FDA product code
MJP: Lens, Intraocular, Toric Optics (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class III
- Regulation
- 21 CFR 886.3600
- Review panel
- Ophthalmic
- Premarket route
- PMA
- Implant
- Yes
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- PMA decisions
- 224 original PMAs, the rest supplements
- Median review time
- 82 daysthe latest 22 decisions
- Recall records
- 40 in the last 12 months
- Adverse event reports
- 1,688last 12 months, 0 deaths
PMA decisions
Original PMAs and supplements for product code MJP per year, by decision date, 22 since 1998. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| P980040 S182 | TECNIS Eyhance™ Toric II IOLs with SmartLOAD™ Delivery Technology (GIU150-600)Johnson & Johnson Surgical Vision, Inc. | May 19, 202681 days |
| P980040 S111 | TECNIS Toric 1-Piece Intraocular Lenses (IOL)Johnson & Johnson Surgical Vision, Inc. | Jun 11, 2020134 days |
| P910056 S027 | enVista One-Piece Hydrophobic Acrylic Toric Intraocular Lens (Model MX60T)Bausch & Lomb, Inc. | Jun 8, 2018221 days |
| P980040 S079 | TECNIS Toric 1-Piece IOL Model ZCTJohnson & Johnson Surgical Vision, Inc. | Mar 24, 201729 days |
| P030002 S038 | CRYSTALENS ACCOMMODATING INTRAOCULAR LENS & TRULIGN TORIC INTRAOCULAR LENSBausch & Lomb, Inc. | Jul 27, 2016177 days |
| P030002 S035 | CRYSTALENS AND TRULIGN TORIC INTRAOCULAR LENSES (IOLS)Bausch & Lomb, Inc. | Mar 24, 2016266 days |
| P030002 S036 | TRULIGN TORIC WEB BASED TORIC CALCULATORBausch & Lomb, Inc. | Feb 17, 201664 days |
| P030002 S037 | CRYSTALENS AND TRULIGN TORIC POSTERIOR CHAMBER INTRAOCULAR LENSBausch & Lomb, Inc. | Feb 9, 201653 days |
| P980040 S058 | TECNIS TORIC 1-PIECE INTRAOCULAR LENS (IOL), TENCNIS TORIC 1-PIECE INTRAOCULAR LENS (IOL), TECNIS TORIC 1-PIECE INTRAOCUJohnson & Johnson Surgical Vision, Inc. | Jul 31, 2015179 days |
| P030002 S034 | TRULIGN TORIC POSTERIOR CHAMBER INTRAOCULAR LENSBausch & Lomb, Inc. | Jun 2, 201562 days |
Recalls
Recall records for product code MJP per year, by the date the recall began. Each recalled product is one record; these 4 records belong to 4 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 2 |
| 2022 | 0 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0735-2024Class II | Johnson & Johnson Surgical Vision, Inc.Due to intraocular lens containing an angle out of specification (OOS). The cylinder axis marks do not meet product specifications for the allowable angle betw… | Dec 12, 2023 |
| Z-0028-2022Class II | Johnson & Johnson Surgical Vision IncNonconforming product was distributed in error. Product was nonconforming due endotoxin levels higher than both the Firm's internal endotoxin specification lim… | Aug 6, 2021 |
| Z-1163-2021Class II | Johnson & Johnson Surgical Vision IncDue to the release of nonconforming Intraocular Lenses (IOLs). | Jan 27, 2021 |
| Z-0519-2021Class II | Bausch & Lomb Surgical, Inc.Product may be missing toric axis marks. | Nov 7, 2020 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code MJP. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 311 |
| 2018 | 364 |
| 2019 | 382 |
| 2020 | 211 |
| 2021 | 285 |
| 2022 | 513 |
| 2023 | 748 |
| 2024 | 1,155 |
| 2025 | 1,762 |
| 2026 (so far) | 1,532 |
Last 12 months by event type
- Deaths0
- Injuries706
- Malfunctions982
- Other or not stated0
Most reported device problems, last 12 months
- Adverse Event Without Identified Device or Use Problem565
- Break207
- Device Difficult to Setup or Prepare138
- Scratched Material133
- Failure to Deliver129
- Material Twisted/Bent103
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 3,993in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 35registered establishments that list this product code
Other codes under 21 CFR 886.3600
Product codes classified under the same regulation.
Watch MJP in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get MJP every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in MJP. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:MJPRetrieved Nov 29, 2026, 4:17 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:MJPRetrieved Nov 29, 2026, 4:17 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:MJP · count decision_dateRetrieved Nov 29, 2026, 4:17 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:MJP · count pma_numberRetrieved Nov 29, 2026, 4:17 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:MJPRetrieved Nov 29, 2026, 4:17 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MJP · count event_date_initiatedRetrieved Nov 29, 2026, 4:17 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MJP · count res_event_numberRetrieved Nov 29, 2026, 4:17 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0735-2024" "Z-0028-2022" "Z-1163-2021" "Z-0519-2021")Retrieved Nov 29, 2026, 4:17 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MJP AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:17 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MJP AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:17 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MJP AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:17 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:MJPRetrieved Nov 29, 2026, 4:17 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:MJPRetrieved Nov 29, 2026, 4:17 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:886.3600Retrieved Nov 29, 2026, 4:14 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil