FDA product code
LKM: Counter, Urine Particle (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 864.5200
- Review panel
- Hematology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 150 in the last 12 months
- Median review time
- 64 daysthe latest 15 decisions
- Recall records
- 60 in the last 12 months
- Adverse event reports
- 1last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code LKM per year, by decision date, 15 since 1982. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 1 |
| 2025 | 1 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K252580 | iQ200 SeriesBeckman Coulter, Inc. | Sep 10, 202526 days |
| K232416 | AUTION EYE AI-4510 Urine Particle Analysis SystemArkray, Inc. | May 3, 2024266 days |
| K210127 | iQ200 System, iChemVELOCITY Automated Urine Chemistry SystemBeckman Coulter, Inc. | Jul 2, 2021164 days |
| K200811 | cobas u 701 microscopy analyzerRoche Diagnostics | Nov 6, 2020224 days |
| K171883 | Sysmex UF-5000 Fully Automated Urine Particle AnalyzerSysmex America, Inc. | Apr 23, 2018304 days |
| K093861 | IQ 20 SERIES ANALYZER, IQ 200 SERIES SYSTEM, NAD IQ LAMINA CRADLEIris International, Inc. | Feb 5, 201051 days |
| K083002 | SYSMEX UF-500I AUTOMATED URINE PARTICLE ANALYZERSysmex America, Inc. | Feb 3, 2009118 days |
| K080887 | SYSMEX UF-1000I, AUTOMATED URINE PARTICLE ANALYZER WITH URINALYSIS WAM SOFTWARESysmex America, Inc. | May 2, 200832 days |
| K070910 | SYSMEX MODEL UF1000ISysmex America, Inc. | May 25, 200753 days |
| K053135 | SYSMEX UF-100I AUTOMATED URINE CELL ANALYZERSysmex America, Inc. | Jan 4, 200657 days |
Recalls
Recall records for product code LKM per year, by the date the recall began. Each recalled product is one record; these 6 records belong to 3 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 4 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0784-2024Class II | Beckman Coulter, Inc.Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula ha… | Dec 19, 2023 |
| Z-0785-2024Class II | Beckman Coulter, Inc.Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula ha… | Dec 19, 2023 |
| Z-0786-2024Class II | Beckman Coulter, Inc.Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula ha… | Dec 19, 2023 |
| Z-0787-2024Class II | Beckman Coulter, Inc.Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula ha… | Dec 19, 2023 |
| Z-0914-2020Class II | Beckman Coulter Inc.Customers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record… | Oct 28, 2019 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code LKM. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 2 |
| 2020 | 2 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 1 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions1
- Other or not stated0
Most reported device problems, last 12 months
- Non Reproducible Results1
In plain English: what the newest reports describe Written by AI
Several reports describe the analyzer giving false negative results for bacteria or white blood cells in urine samples, even when bacteria were visible in images or under a microscope. In one case the results were not checked by microscope and an infection was missed, and in another a urine culture was delayed.
3 reports: 1823260-2026-00081, 2023446-2016-00381, 1000515253-2020-00022
In one report, a burning odor, sparks, and smoke came from the sample transport unit after urine spilled onto its conveyor, and a fire extinguisher was used. In another report, a burning odor came from the analyzer, and a technician found the wiring cord and nearby tubing burnt. No harm to users or patients was reported in either case.
2 reports: 1000515253-2019-00014, 1000515253-2019-00007
One report describes intermittent false negative results for red blood cells, casts, and crystals in urine samples from multiple patients. The manufacturer's investigation is pending in the newest report, and erroneous results were released.
Report: 1000515253-2021-00013
In one report, a cerebrospinal fluid sample gave falsely high counts of one white cell type and falsely low counts of another, with no error flags. The patient was treated with antibiotics and antiviral medicine before a second sample, checked by microscope, gave different results and steroid treatment was started.
Report: 1000515253-2020-00003
In one report, a reagent used for the analyzer's weekly maintenance was accidentally replaced with deionized water on all analyzers, because the rinse water bottle looks similar to the reagent. This led to falsely low white cell and bacteria results for a sample.
Report: 1000515253-2020-00022
Written by AI from the narratives of the 14 newest reports that have one, received Jan 30, 2013 to Jan 7, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 9in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 14registered establishments that list this product code
Other codes under 21 CFR 864.5200
Product codes classified under the same regulation.
Watch LKM in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get LKM every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LKM. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:LKMRetrieved Nov 29, 2026, 7:00 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:LKMRetrieved Nov 29, 2026, 7:00 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:LKM · count decision_dateRetrieved Nov 29, 2026, 7:00 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:LKMRetrieved Nov 29, 2026, 7:00 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LKM · count event_date_initiatedRetrieved Nov 29, 2026, 7:00 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LKM · count res_event_numberRetrieved Nov 29, 2026, 7:00 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0784-2024" "Z-0785-2024" "Z-0786-2024" "Z-0787-2024" "Z-0914-2020")Retrieved Nov 29, 2026, 7:00 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LKM AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:00 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LKM AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:00 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LKM AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:00 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:LKMRetrieved Nov 29, 2026, 7:00 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:LKMRetrieved Nov 29, 2026, 7:00 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:864.5200Retrieved Nov 29, 2026, 3:54 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil