-0320-2027" "Z-0321-2027" "Z-0322-2027")Retrieved Nov 29, 2026, 3:36 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PFF AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 3:36 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PFF AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 3:36 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PFF AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 12:58 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:PFFRetrieved Nov 29, 2026, 3:18 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:PFFRetrieved Nov 29, 2026, 3:18 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:862.2265Retrieved Nov 29, 2026, 3:18 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.