Product codes

FDA product code

NNW: Bag, Urine Collection, Leg, For External Use, Non-Sterile (FDA product code)

Follow NNW: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class I
Regulation
21 CFR 876.5250
Review panel
Gastroenterology and Urology
Premarket route
510(k) exempt
Implant
No
Life-sustaining or supporting
No
GMP exempt
Yes
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
00 in the last 12 months
Median review time
Not availablethe latest 0 decisions
Recall records
20 in the last 12 months
Adverse event reports
2last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code NNW per year, by decision date. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

No decisions on record for this product code.

Recalls

Recall records for product code NNW per year, by the date the recall began. Each recalled product is one record; these 2 records belong to 2 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1485-2017Class II Purewick CorporationPureWick is recalling external catheter for women Wicks because the labeling did not contain the phrase "This product contains dry natural rubber" and/or erron… Feb 24, 2016
Z-0155-2015Class II Coloplast Manufacturing US, LLCColoplast is recalling the Conveen Contour Leg Bag due to potential leakage from the bag seam when filled. Sep 29, 2014

Adverse event reports (MAUDE)

Medical device reports received per year that name product code NNW. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20174
20184
20190
20201
20211
20221
20232
20241
20251
2026 (so far)2

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions2
  • Other or not stated0

Most reported device problems, last 12 months

  1. Restricted Flow rate2
  2. Difficult to Open or Close1

In plain English: what the newest reports describe Written by AI

  • Several reports describe the bag not filling: urine stays in the tubing or does not enter the bag, even after the bag was checked and the tubing had no kinks. One report says the patient blew through the bag and urine still would not enter; the manufacturer's advice in that report was to keep the bag and send it to its complaints department.

    5 reports: 1018233-2026-03990, 1018233-2026-04036, 22851327, MW5105209, 1018233-2018-04384

  • Several reports describe skin injuries near the device. These include blistering, pressure ulcers, friction abrasions and a skin ulcer. In one report the wound was described as resembling suction marks, and the reporter said the device was not properly placed and the suction was set at 200 mmHg. Another report said the suction was set at 160 mmHg and described redness rather than bleeding.

    5 reports: 1018233-2017-05510, 1018233-2018-00976, 1018233-2017-02984, 1018233-2017-01706, 1018233-2018-00429

  • Several reports describe leaks or failures at the connections. One report says a bag connection came apart, which broke the closed system and forced a catheter change. In that case the patient's re-catheterisation was traumatic, with blood loss, and the patient was hospitalized before recovering.

    3 reports: 3006606901-2023-00018, MW5163621, MW5066471

  • Two reports describe a bag or tubing that kinks or collapses. One says the bag kinks when full at night, causing urine to back up and cause pain. Another reports the tubing kinked. A third says the tubing was missing a hose.

    3 reports: MW5107260, 1018233-2017-02427, 1018233-2023-00956

  • A report describes a patient whose leg straps cut into the legs, reportedly down to the bone, and sliced through the bag. The patient reportedly has no feeling in the legs and so does not notice the damage.

    Report: 1018233-2016-00897

Written by AI from the narratives of the 20 newest reports that have one, received Jul 13, 2016 to Jun 19, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
142in the FDA’s Global Unique Device Identification Database
Establishment listings
98registered establishments that list this product code

Watch NNW in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get NNW every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in NNW. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.