Product codes

FDA product code

OAD: Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Flutter (FDA product code)

Follow OAD: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class III
Regulation
None listed
Review panel
Cardiovascular
Premarket route
PMA
Implant
No
Life-sustaining or supporting
Yes
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
PMA decisions
5599 original PMAs, the rest supplements
Median review time
30 daysthe latest 25 decisions
Recall records
221 in the last 12 months
Adverse event reports
251last 12 months, 10 deaths

PMA decisions

Original PMAs and supplements for product code OAD per year, by decision date, 559 since 2002. Review time is the days from receipt to decision.

All of them in the FDA’s PMA database

PMA decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
201764
201829
201939
202042
202150
202220
202319
202419
202519
2026 (so far)14

Latest decisions

NumberDevice and applicantDecision and review time
P030031 S146 EZ STEER THERMOCOOL NAV BI-DIRECTIONAL ELECTROPHYSIOLOGY CATHETER; THERMOCOOL SMARTTOUCH BI-DIRECTIONAL NAVIGATION CATHEBiosense Webster, Inc. Sep 15, 202628 days
P040042 S060 Therapy™ Dual-8™ Ablation CatheterIrvine Biomedical,Inc.(Ibi) Sep 3, 202627 days
P030031 S142 THERMOCOOL SMARTTOUCH® SF CatheterBiosense Webster, Inc. Sep 2, 2026180 days
P040042 S059 Ampere GeneratorIrvine Biomedical,Inc.(Ibi) Aug 7, 202688 days
P060019 S059 Ampere GeneratorIrvine Biomedical, Inc. Aug 7, 202688 days
P110016 S095 Ampere GeneratorABBOTT MEDICAL Aug 7, 202688 days
P030031 S141 Dual Energy THERMOCOOL SMARTTOUCH SF™ SystemBiosense Webster, Inc. Jul 1, 2026279 days
P030031 S143 VISITAG SURPOINT™ External Processing Unit (“EPU”); CARTO 3 SystemBiosense Webster, Inc. Jun 30, 202690 days
P030031 S144 THERMOCOOL SMARTTOUCH SF (Uni-Directional and Bi-Directional)/DUAL ENERGYTHERMOCOOL SMARTTOUCH SF CATHETERSBiosense Webster, Inc. Jun 10, 20267 days
P110016 S094 FlexAbility™ Ablation Catheter; FlexAbilityTM Ablation Catheter, Sensor EnabledTMABBOTT MEDICAL Jun 5, 202635 days

Recalls

Recall records for product code OAD per year, by the date the recall began. Each recalled product is one record; these 22 records belong to 11 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20184
20190
20200
20210
20221
20230
20240
20250
2026 (so far)1

Latest recalls

Recall and classFirm and reasonBegan
Z-1948-2026Class II Medtronic, Inc.Ventricular fibrillation (VF) occurred during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik Implantable Car… Mar 11, 2026
Z-0169-2023Class II Johnson & Johnson Surgical Vision, Inc.Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure o… Sep 7, 2022
Z-0699-2019Class III Boston Scientific CorporationSupplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions… Nov 28, 2018
Z-0700-2019Class III Boston Scientific CorporationSupplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions… Nov 28, 2018
Z-0701-2019Class III Boston Scientific CorporationSupplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions… Nov 28, 2018

Adverse event reports (MAUDE)

Medical device reports received per year that name product code OAD. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017403
2018423
2019475
2020389
2021402
2022344
2023319
2024290
2025196
2026 (so far)225

Last 12 months by event type

  • Deaths10
  • Injuries103
  • Malfunctions131
  • Other or not stated7

Most reported device problems, last 12 months

  1. Adverse Event Without Identified Device or Use Problem77
  2. Patient Device Interaction Problem38
  3. Nonstandard Device20
  4. Fluid/Blood Leak13
  5. Temperature Problem9
  6. No Flow8

In plain English: what the newest reports describe Written by AI

  • Fluid built up around the heart, called pericardial effusion or cardiac tamponade, which can squeeze the heart. Patients needed a needle drainage procedure called pericardiocentesis.

    5 reports: 3013300026-2026-03608, 3008452825-2026-00527, 3013300026-2026-03204, 3013300026-2026-03206, 3008452825-2026-00545

  • Catheters or the mapping system showed error messages or temperature problems, such as a faulty temperature sensor, high impedance (resistance to energy flow), or an inability to reach the needed power. Where stated, the catheter was replaced and the procedure continued.

    4 reports: 2124215-2026-52972, 2124215-2026-53000, MW6004280, 3013300026-2026-03206

  • Two reports describe the patient going into ventricular fibrillation, a dangerous heart rhythm, or cardiac arrest. One happened as energy was delivered and a shock was needed; the other happened in recovery and CPR was needed.

    2 reports: 3013300026-2026-03328, 3013300026-2026-03463

  • Two reports describe low blood pressure or cardiogenic shock, where the heart cannot pump enough blood, after pulsed field ablation. One cause was never found; in the other, the patient later died, and the cause and date of death were not yet provided.

    2 reports: 2124215-2026-55140, 2124215-2026-55149

  • One report describes a catheter pushed past the heart wall in the left atrium, causing a hole and tamponade. Surgeons repaired the wall with surgical adhesive, and the patient was reported to be doing well afterward.

    Report: 3013300026-2026-03261

Written by AI from the narratives of the 17 newest reports that have one, received Sep 29, 2026 to Oct 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
310in the FDA’s Global Unique Device Identification Database
Establishment listings
108registered establishments that list this product code

Watch OAD in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get OAD every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in OAD. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.