MDS: Sensor, Glucose, Invasive (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class III
- Regulation
- None listed
- Review panel
- Clinical Chemistry
- Premarket route
- PMA
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- PMA decisions
- 57311 original PMAs, the rest supplements
- Median review time
- 28 daysthe latest 25 decisions
- Recall records
- 570 in the last 12 months
- Adverse event reports
- 2,215last 12 months, 1 death
PMA decisions
Original PMAs and supplements for product code MDS per year, by decision date, 573 since 1999. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 40 |
| 2018 | 79 |
| 2019 | 47 |
| 2020 | 14 |
| 2021 | 13 |
| 2022 | 8 |
| 2023 | 3 |
| 2024 | 4 |
| 2025 | 7 |
| 2026 (so far) | 5 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| P150021 S061 | Freestyle Libre Pro Flash Glucose Monitoring SystemAbbott Diabetes Care, Inc. | Sep 1, 202626 days |
| P160007 S063 | Simplera System; Simplera SyncMedtronic Minimed | Jul 21, 202629 days |
| P160007 S060 | Simplera systemMedtronic Minimed | Jun 12, 2026133 days |
| P160007 S062 | Simplera and Simplera SyncMedtronic Minimed | May 26, 202629 days |
| P160007 S058 | MiniMed Go app; Simplera sensorMedtronic Minimed | Jan 23, 202660 days |
| P160007 S059 | Simplera SystemMedtronic Minimed | Dec 8, 202512 days |
| P160007 S057 | Simplera SystemMedtronic Minimed | Dec 5, 202518 days |
| P160007 S056 | Guardian Connect System Simplera SystemMedtronic Minimed | Oct 16, 202527 days |
| P160007 S054 | Guardian Connect SystemMedtronic Minimed | Oct 9, 202555 days |
| P160007 S055 | Simplera SystemMedtronic Minimed | Sep 12, 202524 days |
Recalls
Recall records for product code MDS per year, by the date the recall began. Each recalled product is one record; these 57 records belong to 14 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 1 |
| 2019 | 2 |
| 2020 | 3 |
| 2021 | 20 |
| 2022 | 1 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 1 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0033-2026Class II | Medtronic MiniMed, Inc.The device cap label may not adhere to the sensor, compromising the sterility of the sensor, and increasing the risk of infection. | May 7, 2025 |
| Z-0776-2023Class II | Medtronic MiniMedAn app, part of a continuous glucose monitoring system, for use with smartphone devices may automatically log out from CareLink, then the app is not able to up… | Nov 21, 2022 |
| Z-1599-2021Class II | Medtronic Inc.Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose r… | Feb 26, 2021 |
| Z-1600-2021Class II | Medtronic Inc.Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose r… | Feb 26, 2021 |
| Z-1601-2021Class II | Medtronic Inc.Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose r… | Feb 26, 2021 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code MDS. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 48,848 |
| 2018 | 9,085 |
| 2019 | 2,127 |
| 2020 | 935 |
| 2021 | 590 |
| 2022 | 10,571 |
| 2023 | 13,882 |
| 2024 | 4,138 |
| 2025 | 2,835 |
| 2026 (so far) | 1,969 |
Last 12 months by event type
- Deaths1
- Injuries97
- Malfunctions2,117
- Other or not stated0
Most reported device problems, last 12 months
- Wireless Communication Problem1,649
- Data Problem242
- Incorrect, Inadequate or Imprecise Result or Readings199
- Battery Problem81
- Computer Software Problem65
- Calibration Problem49
In plain English: what the newest reports describe Written by AI
Seven reports describe the insulin pump and the transmitter losing contact, so the pump stops receiving glucose readings from the sensor. Troubleshooting was often not performed, and where it was, the outcome was unknown, unresolved, or lacked enough information to escalate.
7 reports: 2032227-2026-263519, 2032227-2026-263802, 2032227-2026-263890, 2032227-2026-262920, 2032227-2026-262386, 2032227-2026-262479, 2032227-2026-263101
Four reports describe the report feature in the companion software producing blank sections, wrong dates, or data that does not match. The manufacturer could not resolve these with standard steps and escalated them, and the device is software, so no product return applies.
4 reports: 2032227-2026-264876, 2032227-2026-263897, 2032227-2026-262884, 2032227-2026-262963
Several reports describe trouble connecting devices to a phone or app. One could not pair the transmitter with the pump, two lost the link between the mobile device and the transmitter or sensor, and one showed an app connection error alongside a pump connection loss.
4 reports: 2032227-2026-261857, 2032227-2026-262124, 2032227-2026-262436, 2032227-2026-263101
Three reports describe the sensor glucose reading differing sharply from a fingerstick blood glucose test. In one case the sensor read 350 mg/dL while blood glucose was 54 mg/dL, and in another the sensor read 52 mg/dL while blood glucose was 597 mg/dL, with the manufacturer finding the gap outside the acceptable range.
3 reports: 2032227-2026-264759, 2032227-2026-263794, 2032227-2026-264021
One report describes a sensor that would not connect and the customer experiencing bleeding. Troubleshooting was partly done for a possible site problem, and the manufacturer did not perform troubleshooting for the connection loss.
Report: 2032227-2026-261969
Written by AI from the narratives of the 19 newest reports that have one, received Oct 27, 2026 to Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 871in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 32registered establishments that list this product code
Watch MDS in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get MDS every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in MDS. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:MDSRetrieved Nov 29, 2026, 7:18 AM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:MDSRetrieved Nov 29, 2026, 7:18 AM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:MDS · count decision_dateRetrieved Nov 29, 2026, 7:18 AM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:MDS · count pma_numberRetrieved Nov 29, 2026, 7:18 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:MDSRetrieved Nov 29, 2026, 7:18 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MDS · count event_date_initiatedRetrieved Nov 29, 2026, 7:18 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MDS · count res_event_numberRetrieved Nov 29, 2026, 7:18 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0033-2026" "Z-0776-2023" "Z-1599-2021" "Z-1600-2021" "Z-1601-2021")Retrieved Nov 29, 2026, 7:18 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MDS AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:13 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MDS AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:13 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MDS AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 3:31 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:MDSRetrieved Nov 29, 2026, 7:18 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:MDSRetrieved Nov 29, 2026, 7:18 AM UTC. FDA data updated Nov 16, 2026.