FDA product code
LWS: Implantable Cardioverter Defibrillator (Non-Crt) (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class III
- Regulation
- None listed
- Review panel
- Cardiovascular
- Premarket route
- PMA
- Implant
- Yes
- Life-sustaining or supporting
- Yes
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- PMA decisions
- 2,84427 original PMAs, the rest supplements
- Median review time
- 27 daysthe latest 25 decisions
- Recall records
- 2190 in the last 12 months
- Adverse event reports
- 21,561last 12 months, 177 deaths
PMA decisions
Original PMAs and supplements for product code LWS per year, by decision date, 2,844 since 1985. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 162 |
| 2018 | 142 |
| 2019 | 98 |
| 2020 | 125 |
| 2021 | 111 |
| 2022 | 91 |
| 2023 | 101 |
| 2024 | 108 |
| 2025 | 97 |
| 2026 (so far) | 99 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| P220012 S077 | Aurora EV ICD DVEA3E4Medtronic, Inc. | Nov 11, 202622 days |
| P980016 S968 | Cobalt MRI ICD DDPB3D1, DDPB3D4, DVPB3D1, DVPB3D4 Cobalt XT MRI ICD DDPA2D1, DDPA2D4, DVPA2D1, DVPA2D4 Crome MRI ICD DDPMedtronic Cardiac Rhythm Disease Management | Nov 11, 202622 days |
| P910073 S195 | ENDOTAK® RELIANCE™ Lead Models: 0137, 0138, 0157, 0158 RELIANCE 4-FRONT™ Lead Models: 0636, 0662, 0663, 0665, 0672, 0673Boston Scientific | Nov 11, 202614 days |
| P950022 S153 | UltiSynq™ Defibrillation lead; SurGrip™ Suture Sleeve AccessoryABBOTT MEDICAL | Nov 5, 2026196 days |
| P980016 S964 | Cobalt MRI ICD (DDPB3D1, DDPB3D4, DVPB3D1, DVPB3D4); Cobalt XT MRI ICD (DDPA2D1, DDPA2D4, DVPA2D1, DVPA2D4); Crome MRI IMedtronic Cardiac Rhythm Disease Management | Nov 4, 202690 days |
| P980016 S967 | Cobalt MRI ICD, Cobalt XT MRI ICD, Crome MRI ICDMedtronic Cardiac Rhythm Disease Management | Oct 29, 202627 days |
| P980023 S153 | Ilivia Neo 7 VRT DX, Intica Neo 7 VRT DX, Intica Neo 5 VRT DX, Acticor Sky VR-T, Rivacor Sky VR-T, Rivacor Rise VR-T, RiBiotronik, Inc. | Oct 22, 202630 days |
| P980023 S152 | Ilivia Neo 7 VR-T DX DF-1 ProMRI, Ilivia Neo 7 VR-T DF-1 ProMRI, Intica Neo 7 VR-T DX DF-1 ProMRI, Intica Neo 7 VR-T DF-Biotronik, Inc. | Oct 13, 202625 days |
| P980023 S149 | Acticor Sky VR-T 481989; Rivacor Sky VR-T 481984; Rivacor Rise VR-T 482020; Rivacor Aura VR-T 482015; Acticor Sky VR-T DBiotronik, Inc. | Oct 9, 202688 days |
| P980016 S966 | Cobalt MRI ICD (DDPB3Dl, DDPB3D4, DVPB3Dl, DVPB3D4); Cobalt XT MRI ICD (DDPA2Dl, DDPA2D4, DVPA2Dl, DVPA2D4); Crome MRI IMedtronic Cardiac Rhythm Disease Management | Oct 5, 202624 days |
Recalls
Recall records for product code LWS per year, by the date the recall began. Each recalled product is one record; these 219 records belong to 73 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 11 |
| 2018 | 19 |
| 2019 | 8 |
| 2020 | 8 |
| 2021 | 5 |
| 2022 | 7 |
| 2023 | 66 |
| 2024 | 17 |
| 2025 | 3 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0726-2026Class II | Medtronic, Inc.There is a potential for delayed time to high-voltage (HV) therapy should a rare sequence of events occur. | Oct 29, 2025 |
| Z-0093-2026Class I | Boston Scientific CorporationSoftware to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance… | Aug 20, 2025 |
| Z-2325-2025Class I | Boston Scientific CorporationAffected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead t… | Jul 24, 2025 |
| Z-2196-2024Class II | Medtronic Inc.Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing. | Apr 16, 2024 |
| Z-2197-2024Class II | Medtronic Inc.Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing. | Apr 16, 2024 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code LWS. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 18,454 |
| 2018 | 17,228 |
| 2019 | 16,183 |
| 2020 | 16,908 |
| 2021 | 19,654 |
| 2022 | 23,453 |
| 2023 | 28,702 |
