Product codes

FDA product code

NJE: Intracranial Neurovascular Stent (FDA product code)

Follow NJE: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Not stated
Regulation
None listed
Review panel
Neurology
Premarket route
Not stated
Implant
Yes
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Ineligible
510(k) and De Novo decisions
00 in the last 12 months
Median review time
Not availablethe latest 0 decisions
Recall records
70 in the last 12 months
Adverse event reports
938last 12 months, 5 deaths

510(k) and De Novo decisions

Decisions for product code NJE per year, by decision date. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

No decisions on record for this product code.

Recalls

Recall records for product code NJE per year, by the date the recall began. Each recalled product is one record; these 7 records belong to 7 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20181
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-3020-2018Class II Stryker NeurovascularUse-related coating damage on delivery catheter shafts of Wingspan delivery catheters were reported Jun 5, 2018
Z-0482-2015Class II Stryker NeurovascularUnits of GDC 360 coils, Target coil and Gateway PTA Balloon Catheter were shipped to US customers beyond the expiration date. Oct 22, 2014
Z-1808-2013Class II Stryker NeurovascularThe device is labeled incorrectly, in that, the device Model Number on the outer carton and inner pouch labels did not match, although both referenced the same… Jun 24, 2013
Z-2701-2011 Codman & Shurtleff, Inc.Mislabeled: One of the stick on stickers on the sterile pouch inside of the product box indicates the stent size to be 22mm in length, when in fact the product… Mar 21, 2011
Z-0627-2011 Boston Scientific CorporationProducts may not meet radial force specifications. Aug 25, 2010

Adverse event reports (MAUDE)

Medical device reports received per year that name product code NJE. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017285
2018422
2019272
2020193
2021382
2022395
2023579
2024727
2025826
2026 (so far)834

Last 12 months by event type

  • Deaths5
  • Injuries61
  • Malfunctions872
  • Other or not stated0

Most reported device problems, last 12 months

  1. Failure to Advance637
  2. Premature Activation446
  3. Activation Failure158
  4. Improper or Incorrect Procedure or Method26
  5. Adverse Event Without Identified Device or Use Problem22
  6. Difficult to Advance20

In plain English: what the newest reports describe Written by AI

  • The stent, a mesh tube used to hold open a blood vessel, got stuck when pushed into the microcatheter, the thin tube that carries it. In six reports, it then came apart from its delivery wire inside a connector or the microcatheter hub, and the stent and microcatheter were replaced.

    6 reports: 3008114965-2026-01399, 3008114965-2026-01349, 3008114965-2026-01346, 3008114965-2026-01360, 3008114965-2026-01357, 3008114965-2026-01380

  • In two reports, the stent got stuck in the microcatheter and could not advance. The stent was still attached to the delivery wire when it was pulled out, and in one report the physician had to increase force to remove the wire.

    2 reports: 3008114965-2026-01385, 3008114965-2026-01365

  • In two reports, the stent's distal markers, small radiopaque bands at its tip, stayed closed and would not open when release was started. The physician pulled the stent back and used a new one to finish the procedure.

    2 reports: 3008114965-2026-01369, 3008114965-2026-01301

  • In one report, after the stent was released, it moved to a different part of the artery and did not cover the treated lesion. The physician removed the microcatheter and placed a second stent at the lesion site.

    Report: 3008114965-2026-01408

  • In one report, the tip of the stent's delivery wire broke into two or more pieces while the stent was being pushed into the microcatheter. The stent was still attached when it was removed, and a new stent was used.

    Report: 3008114965-2026-01327

Written by AI from the narratives of the 13 newest reports that have one, received Oct 15, 2026 to Oct 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
179in the FDA’s Global Unique Device Identification Database
Establishment listings
25registered establishments that list this product code

Watch NJE in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get NJE every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in NJE. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.