Product codes

FDA product code

LPB: Cardiac Ablation Percutaneous Catheter (FDA product code)

Follow LPB: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class III
Regulation
None listed
Review panel
Cardiovascular
Premarket route
PMA
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
PMA decisions
90718 original PMAs, the rest supplements
Median review time
33 daysthe latest 25 decisions
Recall records
210 in the last 12 months
Adverse event reports
1,003last 12 months, 32 deaths

PMA decisions

Original PMAs and supplements for product code LPB per year, by decision date, 907 since 1994. Review time is the days from receipt to decision.

All of them in the FDA’s PMA database

PMA decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
201746
201835
201930
202033
202139
202225
202322
202418
202515
2026 (so far)8

Latest decisions

NumberDevice and applicantDecision and review time
P040036 S101 THERMOCOOL SMARTTOUCH BI-DIRECTIONAL NAVIGATION CATHETER; THERMOCOOL SMARTTOUCH UNI-DIRECTIONAL NAVIGATION CATHETERBiosense Webster, Inc. Sep 15, 202628 days
P040014 S052 Therapy™ (including Bi-Directional and 4mm Thermistor) Ablation CatheterIrvine Biomedical, Inc. Sep 3, 202627 days
P960016 S094 Ampere GeneratorSt Jude Medical Aug 7, 202688 days
P040014 S051 Ampere GeneratorIrvine Biomedical, Inc. Aug 7, 202688 days
P240014 S001 MAGiC Ablation CatheterStereotaxis, Inc. Jul 31, 2026177 days
P020045 S113 Freezor 207Fl,207F3,207F5, Freezor Xtra 217Fl,217F3,217F5Medtronic Cryocath, LP Apr 23, 202685 days
P960016 S093 Livewire™ TC XLS™St Jude Medical Jan 15, 202629 days
P240014 MAGiC Ablation CatheterStereotaxis, Inc. Jan 5, 2026633 days
P990071 S068 nGEN™ PumpBiosense Webster, Inc. Dec 23, 202584 days
P920047 S138 RHYTHMIA HDx™ Ablation Connection Box (for INTELLAGEN™) (M004RA65010)Boston Scientific Corp Oct 17, 202528 days

Recalls

Recall records for product code LPB per year, by the date the recall began. Each recalled product is one record; these 21 records belong to 14 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20176
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0054-2018Class II Biosense Webster, Inc.Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of TH… Sep 6, 2017
Z-0055-2018Class II Biosense Webster, Inc.Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of TH… Sep 6, 2017
Z-0056-2018Class II Biosense Webster, Inc.Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of TH… Sep 6, 2017
Z-0057-2018Class II Biosense Webster, Inc.Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of TH… Sep 6, 2017
Z-0058-2018Class II Biosense Webster, Inc.Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of TH… Sep 6, 2017

Adverse event reports (MAUDE)

Medical device reports received per year that name product code LPB. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20171,115
2018975
20191,124
20201,182
20211,240
20221,336
20231,500
20241,839
20251,199
2026 (so far)884

Last 12 months by event type

  • Deaths32
  • Injuries425
  • Malfunctions546
  • Other or not stated0

Most reported device problems, last 12 months

  1. Patient Device Interaction Problem394
  2. Incorrect, Inadequate or Imprecise Result or Readings103
  3. Obstruction of Flow100
  4. Failure to Sense87
  5. Contamination /Decontamination Problem81
  6. Adverse Event Without Identified Device or Use Problem74

In plain English: what the newest reports describe Written by AI

  • The reports describe fluid building up around the heart (pericardial effusion), which can squeeze the heart (cardiac tamponade). Each needed fluid drained with a needle (pericardiocentesis). One report also describes cardiac arrest. Causes given differ: one physician linked it to the catheter shifting with breathing, and another was unsure of the cause.

    4 reports: 3013300026-2026-03572, 3013300026-2026-03595, 3013300026-2026-03551, 3013300026-2026-03566

  • Several catheters would not connect to the mapping system or showed no temperature or impedance (electrical resistance) reading, so ablation could not start. The reports do not state a cause.

    4 reports: 51202595, 51202636, 51202676, 51202726

  • The ECG heart-rhythm signal was lost on all channels in one report, and a leakage current error stopped ablation in another. The first report says the port was wet and replacing the cable and catheter fixed it. The second says a full reboot of the system and a new cable resolved it.

    2 reports: 3013300026-2026-03556, 3013300026-2026-03574

  • In one report, the heart's electrical signal from the upper chambers to the lower chambers stopped during ablation of the slow pathway. The patient was supported with a temporary pacemaker and was still in heart block when last reported.

    Report: 3013300026-2026-03486

  • In one report, a hole formed between the upper left chamber of the heart and the esophagus (atrioesophageal fistula) weeks after an ablation. The patient died. The manufacturer's analysis of a possible over-ablation reading was unclear, and the report says there was no malfunction.

    Report: 3013300026-2026-03548

Written by AI from the narratives of the 17 newest reports that have one, received Oct 22, 2026 to Oct 28, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
530in the FDA’s Global Unique Device Identification Database
Establishment listings
132registered establishments that list this product code

Watch LPB in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get LPB every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LPB. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.