| 2024 | 25,203 |
| 2025 | 24,629 |
| 2026 (so far) | 19,595 |
Last 12 months by event type
- Deaths177
- Injuries11,489
- Malfunctions9,830
- Other or not stated65
Most reported device problems, last 12 months
- Over-Sensing6,674
- Inappropriate/Inadequate Shock/Stimulation4,531
- Signal Artifact/Noise3,590
- High impedance3,189
- Adverse Event Without Identified Device or Use Problem2,959
- Failure to Read Input Signal1,746
In plain English: what the newest reports describe Written by AI
Reports on S-ICD systems describe the device sensing electrical noise, including muscle activity, and delivering inappropriate shocks. Technical staff reproduced the noise in clinic, and reprogramming, such as changing the sensing vector or gain, was used to reduce the false sensing.
4 reports: 2124215-2026-57737, 2124215-2026-57825, 2124215-2026-57827, 2124215-2026-57832
Reports on right ventricular leads describe noise being sensed as fast heart rhythms, with episodes misclassified and high sensing counters. Reprogramming was used, and the leads remain in use. One report links the noise to electromagnetic interference, a form of outside electrical signal.
3 reports: 2649622-2026-32315, 2649622-2026-32316, 9614453-2026-04431
Reports describe infection at the lead sleeve, in some cases at the xiphoid process (the tip of the breastbone), where the area was suppurating, meaning leaking pus. Medication was given, and the affected devices were explanted, meaning removed from the body.
3 reports: 2124215-2026-57835, 2124215-2026-57836, 2124215-2026-57837
Reports on right ventricular leads describe a fracture, low pacing impedance (electrical resistance too low for pacing), and high defibrillation impedance. The fractured and low-impedance leads were capped and replaced, while the high-impedance lead remains in use.
3 reports: 2649622-2026-32318, 2649622-2026-32311, 2649622-2026-32384
One report describes a right ventricular lead helix, the small screw-like tip that anchors the lead, that would not extend during implant. Once removed, the helix was bent and popped out quickly, and another lead was implanted.
Report: 2649622-2026-32322
Written by AI from the narratives of the 19 newest reports that have one, received Oct 30, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 444in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 137registered establishments that list this product code
Watch LWS in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get LWS every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LWS. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:LWSRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:LWSRetrieved Nov 29, 2026, 4:12 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:LWS · count decision_dateRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 16, 2026.
- device/pma · search product_code:LWS · count pma_numberRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:LWSRetrieved Nov 29, 2026, 4:12 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LWS · count event_date_initiatedRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LWS · count res_event_numberRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0726-2026" "Z-0093-2026" "Z-2325-2025" "Z-2196-2024" "Z-2197-2024")Retrieved Nov 29, 2026, 4:12 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LWS AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 4:12 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LWS AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 4:12 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LWS AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 3:18 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:LWSRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:LWSRetrieved Nov 28, 2026, 11:26 PM UTC. FDA data updated Nov 16, 2026.