FDA product code
MKZ: Dna Probe, Nucleic Acid Amplification, Chlamydia (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class I
- Regulation
- 21 CFR 866.3120
- Review panel
- Microbiology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 190 in the last 12 months
- Median review time
- 170 daysthe latest 19 decisions
- Recall records
- 30 in the last 12 months
- Adverse event reports
- 13last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code MKZ per year, by decision date, 19 since 1993. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K230451 | Aptima® Chlamydia trachomatis AssayHologic, Inc. | Nov 16, 2023268 days |
| K140446 | BD PROBETEC CHLAMYDIA TRACHOMATIS (CT) QX AMPLIFIED DNA ASSAYBecton, Dickinson & CO | May 20, 201488 days |
| K110923 | COBAS 4800 CT / NG TESTRoche Molecular Systems, Inc. | Jan 24, 2012298 days |
| K091724 | BD PROBETEC CHLAMYDIA TRACHOMATIS (CT) Q AMPLIFIED DNA ASSAYBecton, Dickinson & CO | Nov 13, 2009155 days |
| K090824 | BD PROBETEC CHLAMYDIA TRACHOMATIS (CT) QX AMPLIFIED DNA ASSAYBecton, Dickinson & CO | Jun 2, 200968 days |
| K081824 | BD PROBETEC CHLAMYDIA TRACHOMATIS (CT) Q AMPLIFIED DNA ASSAYBecton, Dickinson & CO | Dec 11, 2008167 days |
| K080739 | ABBOTT REALTIME CT/NG ASSAY AND MULTI-COLLECT SPECIMEN COLLECTION KITAbbott Molecular, Inc. | Jul 10, 2008115 days |
| K070174 | AMPLICOR CT/NG TEST FOR CHLAMYDIA TRACHOMATIS; ROCHE SCRIPTS FOR CT/NG TEST ACCESSORYRoche Diagnostics Corporation | Apr 16, 200788 days |
| K063451 | GEN-PROBE APTIMA ASSAY FOR CHLAMYDIA TRACHOMATIS, MODEL 1199Gen-Probe, Inc. | Jan 22, 200768 days |
| K061413 | APTIMA CT ASSAY ON THE TIGRIS DTS SYSTEMGen-Probe, Inc. | Oct 13, 2006144 days |
Recalls
Recall records for product code MKZ per year, by the date the recall began. Each recalled product is one record; these 3 records belong to 3 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2658-2011 | Becton Dickinson & Co.In vitro diagnostic reagent packs may be underfilled and contribute to invalid quality control testing or false positive patient results. | Sep 29, 2010 |
| Z-0969-2010 | Becton Dickinson & Co.Product was shipped beyond its expiration date. | Sep 18, 2009 |
| Recall | Roche Molecular Systems IncAn increased frequency of 'blue foci' that potentially can cause elevated A450 background in microwell plate wells after the addition of conjugate reagent duri… | Jul 7, 2004 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code MKZ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 6 |
| 2018 | 3 |
| 2019 | 3 |
| 2020 | 7 |
| 2021 | 9 |
| 2022 | 2 |
| 2023 | 4 |
| 2024 | 7 |
| 2025 | 1 |
| 2026 (so far) | 13 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions13
- Other or not stated0
Most reported device problems, last 12 months
- False Positive Result13
In plain English: what the newest reports describe Written by AI
Two reports describe false positive results. When the samples were rerun on a different testing system, its negative control failed, so the reruns could not confirm the result. Both reports say there was no patient impact.
2 reports: 1119779-2026-00256, 1119779-2026-00252
Two reports describe discrepant results on repeat testing of the same specimen. In one, ferric oxide (a rust-like iron compound) was found on the extraction rack and cleaned, and results then went positive, negative, positive. In the other, results went negative, positive, negative. Both reports say there was no patient impact.
2 reports: 1119779-2024-00301, 1119779-2024-00303
Two reports describe abnormal amplification curves, the graphs used to read results. One instrument called a sample positive despite an odd curve, while the other report's curves sloped the wrong way, and its retests came back negative.
2 reports: 3004530258-2024-00018, 3004530258-2024-00016
One report describes positive control samples giving lower than expected signal across at least six instruments, and the customer suspected contamination. Cleaning with a 50% bleach solution, a strength the package insert does not list, did not fix the problem, and technical support found possible inhibition, which can cause false negatives. The customer had not amended the reported results.
Report: 2024800-2025-00012
One report describes a swab breaking during a self-collected anorectal sample, with part of the swab left in the patient's rectum. The patient reportedly went to an emergency room for removal, but the treatment received could not be confirmed. No further harm was reported.
Report: 2243471-2024-02944
Written by AI from the narratives of the 8 newest reports that have one, received Apr 18, 2024 to Feb 24, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 70in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 66registered establishments that list this product code
Other codes under 21 CFR 866.3120
Product codes classified under the same regulation.
Watch MKZ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get MKZ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in MKZ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:MKZRetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:MKZRetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:MKZ · count decision_dateRetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:MKZRetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MKZ · count event_date_initiatedRetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MKZ · count res_event_numberRetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2658-2011" "Z-0969-2010")Retrieved Nov 29, 2026, 4:58 PM UTC.
- device/event · search device.device_report_product_code:MKZ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MKZ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MKZ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:MKZRetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:MKZRetrieved Nov 29, 2026, 4:58 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:866.3120Retrieved Nov 29, 2026, 4:56 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil