FDA product code
NKM: Mitral Valve Repair Devices (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class III
- Regulation
- None listed
- Review panel
- Cardiovascular
- Premarket route
- PMA
- Implant
- Yes
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- PMA decisions
- 1162 original PMAs, the rest supplements
- Median review time
- 29 daysthe latest 25 decisions
- Recall records
- 60 in the last 12 months
- Adverse event reports
- 3,814last 12 months, 428 deaths
PMA decisions
Original PMAs and supplements for product code NKM per year, by decision date, 116 since 2013. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 4 |
| 2018 | 3 |
| 2019 | 9 |
| 2020 | 1 |
| 2021 | 5 |
| 2022 | 12 |
| 2023 | 20 |
| 2024 | 15 |
| 2025 | 16 |
| 2026 (so far) | 11 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| P220003 S032 | PASCAL Precision Transcatheter Valve Repair SystemEdwards Lifesciences, LLC | Oct 22, 202677 days |
| P220003 S028 | PASCAL Ace with Capture Clarity™ Valve Repair System – MR Implant SystemEdwards Lifesciences, LLC | Sep 23, 2026180 days |
| P220003 S033 | PASCAL Precision Transcatheter Valve Repair SystemEdwards Lifesciences, LLC | Sep 9, 202622 days |
| P220003 S031 | PASCAL Precision implant system (20000IS and 20000ISM); PASCAL Precision guide sheath system (20000GS)Edwards Lifesciences, LLC | Jul 29, 202629 days |
| P100009 S086 | MitraClip SystemsABBOTT MEDICAL | Jun 25, 202627 days |
| P220003 S030 | PASCAL Precision Transcatheter Valve Repair SystemEdwards Lifesciences, LLC | Jun 18, 202627 days |
| P100009 S085 | MitraClip Delivery SystemsABBOTT MEDICAL | May 28, 202629 days |
| P100009 S084 | MitraClip SystemsABBOTT MEDICAL | May 14, 202630 days |
| P220003 S027 | PASCAL Precision Transcatheter Valve Repair SystemEdwards Lifesciences, LLC | Apr 21, 202628 days |
| P100009 S083 | MitraClip NTR/XTR Clip Delivery System; MitraClip G4 SystemABBOTT MEDICAL | Mar 6, 202628 days |
Recalls
Recall records for product code NKM per year, by the date the recall began. Each recalled product is one record; these 6 records belong to 4 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 3 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2479-2023Class II | Edwards Lifesciences, LLCDue to a manufacturing issue, when using valve repair systems, intended to repair insufficient mitral and/or tricuspid valves, users experienced resistance whe… | Jul 27, 2023 |
| Z-0115-2023Class II | Abbott VascularAn increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking… | Sep 8, 2022 |
| Z-0116-2023Class II | Abbott VascularAn increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking… | Sep 8, 2022 |
| Z-0117-2023Class II | Abbott VascularAn increase in clip locking malfunctions, Establish Final Arm Angle and Clip Opening While Locked, have been observed. Contributing causes may be clip locking… | Sep 8, 2022 |
| Z-1687-2019Class II | Abbott VascularReports of implantable mitral valve repair system clips unexpectedly opening and becoming nonfunctional, resulting from unintended excessive force applied duri… | May 1, 2019 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code NKM. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 1,138 |
| 2018 | 1,668 |
| 2019 | 2,278 |
| 2020 | 2,724 |
| 2021 | 2,346 |
| 2022 | 2,584 |
| 2023 | 2,475 |
| 2024 | 2,562 |
| 2025 | 2,874 |
| 2026 (so far) | 3,506 |
Last 12 months by event type
- Deaths428
- Injuries2,386
- Malfunctions1,000
- Other or not stated0
Most reported device problems, last 12 months
- Insufficient Device Problem Information1,120
- Incomplete Coaptation556
- Unintended Movement545
- Difficult to Open or Close419
- Adverse Event Without Identified Device or Use Problem396
- Positioning Failure174
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 21in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 21registered establishments that list this product code
Watch NKM in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get NKM every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in NKM. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:NKMRetrieved Nov 28, 2026, 4:23 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:NKMRetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:NKM · count decision_dateRetrieved Nov 28, 2026, 4:23 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:NKM · count pma_numberRetrieved Nov 28, 2026, 4:23 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:NKMRetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:NKM · count event_date_initiatedRetrieved Nov 28, 2026, 4:23 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:NKM · count res_event_numberRetrieved Nov 28, 2026, 4:23 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2479-2023" "Z-0115-2023" "Z-0116-2023" "Z-0117-2023" "Z-1687-2019")Retrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:NKM AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:NKM AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:NKM AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:16 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:NKMRetrieved Nov 28, 2026, 4:23 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:NKMRetrieved Nov 28, 2026, 4:23 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil