FDA product code
MZP: Assay, Hybridization And/Or Nucleic Acid Amplification For Detection Of Hepatitis C Rna,Hepatitis C Virus (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 866.3170
- Review panel
- Microbiology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 40 in the last 12 months
- Median review time
- 151 daysthe latest 4 decisions
- Recall records
- 110 in the last 12 months
- Adverse event reports
- 6last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code MZP per year, by decision date, 4 since 2022. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 2 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K252484 | cobas HCVRoche Molecular Systems, Inc. | Nov 3, 202588 days |
| K252102 | Alinity m HCVAbbott Molecular, Inc. | Sep 25, 202584 days |
| K233352 | Aptima HCV Quant Dx AssayHologic, Inc. | Jul 24, 2024299 days |
| K221007 | cobas HCVRoche Molecular Systems, Inc. | Nov 4, 2022213 days |
Recalls
Recall records for product code MZP per year, by the date the recall began. Each recalled product is one record; these 11 records belong to 9 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 4 |
| 2022 | 1 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0407-2024Class II | Abbott Molecular, Inc.Abbott has received reports of an increase in reactive negative controls and false positive results in patient samples when using the Identified Lots of Alinit… | Oct 30, 2023 |
| Z-1518-2022Class II | Abbott Molecular, Inc.Incorrect optical calibration | Jul 19, 2022 |
| Z-0460-2022Class II | Abbott Molecular, Inc.There is a potential for abnormal (non-sigmoidal) amplification curves. | Dec 6, 2021 |
| Z-0461-2022Class II | Abbott Molecular, Inc.There is a software issue associated with the ability to properly complete a Field Service Engineer Maintenance and Diagnostics procedure. | Dec 6, 2021 |
| Z-0462-2022Class II | Abbott Molecular, Inc.There is a software defect that can cause the Amplification Detection clamp CLEAN position to be set too high. | Dec 6, 2021 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code MZP. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 5 |
| 2018 | 4 |
| 2019 | 5 |
| 2020 | 1 |
| 2021 | 1 |
| 2022 | 1 |
| 2023 | 1 |
| 2024 | 2 |
| 2025 | 12 |
| 2026 (so far) | 4 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions6
- Other or not stated0
Most reported device problems, last 12 months
- False Negative Result3
- False Positive Result3
- Incorrect, Inadequate or Imprecise Result or Readings2
- High Test Results1
- Improper or Incorrect Procedure or Method1
In plain English: what the newest reports describe Written by AI
Several reports describe the same kind of problem: a sample first tests reactive (positive) in a pooled run, then tests non-reactive when repeated, or the reverse. The pooled MPX kit combines samples from several donors before testing, and some reports say the pool result did not match the individual donor result.
6 reports: 2243471-2026-00548, 2243471-2026-00324, 2243471-2025-05316, 2243471-2025-03357, 2243471-2025-03058, 2243471-2025-03042
Some reports describe the test showing a positive result for hepatitis B or C on repeat testing with a very low signal, while antibody (serology) tests on the same patient were negative or borderline. The manufacturer's data in one case showed a sample at the lower limit of detection.
4 reports: 2243471-2025-04101, 2243471-2025-03534, 2243471-2024-03798, 2243471-2025-03350
In a few reports, samples first tested positive for hepatitis C viral load, then tested not detected on repeat, and antibody tests were negative. The results were reported as indeterminate in one case.
2 reports: 2243471-2024-03155, 2243471-2025-04110
A report described a hepatitis B result that was reactive in plasma but non-reactive in serum from the same patient, with weak, irregular amplification curves. The results were interpreted from these weak curves.
Report: 2243471-2025-03534
The manufacturer found that specific lots of the reagent had a reference dye signal outside the expected range, which caused reactive negative controls and false positive patient results. A field action was taken, and the manufacturer reported it.
Report: 3005248192-2023-00279
Written by AI from the narratives of the 19 newest reports that have one, received Dec 2, 2022 to Aug 10, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 34in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 21registered establishments that list this product code
Other codes under 21 CFR 866.3170
Product codes classified under the same regulation.
Watch MZP in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get MZP every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in MZP. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:MZPRetrieved Nov 28, 2026, 10:48 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:MZPRetrieved Nov 29, 2026, 6:02 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:MZP · count decision_dateRetrieved Nov 28, 2026, 10:48 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:MZPRetrieved Nov 29, 2026, 6:02 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MZP · count event_date_initiatedRetrieved Nov 28, 2026, 10:48 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MZP · count res_event_numberRetrieved Nov 28, 2026, 10:48 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0407-2024" "Z-1518-2022" "Z-0460-2022" "Z-0461-2022" "Z-0462-2022")Retrieved Nov 29, 2026, 6:02 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MZP AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 6:02 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MZP AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 6:02 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MZP AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 6:02 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:MZPRetrieved Nov 28, 2026, 10:48 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:MZPRetrieved Nov 28, 2026, 10:48 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:866.3170Retrieved Nov 28, 2026, 10:48 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil