FDA product code
MUE: Full Field Digital, System, X-Ray, Mammographic (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 892.1715
- Review panel
- Radiology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 564 in the last 12 months
- Median review time
- 209 daysthe latest 25 decisions
- Recall records
- 433 in the last 12 months
- Adverse event reports
- 66last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code MUE per year, by decision date, 56 since 2011. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 8 |
| 2018 | 2 |
| 2019 | 2 |
| 2020 | 4 |
| 2021 | 4 |
| 2022 | 3 |
| 2023 | 1 |
| 2024 | 1 |
| 2025 | 3 |
| 2026 (so far) | 4 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K254090 | I-View®Hologic, Inc. | Sep 2, 2026257 days |
| K253678 | MAMMOMAT B.brilliant (11372950)Siemens Medical Solutions USA, Inc. | Jun 18, 2026209 days |
| K254092 | Bellalun 2D (VDMS-1000S)Vieworks Co., Ltd. | May 28, 2026160 days |
| K260087 | Senographe Pristina; SenoBright HD; Pristina Serena; Pristina Serena BrightGE Medical Systems SCS | Mar 24, 202671 days |
| K243420 | HESTIAGenoray Co., Ltd. | Jul 17, 2025255 days |
| K243849 | 2430TCA with Xmaru WRayence Co., Ltd. | Mar 12, 202586 days |
| K241113 | VIVIX-MVieworks Co., Ltd. | Jan 16, 2025269 days |
| K233539 | MAMMOMAT B.brilliantSiemens Medical Solutions USA, Inc. | Mar 27, 2024145 days |
| K220073 | RMF-2000DRTECH Corporation | Jan 26, 2023381 days |
| K211720 | Planmed Clarity 2D, Planmed Clarify SPlanmed OY | Jul 18, 2022409 days |
Recalls
Recall records for product code MUE per year, by the date the recall began. Each recalled product is one record; these 43 records belong to 33 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 7 |
| 2018 | 1 |
| 2019 | 3 |
| 2020 | 2 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 4 |
| 2026 (so far) | 3 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0326-2027Class II | GE Medical Systems, LLCGE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisitio… | Aug 28, 2026 |
| Z-0327-2027Class II | GE Medical Systems, LLCGE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisitio… | Aug 28, 2026 |
| Z-2167-2026Class II | Hologic, IncComplaints have been received of systems developing loose, missing, or broken internal bolts over time. | Apr 8, 2026 |
| Z-0875-2026Class II | GE Medical Systems, LLCSome Senographe Pristina systems either have a missing or an incorrect X-Ray Warning label, which is a nonconformance to FDA Regulation 21 CFR 1020.30(j). | Oct 24, 2025 |
| Z-0877-2026Class II | Siemens Medical Solutions USA, IncThere were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus instal… | Oct 24, 2025 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code MUE. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 115 |
| 2018 | 144 |
| 2019 | 80 |
| 2020 | 49 |
| 2021 | 25 |
| 2022 | 36 |
| 2023 | 29 |
| 2024 | 33 |
| 2025 | 40 |
| 2026 (so far) | 60 |
Last 12 months by event type
- Deaths0
- Injuries3
- Malfunctions61
- Other or not stated2
Most reported device problems, last 12 months
- Unintended System Motion22
- Unintended Movement13
- Break3
- Installation-Related Problem3
- Mechanics Altered3
- Defective Device2
In plain English: what the newest reports describe Written by AI
Multiple reports describe the C-arm moving when no one commanded it, including continuing to rotate after the control was released, lowering on its own, or moving past a set stop. The manufacturer linked some cases to sticking switches or a sticking rotation mechanism, and blamed high room heat and humidity in one case.
6 reports: 3005496266-2026-00334, 3005496266-2026-00159, 3005496266-2026-00186, 3005496266-2026-00143, 3005496266-2026-00133, 3005496266-2026-00179
Multiple reports describe the unit shaking, vibrating, or making loud noises during movement. The problem was traced to loose bolts in the vertical travel assembly, which raises and lowers the C-arm. In one report a broken rotational gear bolt was found, and the assembly was replaced.
5 reports: 3005496266-2026-00317, 3005496266-2026-00269, 3005496266-2026-00273, 3005496266-2026-00191, 3005496266-2026-00107
One report describes the gantry moving unexpectedly during a stereotactic breast biopsy while the needle was still in the patient's breast, pulling the needle out. The team stopped the machine and removed the needle, and no added physical injury was found. A marker clip was not placed, and the manufacturer found no loose bolts on inspection.
Report: 50068059
One report describes a technologist's hand being injured when the C-arm's path was blocked by a gas tank during a patient exam, and the technologist tried to move the tank. The injury needed stitches. The service engineer found the tube head cover slightly bent, realigned it, and confirmed normal operation.
Report: 3005496266-2026-00257
One report describes the compression paddle activating unexpectedly while the technologist was away from the controls, causing a patient to cry out and suffer bruising. Compression was released right away. The service engineer replaced the rotation switch assembly, which system logs suggested might have been accidentally activated.
Report: 3005496266-2026-00093
Written by AI from the narratives of the 19 newest reports that have one, received May 1, 2026 to Oct 28, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 244in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 122registered establishments that list this product code
Watch MUE in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get MUE every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in MUE. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:MUERetrieved Nov 29, 2026, 10:52 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:MUERetrieved Nov 29, 2026, 10:52 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:MUE · count decision_dateRetrieved Nov 29, 2026, 10:52 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:MUERetrieved Nov 29, 2026, 10:52 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MUE · count event_date_initiatedRetrieved Nov 29, 2026, 10:52 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MUE · count res_event_numberRetrieved Nov 29, 2026, 10:52 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0326-2027" "Z-0327-2027" "Z-2167-2026" "Z-0875-2026" "Z-0877-2026")Retrieved Nov 29, 2026, 10:52 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MUE AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 10:52 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MUE AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 10:52 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MUE AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 10:52 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:MUERetrieved Nov 29, 2026, 10:52 AM UTC. FDA data updated Nov 2, 2026.