FDA product code
PEO: Fungal Organisms, Nucleic Acid-Based Assay (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 866.3365
- Review panel
- Microbiology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 20 in the last 12 months
- Median review time
- 88 daysthe latest 2 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 1last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code PEO per year, by decision date, 2 since 2018. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
Recalls
Recall records for product code PEO per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code PEO. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 1 |
| 2025 | 2 |
| 2026 (so far) | 1 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions1
- Other or not stated0
Most reported device problems, last 12 months
- False Negative Result1
In plain English: what the newest reports describe Written by AI
The panel gave negative results for Candida auris in three patient samples, but culture later grew Candida auris from them, so the test missed a fungus that culture found.
Report: 3008632402-2026-00163
Two samples gave conflicting species results on repeat testing. In one, the repeat matched the first result for Candida famata, but culture found Candida dubliniensis. In the other, the repeat dropped Candida dubliniensis, which culture did find.
Report: 3008632402-2025-02245
A blood test reported Staphylococcus aureus with the mecA marker not detected, but culture showed the isolate was methicillin-resistant, which the mecA result should have shown. MecA is a marker for resistance to methicillin, a common antibiotic. The facility then stopped testing positive blood cultures with gram-positive cocci in clusters while it investigated reports of similar problems elsewhere.
Report: 21949700
One sample first gave negative results, but identification of the growth by MALDI, a lab method that identifies organisms by their protein profile, showed Candida lusitaniae with a high confidence score. In a second sample, the test positive for Candida parapsilosis, and culture grew both Candida parapsilosis and Candida lusitaniae.
Report: 3008632402-2024-03436
A facility reported false positive Candida tropicalis results on several blood culture bottle types when tested with the molecular panel. The facility believes the cause is DNA from an organism contaminating the culture media. The manufacturer disputed that there was a problem, and the facility said several other facilities in its region had seen the same issue.
Report: MW5145662
Written by AI from the narratives of the 5 newest reports that have one, received Sep 14, 2023 to Jan 13, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 9in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 10registered establishments that list this product code
Other codes under 21 CFR 866.3365
Product codes classified under the same regulation.
Watch PEO in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get PEO every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PEO. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:PEORetrieved Nov 29, 2026, 12:56 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PEORetrieved Nov 29, 2026, 6:29 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PEO · count decision_dateRetrieved Nov 29, 2026, 12:56 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:PEORetrieved Nov 29, 2026, 6:29 PM UTC.
- device/recall · search product_code:PEO · count event_date_initiatedRetrieved Nov 29, 2026, 12:56 AM UTC.
- device/recall · search product_code:PEO · count res_event_numberRetrieved Nov 29, 2026, 12:56 AM UTC.
- device/event · search device.device_report_product_code:PEO AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:29 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PEO AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:29 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PEO AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 12:56 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:PEORetrieved Nov 29, 2026, 12:56 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:PEORetrieved Nov 29, 2026, 12:56 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:866.3365Retrieved Nov 29, 2026, 4:47 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil