FDA product code

PCD: Powered Stairway Chair Lift (FDA product code)

Follow PCD: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 890.5150
Review panel
Physical Medicine
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
00 in the last 12 months
Median review time
Not availablethe latest 0 decisions
Recall records
30 in the last 12 months
Adverse event reports
47last 12 months, 3 deaths

510(k) and De Novo decisions

Decisions for product code PCD per year, by decision date. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

No decisions on record for this product code.

Recalls

Recall records for product code PCD per year, by the date the recall began. Each recalled product is one record; these 3 records belong to 3 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20171
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-2060-2017Class II Savaria Concord Lifts, Inc.The electrical controller, combined with motor, causes feedback, damaging a component on the control board, starting a fire. Apr 12, 2017
Z-0681-2022Class II Acorn Stairlifts, IncThe design of the seats did not consistently have the adequate strength required to support the customers weight with continued use. Apr 20, 2016
Z-1530-2016Class II Acorn Stairlifts IncAluminum rivets holding the base squab to the framer under prolonged customer use were seen to elongate and subsequently break. Feb 17, 2016

Adverse event reports (MAUDE)

Medical device reports received per year that name product code PCD. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201712
201812
201916
20209
202120
202221
202321
202415
202521
2026 (so far)44

Last 12 months by event type

  • Deaths3
  • Injuries28
  • Malfunctions16
  • Other or not stated0

Most reported device problems, last 12 months

  1. Use of Device Problem5
  2. Improper or Incorrect Procedure or Method4
  3. Mechanical Problem4
  4. Installation-Related Problem3
  5. Patient Device Interaction Problem3
  6. Adverse Event Without Identified Device or Use Problem2

In plain English: what the newest reports describe Written by AI

  • Several reports describe a stairlift stopping partway up or down the stairs, and riders then trying to leave the seat or walk the stairs and falling. In some cases the rider was not wearing the seatbelt, and the reports describe injuries such as broken ribs, fractures and head injuries.

    11 reports: 3003124453-2026-00020, 3003124453-2026-00018, 3003124453-2026-00017, 3003124453-2026-00016, 3003124453-2026-00012, 3011268530-2026-00007, 3003124453-2026-00011, 3011268530-2026-00009, 3003124453-2026-00019, 3042175755-2026-00003, 3042175755-2026-00001

  • Two reports describe a rider who stepped off the lift before reaching the bottom of the stairs, believing they had arrived, and fell. One of these reports says the rider was blind.

    2 reports: 3013423626-2026-00002, 3003124453-2026-00016

  • Two reports describe the lift sliding or running down the rail. One says a rack separated from the rail assembly near the top landing and the lift slid down, injuring a caregiver standing behind it. The other says the unit ran off the rail and stopped itself at the end of the track.

    2 reports: 3003124453-2026-00013, 3005753836-2026-00004

  • Two reports describe the lift sliding down the rail or malfunctioning during maintenance by technicians. One report does not say what caused the slide.

    2 reports: 3013423626-2026-00005, 3005753836-2026-00005

  • One report describes a stairlift that would not move using either toggle switch, though it still had power. The report says the stairlift had previously stopped intermittently for no apparent reason.

    Report: 3003124453-2026-00011

Written by AI from the narratives of the 20 newest reports that have one, received Jun 10, 2026 to Oct 26, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
141in the FDA’s Global Unique Device Identification Database
Establishment listings
24registered establishments that list this product code

Product codes classified under the same regulation.

Watch PCD in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get PCD every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PCD. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.