FDA product code
NLF: Oximeter, Reprocessed (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 870.2700
- Review panel
- Anesthesiology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 270 in the last 12 months
- Median review time
- 149 daysthe latest 25 decisions
- Recall records
- 30 in the last 12 months
- Adverse event reports
- 34last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code NLF per year, by decision date, 27 since 2002. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K241758 | Reprocessed RD SET Adt Pulse Oximeter Sensor (4000, 4003); Reprocessed RD SET Pdt Pulse Oximeter Sensor (4001); Reprocessed RD SET Inf Pulse Oximeter Sensor (4002)Stryker Sustainability Solutions | Oct 22, 2024124 days |
| K222019 | Reprocessed RD SET Adt Pulse Oximeter Sensor (4000), Reprocessed RD SET Pdt Pulse Oximeter Sensor (4001), Reprocessed RD SET Inf Pulse Oximeter Sensor (4002), Reprocessed RD SET Adt Pulse Oximeter Sensor (4003)Stryker Sustainability Solutions | Dec 4, 2022149 days |
| K211138 | Reprocessed Pulse Oximeter Sensor (D-25), Reprocessed Pulse Oximeter Sensor (D-25L), Reprocessed Pulse Oximeter Sensor (D-20), Reprocessed Pulse Oximeter Sensor (N-25), Reprocessed Pulse Oximeter Sensor (I-20), Reprocessed Pulse Oximeter Sensor (Max-A), Reprocessed Pulse Oximeter Sensor (Max-AL), Reprocessed Pulse Oximeter Sensor (Max-P), Reprocessed Pulse Oximeter Sensor (Max-N), Reprocessed Pulse Oximeter Sensor (Max-I)Stryker Sustainability Solutions | Mar 8, 2022326 days |
| K201699 | Medline ReNewal Reprocessed Nellcor OxiMax SpO2 SensorSurgical Instrument and Savings Inc (Dba Medline Renewal) | Jul 22, 202030 days |
| K191018 | Medline ReNewal Reprocessed Masimo LNCS Series Adult and Pediatric SpO2 Adhesive SensorsSurgical Instrument Service and Savings, Inc. | Jul 3, 201977 days |
| K111773 | RENU MEDICAL REPROCESSED OXIMAX SENSORS: ADULT, PEDIATRIC, INFANT AND NEONATERenu Medical, Inc. | Nov 16, 2011146 days |
| K110723 | M-LNCS OXIMETRY SENSORS, REPROCESSED M-LNCS OXIMETRY SENSORSMasimo Corporation | Jul 28, 2011134 days |
| K102560 | REPROCESSED PULSE OXIMETER SENSORSSterilmed, Inc. | Feb 18, 2011164 days |
| K101280 | MIDWESTERN REPROCESSING CENTER REPROCESSED PULSE OXIMETER PROBEMidwest Reprocessing Center, LLC | Oct 7, 2010154 days |
| K092368 | REPROCESSED PULSE OXIMETER SENSORSSterilmed, Inc. | Nov 3, 200990 days |
Recalls
Recall records for product code NLF per year, by the date the recall began. Each recalled product is one record; these 3 records belong to 1 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-1663-2008 | Ascent Healthcare Solutions, Inc.Sterility Compromised -- Ascent Healthcare Solutions cannot validate the sterilization process for a variety of Ascent Healthcare Solutions products. | Mar 11, 2008 |
| Z-1664-2008 | Ascent Healthcare Solutions, Inc.Sterility Compromised -- Ascent Healthcare Solutions cannot validate the sterilization process for a variety of Ascent Healthcare Solutions products. | Mar 11, 2008 |
| Z-1683-2008 | Ascent Healthcare Solutions, Inc.Sterility Compromised -- Ascent Healthcare Solutions cannot validate the sterilization process for a variety of Ascent Healthcare Solutions products. | Mar 11, 2008 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code NLF. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 4 |
| 2018 | 8 |
| 2019 | 11 |
| 2020 | 2 |
| 2021 | 1 |
| 2022 | 2 |
| 2023 | 4 |
| 2024 | 1 |
| 2025 | 3 |
| 2026 (so far) | 34 |
Last 12 months by event type
- Deaths0
- Injuries1
- Malfunctions33
- Other or not stated0
Most reported device problems, last 12 months
- Incorrect, Inadequate or Imprecise Result or Readings10
- Electrical /Electronic Property Problem5
- Low Readings5
- Material Deformation5
- False Alarm4
- Loose or Intermittent Connection4
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 99in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 55registered establishments that list this product code
Other codes under 21 CFR 870.2700
Product codes classified under the same regulation.
Watch NLF in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get NLF every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in NLF. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:NLFRetrieved Nov 28, 2026, 11:42 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:NLFRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:NLF · count decision_dateRetrieved Nov 28, 2026, 11:42 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:NLFRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:NLF · count event_date_initiatedRetrieved Nov 28, 2026, 11:42 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:NLF · count res_event_numberRetrieved Nov 28, 2026, 11:42 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-1663-2008" "Z-1664-2008" "Z-1683-2008")Retrieved Nov 29, 2026, 4:40 PM UTC.
- device/event · search device.device_report_product_code:NLF AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:NLF AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:NLF AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:NLFRetrieved Nov 28, 2026, 11:42 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:NLFRetrieved Nov 28, 2026, 11:42 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:870.2700Retrieved Nov 29, 2026, 4:40 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.