FDA product code
OTD: Endovascular Suturing System (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 870.3460
- Review panel
- Cardiovascular
- Premarket route
- 510(k)
- Implant
- Yes
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 70 in the last 12 months
- Median review time
- 30 daysthe latest 7 decisions
- Recall records
- 10 in the last 12 months
- Adverse event reports
- 46last 12 months, 1 death
510(k) and De Novo decisions
Decisions for product code OTD per year, by decision date, 7 since 2011. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K182957 | Heli-FX EndoAnchor SystemMedtronic Vascular, Inc. | Nov 21, 201828 days |
| K171427 | Heli-FX Applier, Heli-FX Guide, Ancillary EndoAnchor CassetteMedtronic Vascular, Inc. | Jun 13, 201729 days |
| K141041 | APTUS HELI-FX ANCILLARY ENDOANCHOR CASSETTEAptus Endosystems | May 16, 201423 days |
| K140036 | APTUS HELI-FX GUIDE - 22CM DEFLECTABLE LENGTH; APTUS HELI-FX GUIDE - 28CM DEFLECTABLE LENGTH; APTUS HELI-FX APPLIER WITHAptus Endosystems, Inc. | Feb 6, 201430 days |
| K130677 | APTUS HELI-FX AORTIC SECUREMENT SYSTEMAptus Endosystems, Inc. | Apr 12, 201330 days |
| K121168 | APTUS 18FR HELIFX GUIDE-22 DEFLECTABLE LENGTHAptus Endosystems, Inc. | Aug 8, 2012113 days |
| DEN100026 | APTUS STEERABLE ENDOGUIDE WITH OBTURATOR, APTUS ENDOSTAPLE APPLIER WITH ENDOSTAPLE CASSETTE, APTUS ENDOSTAPLE CASSETTEAptus Endosystems, Inc. | Nov 21, 2011342 days |
Recalls
Recall records for product code OTD per year, by the date the recall began. Each recalled product is one record; these 1 records belong to 1 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-3227-2017Class II | Medtronic Vascular, Inc.It was determined that the deflection length indicated on the Guide catheter handle does not match the label on the box and sterile packaging for two lots. | Sep 13, 2017 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code OTD. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 161 |
| 2018 | 215 |
| 2019 | 141 |
| 2020 | 101 |
| 2021 | 59 |
| 2022 | 52 |
| 2023 | 52 |
| 2024 | 62 |
| 2025 | 56 |
| 2026 (so far) | 39 |
Last 12 months by event type
- Deaths1
- Injuries17
- Malfunctions27
- Other or not stated1
Most reported device problems, last 12 months
- Fracture15
- Activation, Positioning or Separation Problem11
- Leak/Splash11
- Adverse Event Without Identified Device or Use Problem9
- Device Difficult to Setup or Prepare5
- Device Dislodged or Dislocated5
In plain English: what the newest reports describe Written by AI
The anchor broke, snapped, or detached during deployment or removal, and pieces were left in the graft or fell into blood vessels. In one report the anchor was retrieved with a snare, and in another part of it stayed embedded in the graft.
4 reports: 9612164-2026-04758, 9612164-2026-05135, 9612164-2026-04075, 9612164-2026-02390
Anchors were later found to have moved out of place. In one report an anchor was found in a leg artery about a year after placement, and in another a displaced anchor was found about five years later and removed. The manufacturer's cause is unknown in both cases.
2 reports: 9612164-2026-05067, 9612164-2026-03954
The anchor applier had trouble loading anchors. In one report the applier showed an error light and the anchor was fully deployed without penetrating the graft fabric. In the other, an anchor did not seat in the applier even though the forward button flashed as if loading was complete. The cause was not determined.
2 reports: 9612164-2026-03917, 9612164-2026-03347
Items were found missing from opened packaging. In one report all five anchors in an accessory cassette were gone, and in the other the dilator was missing from a guide package. The manufacturer attributed the guide case to a packaging error, and the cassette cause was undetermined.
2 reports: 9612164-2026-03117, 9612164-2026-02216
Patients had symptoms after the procedure, such as abdominal pain, fever, low potassium, or liver injury. The site and sponsor assessed these differently, with some judged possibly related to the procedure but not to the device.
2 reports: 9612164-2026-05263, 9612164-2026-04993
Written by AI from the narratives of the 18 newest reports that have one, received Apr 24, 2026 to Oct 22, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 36in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 22registered establishments that list this product code
Watch OTD in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get OTD every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in OTD. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:OTDRetrieved Nov 29, 2026, 8:11 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OTDRetrieved Nov 29, 2026, 8:11 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:OTD · count decision_dateRetrieved Nov 29, 2026, 8:11 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:OTDRetrieved Nov 29, 2026, 8:11 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:OTD · count event_date_initiatedRetrieved Nov 29, 2026, 8:11 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:OTD · count res_event_numberRetrieved Nov 29, 2026, 8:11 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-3227-2017")Retrieved Nov 29, 2026, 8:11 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OTD AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 8:11 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OTD AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 8:11 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:OTD AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 8:11 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:OTDRetrieved Nov 29, 2026, 8:11 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:OTDRetrieved Nov 29, 2026, 8:11 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:870.3460Retrieved Nov 29, 2026, 8:11 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.