MAY: Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous Cemented, Osteophilic Finish (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 888.3353
- Review panel
- Orthopedic
- Premarket route
- 510(k)
- Implant
- Yes
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 40 in the last 12 months
- Median review time
- 78 daysthe latest 4 decisions
- Recall records
- 190 in the last 12 months
- Adverse event reports
- 3last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code MAY per year, by decision date, 4 since 2005. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K123604 | ABG III MONOLITHIC HIP STEMHowmedica Osteonics Corp. | Aug 16, 2013268 days |
| K110807 | ABG II MONOLITHIC HIP STEMStryker Corp. | May 25, 201163 days |
| K090103 | 100KGY E-POLY ACETABULAR LINERS- ADDITIONAL PROFILES: +3 MAXROM AND +3 HI-WALLBiomet, Inc. | Feb 11, 200927 days |
| K051844 | EQUATOR PLUS ACETABULAR CUP SYSTEMPortland Orthopaedics Pty. , Ltd. | Oct 7, 200592 days |
Recalls
Recall records for product code MAY per year, by the date the recall began. Each recalled product is one record; these 19 records belong to 2 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0867-2016Class II | Stryker Howmedica Osteonics Corp.During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased. | Jan 21, 2016 |
| Z-0869-2016Class II | Stryker Howmedica Osteonics Corp.During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased. | Jan 21, 2016 |
| Z-0871-2016Class II | Stryker Howmedica Osteonics Corp.During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased. | Jan 21, 2016 |
| Z-0872-2016Class II | Stryker Howmedica Osteonics Corp.During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased. | Jan 21, 2016 |
| Z-0873-2016Class II | Stryker Howmedica Osteonics Corp.During a review of the packaging, it was determined that the seal integrity of the pouch may be decreased. | Jan 21, 2016 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code MAY. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 33 |
| 2018 | 33 |
| 2019 | 11 |
| 2020 | 14 |
| 2021 | 24 |
| 2022 | 11 |
| 2023 | 8 |
| 2024 | 8 |
| 2025 | 2 |
| 2026 (so far) | 2 |
Last 12 months by event type
- Deaths0
- Injuries2
- Malfunctions1
- Other or not stated0
Most reported device problems, last 12 months
- Adverse Event Without Identified Device or Use Problem1
- Patient-Device Incompatibility1
In plain English: what the newest reports describe Written by AI
Several reports describe instruments or parts breaking during surgery, such as a flexible reamer tip or shaft breaking inside the bone, a hex driver tip breaking off, or an implant component coming loose inside the patient. Where stated, the broken piece was usually found and removed, but in one report a reamer tip was left in the thigh bone and a later surgery was planned to deal with it.
7 reports: 0009610622-2024-00897, 0009610622-2023-00341, 0009610622-2023-00222, 0001825034-2023-00437, 0009610622-2024-00790, 0009610622-2024-00925, 0009610622-2024-00832
Several reports describe surgical delays or trouble removing reamers, and one report says the surgeon had to pull a broken reamer tip out with pliers. These reports describe the broken parts as a problem during the operation itself.
3 reports: 0009610622-2023-00074, 0009610622-2023-00341, 0009610622-2024-00790
One report describes a hip revision about thirteen years after an original implant, done because of instability and pain. The surgeon found wear on the liner and misalignment of the cup, and both were replaced, but the report gives no further details.
Report: 0001825034-2025-00034
One report describes a deep infection (abscess) in the left hip about two months after a hip replacement. The patient needed two rounds of surgery to clean out the infection and replace parts, and the second-stage surgery details are unknown because the manufacturer could not obtain more information.
Report: 0001825034-2026-01415
One report describes a hip replacement component that was taken out during revision surgery and fell to the floor. The surgeon then chose to open a different implant that was not in the tray, which the report says the surgeon was told was off-label use.
Report: MW5182651
Written by AI from the narratives of the 17 newest reports that have one, received Mar 2, 2023 to May 20, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 4,122in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 240registered establishments that list this product code
Other codes under 21 CFR 888.3353
Product codes classified under the same regulation.
Watch MAY in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get MAY every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in MAY. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:MAYRetrieved Nov 28, 2026, 10:58 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:MAYRetrieved Nov 29, 2026, 4:42 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:MAY · count decision_dateRetrieved Nov 28, 2026, 10:58 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:MAYRetrieved Nov 29, 2026, 4:42 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MAY · count event_date_initiatedRetrieved Nov 28, 2026, 10:58 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:MAY · count res_event_numberRetrieved Nov 28, 2026, 10:58 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0867-2016" "Z-0869-2016" "Z-0871-2016" "Z-0872-2016" "Z-0873-2016")Retrieved Nov 29, 2026, 4:42 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MAY AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:42 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MAY AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:42 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:MAY AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 4:15 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:MAYRetrieved Nov 28, 2026, 10:58 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:MAYRetrieved Nov 28, 2026, 10:58 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:888.3353Retrieved Nov 28, 2026, 6:51 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.