FDA product code
LJE: Catheter, Nephrostomy (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Unclassified
- Regulation
- None listed
- Review panel
- Gastroenterology and Urology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 560 in the last 12 months
- Median review time
- 184 daysthe latest 25 decisions
- Recall records
- 120 in the last 12 months
- Adverse event reports
- 49last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code LJE per year, by decision date, 56 since 1982. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 3 |
| 2019 | 4 |
| 2020 | 2 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 1 |
| 2024 | 0 |
| 2025 | 1 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K250448 | Disposable Percutaneous Nephrostomy Dilatation KitShenzhen Trious Medical Technology Co., Ltd. | Jul 3, 2025135 days |
| K222705 | Introducer NeedleYoucare Technology Co.,Ltd. (Wuhan) | May 22, 2023257 days |
| K211911 | Kolibri Percutaneous Nephrostomy Catheters, Kolibri Percutaneous Nephrostomy Kits, Kolibri Direct Puncture SetsColoplast Corp. | Mar 10, 2022262 days |
| K201165 | In-Ka Percutaneous Balloon Dilatation Catheters, Amplatz SheathColoplast Corp. | Dec 17, 2020230 days |
| K191498 | Cook-Cope Loop Nephrostomy Set, Ultrathane Nephrostomy Set with Mac-Loc, Ultrathane Suprapubic Set with Mac-LocCook Incorporated | Jan 29, 2020238 days |
| K190903 | Lawson Retrograde Nephrostomy Wire Puncture SetCook Incorporated | Dec 16, 2019252 days |
| K183051 | Kaye Nephrostomy Tamponade Balloon Catheter and Stent SetCook Incorporated | Jul 16, 2019256 days |
| K183035 | Percutaneous Entry Set, Skinny Needle(R) with Chiba Tip, Disposable Two-part Trocar NeedleCook Incorporated | May 17, 2019197 days |
| K181735 | Percutaneous Neonatal Pigtail Nephrostomy Set, Pediatric Nephrostomy Stent SetCook Incorporated | Mar 27, 2019268 days |
| K181713 | Ultraxx Nephrostomy Balloon Catheter SetCook Incorporated | Aug 17, 201850 days |
Recalls
Recall records for product code LJE per year, by the date the recall began. Each recalled product is one record; these 12 records belong to 9 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 1 |
| 2020 | 2 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 2 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0572-2026Class II | Olympus Corporation of the AmericasDevice kits contained an incorrectly sized 4mm x 4cm urinary tract balloon catheter (BURS0404) instead of the correct 8mm x 15cm nephrostomy tract balloon cath… | Oct 17, 2025 |
| Z-2535-2025Class II | Boston Scientific CorporationThe Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal in… | Aug 5, 2025 |
| Z-0414-2022Class II | Cook Inc.Specific lots may contain black debris on the exterior of the inner stylet and/or on the interior of the outer cannula. Local inflammatory reactions may occur… | Oct 13, 2021 |
| Z-0105-2021Class II | Cook Inc.The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. The current… | Aug 31, 2020 |
| Z-0106-2021Class II | Cook Inc.The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. The current… | Aug 31, 2020 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code LJE. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 60 |
| 2018 | 69 |
| 2019 | 78 |
| 2020 | 50 |
| 2021 | 38 |
| 2022 | 42 |
| 2023 | 131 |
| 2024 | 47 |
| 2025 | 61 |
| 2026 (so far) | 45 |
Last 12 months by event type
- Deaths0
- Injuries5
- Malfunctions44
- Other or not stated0
Most reported device problems, last 12 months
- Loss of or Failure to Bond6
- Material Puncture/Hole5
- Nonstandard Device5
- Dull, Blunt4
- Burst Container or Vessel3
- Device Misassembled During Manufacturing /Shipping3
In plain English: what the newest reports describe Written by AI
Balloon dilation catheters used in kidney procedures were reported to leak, rupture or fail to expand, sometimes found before use and sometimes during the procedure. In these cases a replacement was used and the procedure was completed.
6 reports: 1018233-2026-06110, 1018233-2026-04599, 1018233-2026-03549, 1018233-2026-02810, 2124215-2026-21373, 1820334-2026-00614
Several reports describe the plastic stylet (the inner wire guide that stiffens the catheter) with a connector or plastic part coming loose. In one report a surgeon had to pull the stainless steel part out with forceps, and in another the stylet was hard to remove and had to be pulled quite hard.
3 reports: 9610711-2026-00261, 9610711-2026-00246, 9610711-2026-00194
A ureteral stent (a thin tube left in the ureter to keep urine flowing) was reported to be kinked or flattened, so a guidewire could not pass through it. In one case the procedure was cancelled, and in another the stent shaft was reported to have broken.
2 reports: 2124215-2026-20064, 2124215-2026-33287
Pieces of kidney tubes were reported to break off inside patients. One tube snapped at the tip and was left in the kidney, and in another a catheter broke at the hub, the connection point.
2 reports: 50115906, 25812152
One report describes a catheter that was difficult to remove, which the reporting physician thinks may have caused a tear in the patient's bladder. The patient was admitted to hospital, and the final outcome is unknown. Another report describes a catheter that came out on its own, and the patient was later hospitalized with an infection.
2 reports: 1820334-2026-01110, MW6001345
Written by AI from the narratives of the 20 newest reports that have one, received Apr 6, 2026 to Oct 23, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 382in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 118registered establishments that list this product code
Watch LJE in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get LJE every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LJE. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:LJERetrieved Nov 28, 2026, 10:24 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:LJERetrieved Nov 29, 2026, 7:34 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:LJE · count decision_dateRetrieved Nov 28, 2026, 10:24 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:LJERetrieved Nov 29, 2026, 7:34 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LJE · count event_date_initiatedRetrieved Nov 28, 2026, 10:24 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LJE · count res_event_numberRetrieved Nov 28, 2026, 10:24 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0572-2026" "Z-2535-2025" "Z-0414-2022" "Z-0105-2021" "Z-0106-2021")Retrieved Nov 29, 2026, 7:34 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LJE AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:34 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LJE AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:34 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LJE AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:34 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:LJERetrieved Nov 28, 2026, 10:24 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:LJERetrieved Nov 28, 2026, 10:24 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil