Product codes

FDA product code

OMA: Amplitude-Integrated Electroencephalograph (FDA product code)

Follow OMA: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 882.1400
Review panel
Neurology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
160 in the last 12 months
Median review time
191 daysthe latest 16 decisions
Recall records
30 in the last 12 months
Adverse event reports
0last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code OMA per year, by decision date, 16 since 1979. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20171
20180
20190
20201
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K192889 NeoEemagine Medical Imaging Solutions GmbH Jan 30, 2020112 days
K163644 Nihon Kohden QP-160AK EEG Trend ProgramNihon Kohden Corporation May 19, 2017147 days
K161027 Cadwell AmpliScanCadwell Industries, Inc. Nov 8, 2016210 days
K152301 Background Pattern Classification [BPc(TM)]Natus Medical Incorporated Dba Excel-Tech Ltd. (Xltek) Jun 3, 2016294 days
K130238 NIHON KOHDEN AE-918P NEURO UNITNihon Kohden Corp. Mar 4, 2015762 days
K131789 CEREBRALOGIK- AEEGMennen Medical , Ltd. Dec 27, 2013192 days
K123079 OLYMPIC BRAINZ MONITORExcel-Tech Ltd. (Xltek) May 8, 2013219 days
K120485 NIHON KOHDEN QP-160 AK TREND PROGRAM WITH THE ADDITION OF DSA ASYMMETRY TREND AND FFT PNihon Kohden America, Inc. Mar 16, 201228 days
K092573 NIHON KOHDEN QP-160AK EEG TREND PROGRAMNihon Kohden America, Inc. Jul 9, 2010323 days
K093949 OLYMPIC BRAINZ MONITOR MODEL OBM70001Natus Medical, Inc. Jun 16, 2010175 days

Recalls

Recall records for product code OMA per year, by the date the recall began. Each recalled product is one record; these 3 records belong to 3 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20201
20210
20220
20231
20240
20251
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1925-2025Class II Natus Neurology DBA Excel Tech., Ltd. (XLTEK)Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately lead to misdiagnosis of seizure location in neonate… May 7, 2025
Z-1338-2023Class II Moberg Research, Inc.Battery leakage can cause corrosion to the metal enclosure near the electrical interface. Prolonged contact to corrosive materials could cause injury to skin.… Feb 20, 2023
Z-1630-2020Class II Natus Neurology DBA Excel Tech., Ltd. (XLTEK)The OBM system functional test (impedance and noise) was not carried out on the affected items prior to release to the customer. The manufacturer cannot attest… Mar 5, 2020

Adverse event reports (MAUDE)

Medical device reports received per year that name product code OMA. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20170
20180
20190
20200
20210
20220
20231
20242
20250
2026 (so far)0

Last 12 months by event type

No reports received in the last 12 months.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
77in the FDA’s Global Unique Device Identification Database
Establishment listings
45registered establishments that list this product code

Watch OMA in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get OMA every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in OMA. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.