FDA product code
LIP: Enzyme Linked Immunoabsorption Assay, Treponema Pallidum (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 866.3830
- Review panel
- Microbiology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 350 in the last 12 months
- Median review time
- 139 daysthe latest 25 decisions
- Recall records
- 182 in the last 12 months
- Adverse event reports
- 104last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code LIP per year, by decision date, 35 since 1982. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 1 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 2 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K241427 | Access SyphilisBeckman Coulter, Inc. | Sep 6, 2024109 days |
| K241534 | VITROS Immunodiagnostic Products Syphilis Reagent PackOrtho Clinical Diagnostics | Aug 28, 202490 days |
| K211302 | Elecsys SyphilisRoche Diagnostics | Jul 20, 202182 days |
| K170413 | BioPlex 2200 Syphilis Total & RPR, BioPlex 2200 Syphilis Total & RPR Calibrator Set, BioPlex 2200 Syphilis Total & RPR Control SetBio-Rad Laboratories | May 11, 201790 days |
| K160910 | Elecsys SyphilisRoche Diagnostics | Jul 28, 2016118 days |
| K153145 | Lumipulse G TP-N Immunoreaction Cartridge SetFujirebio Diagnostics,Inc. | Jul 5, 2016249 days |
| K153730 | ARCHITECT Syphilis TP Reagent, ARCHITECT Syphilis TP Calibrator, ARCHITECT Syphilis TP ControlAbbott Laboratories | Jun 15, 2016170 days |
| K120439 | BIOPLEX 2200 EBV IGG AND SYPHILIS IGGBio-Rad Laboratories, Inc. | Mar 14, 201230 days |
| K112343 | TREPONEMA PALLIDUM TREPONEMAL TEST REAGENTS; QUALITY CONTROL MATERIALSSiemens Healthcare Diagnostics, Inc. | Jan 20, 2012158 days |
| K102400 | SYPHILIS HEALTH CHECKDiagnostics Direct, LLC | Aug 1, 2011342 days |
Recalls
Recall records for product code LIP per year, by the date the recall began. Each recalled product is one record; these 18 records belong to 15 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 1 |
| 2021 | 3 |
| 2022 | 3 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 2 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-2965-2026Class II | Siemens Healthcare Diagnostics Products Ltd.Syphilis Screen assay may contain a negative patient bias resulting in low reactive patient results being reported as nonreactive. | Jul 13, 2026 |
| Z-2966-2026Class II | Siemens Healthcare Diagnostics Products Ltd.Syphilis Screen assay may contain a negative patient bias resulting in low reactive patient results being reported as nonreactive. | Jul 13, 2026 |
| Z-1096-2022Class II | Bio-Rad Laboratories, Inc.Performance concerns for the CMV IgM analyte. Analyte has demonstrated variable elevated background depending on the handling of the reagent pack. | Apr 14, 2022 |
| Z-0806-2022Class II | Bio-Rad Laboratories, Inc.Due to current concerns related to COVID-19 vaccine interference with the RPR portion of the BioPlex 2200 Syphilis Total & RPR panel, as well as intermittent m… | Feb 8, 2022 |
| Z-0807-2022Class II | Bio-Rad Laboratories, Inc.Due to current concerns related to COVID-19 vaccine interference with the RPR portion of the BioPlex 2200 Syphilis Total & RPR panel, as well as intermittent m… | Feb 8, 2022 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code LIP. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 3 |
| 2018 | 5 |
| 2019 | 9 |
| 2020 | 46 |
| 2021 | 54 |
| 2022 | 44 |
| 2023 | 66 |
| 2024 | 73 |
| 2025 | 112 |
| 2026 (so far) | 88 |
Last 12 months by event type
- Deaths0
- Injuries1
- Malfunctions103
- Other or not stated0
Most reported device problems, last 12 months
- False Negative Result83
- False Positive Result20
- Compatibility Problem1
- High Test Results1
- Incorrect Measurement1
- Low Test Results1
In plain English: what the newest reports describe Written by AI
Multiple reports describe the syphilis test giving nonreactive (negative) results on patient samples, while other tests on the same samples, such as TPPA, colloidal gold, or tests from other manufacturers, were positive. Most of the nonreactive values were just under the 1 S/CO cutoff.
15 reports: 3002809144-2026-00273, 3002809144-2026-00262, 3002809144-2026-00261, 3002809144-2026-00260, 3002809144-2026-00249, 3002809144-2026-00244, 3002809144-2026-00245, 3002809144-2026-00228, 3002809144-2026-00230, 3002809144-2026-00204, 3002809144-2026-00196, 3002809144-2026-00184, 3002809144-2026-00185, 3002809144-2026-00181, 3002809144-2026-00176
Several reports note that a rapid test called TRUST gave negative results for patients whose other syphilis tests were positive, so the methods disagreed. The reports do not say which method was correct.
6 reports: 3002809144-2026-00260, 3002809144-2026-00230, 3002809144-2026-00181, 3002809144-2026-00185, 3002809144-2026-00249, 3002809144-2026-00273
Two reports describe negative results on an automated analyzer that were later questioned. One donor sample read negative at 0.501 COI, but a new sample was positive. In another, a patient sample read negative three times, while another method and a colloidal gold test were positive.
2 reports: 1823260-2026-04304, 1823260-2026-04209
One report describes two reagent lots giving different results on one patient sample. One lot gave 0.88 and 0.80 S/CO (nonreactive), while the other gave 1.23 and 1.16 S/CO (reactive). Quality control was in range for both lots.
Report: 3002809144-2026-00186
One report describes a sample reading 5.89 S/CO (reactive) on the syphilis test, while the same sample tested at another laboratory was expected to be nonreactive. The laboratory did not report the result, and the other lab's method was not provided.
Report: 3007111389-2026-00159
Written by AI from the narratives of the 20 newest reports that have one, received Jul 6, 2026 to Oct 27, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 58in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 34registered establishments that list this product code
Other codes under 21 CFR 866.3830
Product codes classified under the same regulation.
Watch LIP in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get LIP every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LIP. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:LIPRetrieved Nov 28, 2026, 10:22 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:LIPRetrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:LIP · count decision_dateRetrieved Nov 28, 2026, 10:22 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:LIPRetrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LIP · count event_date_initiatedRetrieved Nov 28, 2026, 10:22 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LIP · count res_event_numberRetrieved Nov 28, 2026, 10:22 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-2965-2026" "Z-2966-2026" "Z-1096-2022" "Z-0806-2022" "Z-0807-2022")Retrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LIP AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LIP AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LIP AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 5:05 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:LIPRetrieved Nov 28, 2026, 10:22 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:LIPRetrieved Nov 28, 2026, 10:22 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:866.3830Retrieved Nov 29, 2026, 4:18 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil