Product codes

FDA product code

OUT: Intracranial Aneurysm Flow Diverter (FDA product code)

Follow OUT: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class III
Regulation
None listed
Review panel
Neurology
Premarket route
PMA
Implant
Yes
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
PMA decisions
933 original PMAs, the rest supplements
Median review time
30 daysthe latest 25 decisions
Recall records
143 in the last 12 months
Adverse event reports
1,682last 12 months, 41 deaths

PMA decisions

Original PMAs and supplements for product code OUT per year, by decision date, 93 since 2011. Review time is the days from receipt to decision.

All of them in the FDA’s PMA database

PMA decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20171
20187
20195
20209
202113
20227
20238
202414
202510
2026 (so far)6

Latest decisions

NumberDevice and applicantDecision and review time
P170024 S027 Surpass Elite™ Flow DiverterStryker Neurovascular Aug 28, 202639 days
P100018 S055 Pipeline™ Flex Embolization Device (Pipeline™ Flex); Pipeline™ Flex Embolization Device with Shield Technology™ (PipelinMicro Therapeutics, Inc. d/b/a ev3 Neurovascular Jul 21, 202629 days
P170024 S026 Surpass Evolve; Surpass Elite FDSStryker Neurovascular Jun 26, 202629 days
P100018 S054 Pipeline™ Flex Embolization Device with Shield Technology™ (Pipeline™ Shield); Pipeline™ Vantage Embolization Device witMicro Therapeutics, Inc. d/b/a ev3 Neurovascular Jun 26, 202629 days
P100018 S053 Pipeline™ Flex Embolization Device (Pipeline™ Flex); Pipeline™ Flex Embolization Device with Shield Technology™ (PipelinMicro Therapeutics, Inc. d/b/a ev3 Neurovascular May 15, 202628 days
P100018 S052 PIPELINE EMBOLIZATION DEVICEMicro Therapeutics, Inc. d/b/a ev3 Neurovascular Feb 20, 202692 days
P180027 S013 Flow Re-Direction Endoluminal Device (FRED®) SystemMicroVention, Inc. Nov 2, 2025178 days
P180027 S012 Flow Re-Direction Endoluminal Device (FRED) X SystemMicroVention, Inc. Sep 20, 2025159 days
P180027 S014 FRED X SystemMicroVention, Inc. Sep 18, 202590 days
P170024 S024 Surpass Streamline Flow DiverterStryker Neurovascular Aug 20, 202527 days

Recalls

Recall records for product code OUT per year, by the date the recall began. Each recalled product is one record; these 14 records belong to 10 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20202
20212
20220
20231
20240
20253
2026 (so far)3

Latest recalls

Recall and classFirm and reasonBegan
Z-3326-2026Class II Micro Therapeutics, Inc.Due to incorrect label/product dimensions. Product is supposed to be 3.00 mm in diameter, but unit is packaging was a 2.50 mm in diameter. Jun 25, 2026
Z-2773-2026Class II Micro Therapeutics, Inc.Incorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20… Jun 8, 2026
Z-2774-2026Class II Micro Therapeutics, Inc.Incorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20… Jun 8, 2026
Z-0108-2026Class II MICROVENTION INC.Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely op… Sep 5, 2025
Z-1245-2025Class I Micro Therapeutics, Inc.Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid nar… Jan 30, 2025

Adverse event reports (MAUDE)

Medical device reports received per year that name product code OUT. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017448
2018425
2019528
2020809
2021850
20221,118
20231,186
20241,326
20251,643
2026 (so far)1,526

Last 12 months by event type

  • Deaths41
  • Injuries304
  • Malfunctions1,336
  • Other or not stated1

Most reported device problems, last 12 months

  1. Activation Failure991
  2. Physical Resistance/Sticking462
  3. Material Deformation404
  4. Retraction Problem306
  5. Adverse Event Without Identified Device or Use Problem248
  6. Unintended Movement131

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
965in the FDA’s Global Unique Device Identification Database
Establishment listings
16registered establishments that list this product code

Watch OUT in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get OUT every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in OUT. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.