Product codes

FDA product code

LKM: Counter, Urine Particle (FDA product code)

Follow LKM: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 864.5200
Review panel
Hematology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
150 in the last 12 months
Median review time
64 daysthe latest 15 decisions
Recall records
60 in the last 12 months
Adverse event reports
1last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code LKM per year, by decision date, 15 since 1982. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20181
20190
20201
20211
20220
20230
20241
20251
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K252580 iQ200 SeriesBeckman Coulter, Inc. Sep 10, 202526 days
K232416 AUTION EYE AI-4510 Urine Particle Analysis SystemArkray, Inc. May 3, 2024266 days
K210127 iQ200 System, iChemVELOCITY Automated Urine Chemistry SystemBeckman Coulter, Inc. Jul 2, 2021164 days
K200811 cobas u 701 microscopy analyzerRoche Diagnostics Nov 6, 2020224 days
K171883 Sysmex UF-5000 Fully Automated Urine Particle AnalyzerSysmex America, Inc. Apr 23, 2018304 days
K093861 IQ 20 SERIES ANALYZER, IQ 200 SERIES SYSTEM, NAD IQ LAMINA CRADLEIris International, Inc. Feb 5, 201051 days
K083002 SYSMEX UF-500I AUTOMATED URINE PARTICLE ANALYZERSysmex America, Inc. Feb 3, 2009118 days
K080887 SYSMEX UF-1000I, AUTOMATED URINE PARTICLE ANALYZER WITH URINALYSIS WAM SOFTWARESysmex America, Inc. May 2, 200832 days
K070910 SYSMEX MODEL UF1000ISysmex America, Inc. May 25, 200753 days
K053135 SYSMEX UF-100I AUTOMATED URINE CELL ANALYZERSysmex America, Inc. Jan 4, 200657 days

Recalls

Recall records for product code LKM per year, by the date the recall began. Each recalled product is one record; these 6 records belong to 3 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20191
20200
20210
20220
20234
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0784-2024Class II Beckman Coulter, Inc.Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula ha… Dec 19, 2023
Z-0785-2024Class II Beckman Coulter, Inc.Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula ha… Dec 19, 2023
Z-0786-2024Class II Beckman Coulter, Inc.Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula ha… Dec 19, 2023
Z-0787-2024Class II Beckman Coulter, Inc.Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula ha… Dec 19, 2023
Z-0914-2020Class II Beckman Coulter Inc.Customers have reported incidents where two sets of results report the same Specimen Identifier (Specimen ID) with different results, different Medical Record… Oct 28, 2019

Adverse event reports (MAUDE)

Medical device reports received per year that name product code LKM. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20171
20180
20192
20202
20211
20220
20230
20240
20250
2026 (so far)1

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions1
  • Other or not stated0

Most reported device problems, last 12 months

  1. Non Reproducible Results1

In plain English: what the newest reports describe Written by AI

  • Several reports describe the analyzer giving false negative results for bacteria or white blood cells in urine samples, even when bacteria were visible in images or under a microscope. In one case the results were not checked by microscope and an infection was missed, and in another a urine culture was delayed.

    3 reports: 1823260-2026-00081, 2023446-2016-00381, 1000515253-2020-00022

  • In one report, a burning odor, sparks, and smoke came from the sample transport unit after urine spilled onto its conveyor, and a fire extinguisher was used. In another report, a burning odor came from the analyzer, and a technician found the wiring cord and nearby tubing burnt. No harm to users or patients was reported in either case.

    2 reports: 1000515253-2019-00014, 1000515253-2019-00007

  • One report describes intermittent false negative results for red blood cells, casts, and crystals in urine samples from multiple patients. The manufacturer's investigation is pending in the newest report, and erroneous results were released.

    Report: 1000515253-2021-00013

  • In one report, a cerebrospinal fluid sample gave falsely high counts of one white cell type and falsely low counts of another, with no error flags. The patient was treated with antibiotics and antiviral medicine before a second sample, checked by microscope, gave different results and steroid treatment was started.

    Report: 1000515253-2020-00003

  • In one report, a reagent used for the analyzer's weekly maintenance was accidentally replaced with deionized water on all analyzers, because the rinse water bottle looks similar to the reagent. This led to falsely low white cell and bacteria results for a sample.

    Report: 1000515253-2020-00022

Written by AI from the narratives of the 14 newest reports that have one, received Jan 30, 2013 to Jan 7, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
9in the FDA’s Global Unique Device Identification Database
Establishment listings
14registered establishments that list this product code

Product codes classified under the same regulation.

Watch LKM in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get LKM every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LKM. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.