FDA product code
PAW: Nonabsorbable Expanded Polytetrafluoroethylene Surgical Suture For Chordae Tendinae Repair Or Replacement (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 870.3470
- Review panel
- Cardiovascular
- Premarket route
- 510(k)
- Implant
- Yes
- Life-sustaining or supporting
- Yes
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 31 in the last 12 months
- Median review time
- 121 daysthe latest 3 decisions
- Recall records
- 10 in the last 12 months
- Adverse event reports
- 1last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code PAW per year, by decision date, 3 since 2012. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 1 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K252126 | MitraPatch Mitral Valve Repair SystemChawla Heart Technologies, LLC | Dec 4, 2025150 days |
| K141060 | CHORD-X PRE-MEASURED LOOPS FOR MITRAL CHORDAL REPLACEMENTOn-X Life Technologies, Inc. | Jun 6, 201443 days |
| K121173 | CHORD-X EPTFE SUTUREOn-Xlifetechnologies | Aug 16, 2012121 days |
Recalls
Recall records for product code PAW per year, by the date the recall began. Each recalled product is one record; these 1 records belong to 1 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1057-2017Class II | On-X Life Technologies, Inc.Four issues were uncovered by the recalling firm's CAPA: 1) Burrs were found to be present on Left Housing, Right Housing and Barrel components, 2) Loose plast… | Dec 13, 2016 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code PAW. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 2 |
| 2025 | 4 |
| 2026 (so far) | 0 |
Last 12 months by event type
- Deaths0
- Injuries1
- Malfunctions0
- Other or not stated0
Most reported device problems, last 12 months
- Break1
- Defective Component1
In plain English: what the newest reports describe Written by AI
In the reports, the suture broke, tore or pulled out at its anchor on the papillary muscle, the muscle that holds the valve's cords. This led to valve replacement in some cases and to longer operations with extra procedures, such as additional clamping of the heart's blood supply.
5 reports: 1649833-2025-00010, 1649833-2025-00045, 1649833-2016-00035, 1649833-2016-00064, 1649833-2020-00043
A small cushion piece at the end of the suture, called a pledget, came loose and was lost inside the heart's left ventricle. Surgeons searched with a camera or by other means but could not find it, and in one case the search needed further procedures.
4 reports: 1649833-2025-00045, 1649833-2025-00076, 1649833-2023-00040, 1649833-2016-00064
Two published studies were reported. One study from Vietnam reported generally favorable results, with cardiac deaths during follow-up. The other reported a return to surgery for bleeding, a longer hospital stay and readmissions within 30 days, for heart rhythm problems or other reasons.
2 reports: 1649833-2025-00012, 1649833-2024-00072
A report describes pre-tied knots sliding sooner than the surgeon expected while tying, after which the suture pulled out of the muscle. The surgeon was not sure whether the suture itself had broken.
Report: 1649833-2016-00064
A report describes a valve repair done about two years earlier, after which a loop holding the valve's cords detached from the papillary muscle. Leaking of the valve (mitral regurgitation) followed, and the valve was replaced. The cause of the detachment is unknown.
Report: 1649833-2024-00001
Written by AI from the narratives of the 10 newest reports that have one, received Aug 24, 2016 to Dec 31, 2025. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 40in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 5registered establishments that list this product code
Other codes under 21 CFR 870.3470
Product codes classified under the same regulation.
Watch PAW in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get PAW every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in PAW. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:PAWRetrieved Nov 29, 2026, 12:51 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PAWRetrieved Nov 29, 2026, 6:33 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:PAW · count decision_dateRetrieved Nov 29, 2026, 12:51 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:PAWRetrieved Nov 29, 2026, 6:33 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PAW · count event_date_initiatedRetrieved Nov 29, 2026, 12:51 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:PAW · count res_event_numberRetrieved Nov 29, 2026, 12:51 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1057-2017")Retrieved Nov 29, 2026, 6:33 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PAW AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:33 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PAW AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:33 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:PAW AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 12:51 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:PAWRetrieved Nov 29, 2026, 12:51 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:PAWRetrieved Nov 29, 2026, 12:51 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:870.3470Retrieved Nov 28, 2026, 8:42 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.