FDA product code
PCU: Pancreatic Stent, Covered, Metallic, Removable (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 876.5015
- Review panel
- Gastroenterology and Urology
- Premarket route
- 510(k)
- Implant
- Yes
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 111 in the last 12 months
- Median review time
- 140 daysthe latest 11 decisions
- Recall records
- 60 in the last 12 months
- Adverse event reports
- 291last 12 months, 4 deaths
510(k) and De Novo decisions
Decisions for product code PCU per year, by decision date, 11 since 2013. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 2 |
| 2020 | 0 |
| 2021 | 1 |
| 2022 | 1 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K252648 | Niti-S SPAXUS StentTaewoong Medical Co., Ltd. | Mar 13, 2026204 days |
| K220112 | AXIOS Stent and Electrocautery-Enhanced Delivery SystemBoston Scientific Corporation | Sep 7, 2022237 days |
| K203132 | AXIOS Stent and Electrocautery Enhanced Delivery SystemBoston Scientific Corporation | Apr 15, 2021178 days |
| K192043 | AXIOS Stent and Electrocautery Enhanced Delivery SystemBoston Scientific Corporation | Aug 26, 201926 days |
| K181905 | AXIOS Stent and Delivery System, AXIOS Stent and Electrocautery Enhanced Delivery SystemBoston Scientific Corporation | Mar 25, 2019252 days |
| K163272 | AXIOS Stent and Electrocautery Enhanced Delivery SystemBoston Scientific Corporation | Feb 18, 201789 days |
| K153088 | AXIOS Stent with Delivery System, AXIOS Stent with Electro-cautery Enhanced Delivery SystemBoston Scientific Corporation | Dec 11, 201546 days |
| K152572 | AXIOS Stent and Delivery SystemBoston Scientific Corporation | Oct 6, 201527 days |
| K150692 | AXIOS Stent with Electrocautery Enhanced Delivery SystemBoston Scientific Corp | Aug 5, 2015140 days |
| K140561 | AXIOS STENT AND DELIVERY SYSTEM (WITH 10MM X 10MM STENT) AXIOS STENT AND DELIVERY SYSTEM (WITH 15MM X 10MM STENT)Xlumena, Inc. | Apr 23, 201449 days |
Recalls
Recall records for product code PCU per year, by the date the recall began. Each recalled product is one record; these 6 records belong to 1 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 6 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0773-2021Class II | Boston Scientific CorporationThe rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent de… | Dec 1, 2020 |
| Z-0774-2021Class II | Boston Scientific CorporationThe rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent de… | Dec 1, 2020 |
| Z-0775-2021Class II | Boston Scientific CorporationThe rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent de… | Dec 1, 2020 |
| Z-0776-2021Class II | Boston Scientific CorporationThe rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent de… | Dec 1, 2020 |
| Z-0777-2021Class II | Boston Scientific CorporationThe rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent de… | Dec 1, 2020 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code PCU. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 95 |
| 2018 | 90 |
| 2019 | 136 |
| 2020 | 144 |
| 2021 | 162 |
| 2022 | 208 |
| 2023 | 260 |
| 2024 | 417 |
| 2025 | 302 |
| 2026 (so far) | 261 |
Last 12 months by event type
- Deaths4
- Injuries67
- Malfunctions220
- Other or not stated0
Most reported device problems, last 12 months
- Positioning Failure135
- Adverse Event Without Identified Device or Use Problem78
- Use of Device Problem44
- Activation Failure39
- Positioning Problem36
- Activation, Positioning or Separation Problem22
In plain English: what the newest reports describe Written by AI
In six reports, the flange (the flared end that holds the stent in place) would not open or release during the procedure. Physicians tried repeated maneuvers, and in some cases the stent was pulled out with the scope or removed with forceps, and another device was used to finish. One report described the distal part as loose.
6 reports: 3005099803-2026-04182, 3005099803-2026-04169, 3005099803-2026-04083, 3005099803-2026-03917, 3005099803-2026-03887, 3005099803-2026-04110
Three reports describe a stent deployed in the wrong place. In one case the stent was placed entirely inside the fluid collection, and in another the distal flange was misdeployed so the stent did not sit against the bile duct wall. Each stent was removed, and the access site was closed or otherwise managed.
3 reports: 3005099803-2026-03932, 3005099803-2026-03704, 3005099803-2026-03933
Two reports, both from articles, describe bleeding after the stent was in place. One patient had bleeding from the gallbladder wall that needed emergency removal of the gallbladder. The other had massive bleeding after the stent was removed, due to a ruptured splenic artery, and needed embolization, a procedure to block the artery.
2 reports: 3005099803-2026-04071, 3005099803-2026-04072
One report, based on an article, describes a stent that became buried, meaning embedded in the cyst-side end of the tract between the stomach and the cyst. Removal failed with a standard scope and succeeded only after the opening was widened with a balloon and a dual-channel endoscope was used.
Report: 3005099803-2026-04101
One report, based on an article, describes a stent that became completely blocked by a residual gallstone. The blockage was broken into smaller pieces with a laser through the tract, which is a passage made between the organs during the procedure.
Report: 3005099803-2026-04073
Written by AI from the narratives of the 16 newest reports that have one, received Sep 10, 2026 to Oct 19, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 11in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 9registered establishments that list this product code
Watch PCU in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get PCU every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PCU. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:PCURetrieved Nov 29, 2026, 12:53 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PCURetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PCU · count decision_dateRetrieved Nov 29, 2026, 12:53 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:PCURetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PCU · count event_date_initiatedRetrieved Nov 29, 2026, 12:53 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PCU · count res_event_numberRetrieved Nov 29, 2026, 12:53 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0773-2021" "Z-0774-2021" "Z-0775-2021" "Z-0776-2021" "Z-0777-2021")Retrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PCU AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PCU AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PCU AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 12:53 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:PCURetrieved Nov 29, 2026, 12:53 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:PCURetrieved Nov 29, 2026, 12:53 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:876.5015Retrieved Nov 29, 2026, 12:53 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.