FDA product code
NNW: Bag, Urine Collection, Leg, For External Use, Non-Sterile (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 876.5250
- Review panel
- Gastroenterology and Urology
- Premarket route
- 510(k) exempt
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- Yes
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 00 in the last 12 months
- Median review time
- Not availablethe latest 0 decisions
- Recall records
- 20 in the last 12 months
- Adverse event reports
- 2last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code NNW per year, by decision date. Review time is the days from receipt to decision.
No decisions on record for this product code.
Recalls
Recall records for product code NNW per year, by the date the recall began. Each recalled product is one record; these 2 records belong to 2 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1485-2017Class II | Purewick CorporationPureWick is recalling external catheter for women Wicks because the labeling did not contain the phrase "This product contains dry natural rubber" and/or erron… | Feb 24, 2016 |
| Z-0155-2015Class II | Coloplast Manufacturing US, LLCColoplast is recalling the Conveen Contour Leg Bag due to potential leakage from the bag seam when filled. | Sep 29, 2014 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code NNW. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 4 |
| 2018 | 4 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 1 |
| 2022 | 1 |
| 2023 | 2 |
| 2024 | 1 |
| 2025 | 1 |
| 2026 (so far) | 2 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions2
- Other or not stated0
Most reported device problems, last 12 months
- Restricted Flow rate2
- Difficult to Open or Close1
In plain English: what the newest reports describe Written by AI
Several reports describe the bag not filling: urine stays in the tubing or does not enter the bag, even after the bag was checked and the tubing had no kinks. One report says the patient blew through the bag and urine still would not enter; the manufacturer's advice in that report was to keep the bag and send it to its complaints department.
5 reports: 1018233-2026-03990, 1018233-2026-04036, 22851327, MW5105209, 1018233-2018-04384
Several reports describe skin injuries near the device. These include blistering, pressure ulcers, friction abrasions and a skin ulcer. In one report the wound was described as resembling suction marks, and the reporter said the device was not properly placed and the suction was set at 200 mmHg. Another report said the suction was set at 160 mmHg and described redness rather than bleeding.
5 reports: 1018233-2017-05510, 1018233-2018-00976, 1018233-2017-02984, 1018233-2017-01706, 1018233-2018-00429
Several reports describe leaks or failures at the connections. One report says a bag connection came apart, which broke the closed system and forced a catheter change. In that case the patient's re-catheterisation was traumatic, with blood loss, and the patient was hospitalized before recovering.
3 reports: 3006606901-2023-00018, MW5163621, MW5066471
Two reports describe a bag or tubing that kinks or collapses. One says the bag kinks when full at night, causing urine to back up and cause pain. Another reports the tubing kinked. A third says the tubing was missing a hose.
3 reports: MW5107260, 1018233-2017-02427, 1018233-2023-00956
A report describes a patient whose leg straps cut into the legs, reportedly down to the bone, and sliced through the bag. The patient reportedly has no feeling in the legs and so does not notice the damage.
Report: 1018233-2016-00897
Written by AI from the narratives of the 20 newest reports that have one, received Jul 13, 2016 to Jun 19, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 142in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 98registered establishments that list this product code
Other codes under 21 CFR 876.5250
Product codes classified under the same regulation.
Watch NNW in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get NNW every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in NNW. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:NNWRetrieved Nov 29, 2026, 7:13 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:NNWRetrieved Nov 29, 2026, 7:13 PM UTC.
- device/510k · search product_code:NNW · count decision_dateRetrieved Nov 29, 2026, 7:13 PM UTC.
- device/recall · search product_code:NNWRetrieved Nov 29, 2026, 7:13 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:NNW · count event_date_initiatedRetrieved Nov 29, 2026, 7:13 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:NNW · count res_event_numberRetrieved Nov 29, 2026, 7:13 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1485-2017" "Z-0155-2015")Retrieved Nov 29, 2026, 7:13 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:NNW AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:13 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:NNW AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:13 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:NNW AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:13 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:NNWRetrieved Nov 29, 2026, 7:13 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:NNWRetrieved Nov 29, 2026, 7:13 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:876.5250Retrieved Nov 29, 2026, 6:39 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil