FDA product code
PCB: External Cerebrospinal Fluid (Csf) Diversion (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 882.5560
- Review panel
- Neurology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 20 in the last 12 months
- Median review time
- 357 daysthe latest 2 decisions
- Recall records
- 50 in the last 12 months
- Adverse event reports
- 2last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code PCB per year, by decision date, 2 since 2014. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
Recalls
Recall records for product code PCB per year, by the date the recall began. Each recalled product is one record; these 5 records belong to 1 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 5 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1162-2024Class I | Medtronic NeurosurgeryPotential for catheter disconnection from the patient line stopcock connectors. | Jan 22, 2024 |
| Z-1163-2024Class I | Medtronic NeurosurgeryPotential for catheter disconnection from the patient line stopcock connectors. | Jan 22, 2024 |
| Z-1164-2024Class I | Medtronic NeurosurgeryPotential for catheter disconnection from the patient line stopcock connectors. | Jan 22, 2024 |
| Z-1165-2024Class I | Medtronic NeurosurgeryPotential for catheter disconnection from the patient line stopcock connectors. | Jan 22, 2024 |
| Z-1166-2024Class I | Medtronic NeurosurgeryPotential for catheter disconnection from the patient line stopcock connectors. | Jan 22, 2024 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code PCB. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 3 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 8 |
| 2024 | 12 |
| 2025 | 1 |
| 2026 (so far) | 2 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions2
- Other or not stated0
In plain English: what the newest reports describe Written by AI
Several reports describe the drain tubing or connection pieces coming apart, usually at a connector or stopcock, so cerebrospinal fluid (the fluid around the brain and spinal cord) leaked out. Staff clamped the line, changed the system or reattached it.
12 reports: 19475524, 19434751, 18826021, MW5151018, MW5149143, MW5149144, MW5149145, 18825982, MW5149142, 24578550, 24578552, 19031339
A few reports describe the tubing, nipple or drainage bag cracking, breaking or leaking from its base. Staff replaced the part and the fluid was caught in linens or on the floor.
5 reports: 19152246, 19087291, 18551942, 18101875, 18573796
In several reports, the drain disconnected or leaked while a patient was being moved, bathed or walked, or without anyone touching it. In one report a nurse found the connector had come apart during a bath, and in another the device fell to the floor when a patient walked to the bathroom.
3 reports: 24578550, 24578552, 19434751
Two reports describe staff who were exposed to cerebrospinal fluid or who found fluid in the lab sample. One report describes a positive fluid culture for a bacterium, and the reports do not say whether the leak caused it.
2 reports: 18573796, 19087291
One report describes the stopcock disconnecting several times, and the patient needed further medical care including scans and a spinal tap. The reporter said this followed a large loss of cerebrospinal fluid, but the reports do not say whether the loss itself caused the harm.
Report: MW5153724
Written by AI from the narratives of the 20 newest reports that have one, received Nov 9, 2023 to Mar 12, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 26in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 9registered establishments that list this product code
Watch PCB in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get PCB every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PCB. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:PCBRetrieved Nov 29, 2026, 8:35 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PCBRetrieved Nov 29, 2026, 8:35 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PCB · count decision_dateRetrieved Nov 29, 2026, 8:35 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:PCBRetrieved Nov 29, 2026, 8:35 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PCB · count event_date_initiatedRetrieved Nov 29, 2026, 8:35 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PCB · count res_event_numberRetrieved Nov 29, 2026, 8:35 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1162-2024" "Z-1163-2024" "Z-1164-2024" "Z-1165-2024" "Z-1166-2024")Retrieved Nov 29, 2026, 8:35 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PCB AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 8:35 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PCB AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 8:35 PM UTC.
- device/event · search device.device_report_product_code:PCB AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 8:35 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:PCBRetrieved Nov 29, 2026, 8:35 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:PCBRetrieved Nov 29, 2026, 8:35 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:882.5560Retrieved Nov 29, 2026, 8:35 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil