Product codes

FDA product code

PEO: Fungal Organisms, Nucleic Acid-Based Assay (FDA product code)

Follow PEO: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 866.3365
Review panel
Microbiology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
510(k) and De Novo decisions
20 in the last 12 months
Median review time
88 daysthe latest 2 decisions
Recall records
00 in the last 12 months
Adverse event reports
1last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code PEO per year, by decision date, 2 since 2018. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20181
20190
20200
20210
20220
20230
20241
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K240627 LIAISON PLEX Yeast Blood Culture AssayLuminex Corporation Jun 4, 202490 days
K182690 ePlex Blood Culture Identification Panel Fungal Pathogen (BCID-FP) PanelGenmark Diagnostics, Incorporated Dec 21, 201885 days

Recalls

Recall records for product code PEO per year, by the date the recall began. Each recalled product is one record.

All of them in the FDA’s recall database

No recalls on record for this product code.

Adverse event reports (MAUDE)

Medical device reports received per year that name product code PEO. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
20180
20190
20200
20210
20220
20231
20241
20252
2026 (so far)1

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions1
  • Other or not stated0

Most reported device problems, last 12 months

  1. False Negative Result1

In plain English: what the newest reports describe Written by AI

  • The panel gave negative results for Candida auris in three patient samples, but culture later grew Candida auris from them, so the test missed a fungus that culture found.

    Report: 3008632402-2026-00163

  • Two samples gave conflicting species results on repeat testing. In one, the repeat matched the first result for Candida famata, but culture found Candida dubliniensis. In the other, the repeat dropped Candida dubliniensis, which culture did find.

    Report: 3008632402-2025-02245

  • A blood test reported Staphylococcus aureus with the mecA marker not detected, but culture showed the isolate was methicillin-resistant, which the mecA result should have shown. MecA is a marker for resistance to methicillin, a common antibiotic. The facility then stopped testing positive blood cultures with gram-positive cocci in clusters while it investigated reports of similar problems elsewhere.

    Report: 21949700

  • One sample first gave negative results, but identification of the growth by MALDI, a lab method that identifies organisms by their protein profile, showed Candida lusitaniae with a high confidence score. In a second sample, the test positive for Candida parapsilosis, and culture grew both Candida parapsilosis and Candida lusitaniae.

    Report: 3008632402-2024-03436

  • A facility reported false positive Candida tropicalis results on several blood culture bottle types when tested with the molecular panel. The facility believes the cause is DNA from an organism contaminating the culture media. The manufacturer disputed that there was a problem, and the facility said several other facilities in its region had seen the same issue.

    Report: MW5145662

Written by AI from the narratives of the 5 newest reports that have one, received Sep 14, 2023 to Jan 13, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
9in the FDA’s Global Unique Device Identification Database
Establishment listings
10registered establishments that list this product code

Product codes classified under the same regulation.

Watch PEO in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get PEO every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PEO. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.