FDA product code
LFZ: Enzyme Linked Immunoabsorbent Assay, Cytomegalovirus (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 866.3175
- Review panel
- Microbiology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 580 in the last 12 months
- Median review time
- 197 daysthe latest 25 decisions
- Recall records
- 90 in the last 12 months
- Adverse event reports
- 47last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code LFZ per year, by decision date, 58 since 1982. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 2 |
| 2018 | 1 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 2 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K220949 | Architect CMV IgGAbbott Laboratories | Oct 27, 2022209 days |
| K220911 | Elecsys CMV IgGRoche Diagnostics | Oct 12, 2022197 days |
| K181213 | ADVIA Centaur CMV IgG and ADVIA Centaur CMV IgG Quality ControlSiemens Healthcare Diagnostics, Inc. | Jul 30, 201884 days |
| K163569 | Elecsys CMV IgMRoche Diagnostics | Mar 17, 201788 days |
| K162969 | LIAISON® CMV IgG and LIAISON® CMV IgG Serum Control SetDiaSorin, Inc. | Jan 6, 201774 days |
| K142133 | ELECYS CMV IGM IMMUNOASSAY; ELECYS PRECICONTROL CMV IGMRoche Diagnostics | Oct 24, 201481 days |
| K131605 | ELECSYS PRECICONTROL CMV IGG, ELECSYS CMV IGG IMMUNOASSAYRoche Diagnostics | Feb 28, 2014270 days |
| K040290 | DIASORIN LIAISON CMV IGM/IGGDiasorin S.P.A | Jun 1, 2005481 days |
| K993952 | IMMULITE 2000 CMV IGG, MODEL L2KCV2, L2KCV6Diagnostic Products Corp. | Dec 16, 199924 days |
| K982485 | SERAQUEST CMV IGMQuest Intl., Inc. | Jan 13, 1999180 days |
Recalls
Recall records for product code LFZ per year, by the date the recall began. Each recalled product is one record; these 9 records belong to 9 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 1 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 1 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 1 |
| 2026 (so far) | 0 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-0758-2026Class II | GET TESTED INTERNATIONAL ABDistribution without premarket approval/clearance. | Nov 3, 2025 |
| Z-0342-2023Class II | Bio-Rad Laboratories, Inc.Due to an unusual increase in the positivity rate with human IgM antibodies to cytomegalovirus (CMV) Enzyme Immunoassay (EIA). | Oct 19, 2022 |
| Z-1200-2017Class II | BioMerieux SADuring development of the VIDAS 3 software version 1.2, some anomalies have been identified and observed to be already present in the current software version… | Jan 11, 2017 |
| Z-1689-2016Class II | Diamedix CorporationProduct contained an incorrect substrate. | Mar 26, 2015 |
| Z-1047-2009 | Abbott LaboratoriesDefective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument. | Jan 19, 2009 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code LFZ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 13 |
| 2018 | 9 |
| 2019 | 10 |
| 2020 | 7 |
| 2021 | 23 |
| 2022 | 11 |
| 2023 | 36 |
| 2024 | 34 |
| 2025 | 54 |
| 2026 (so far) | 41 |
Last 12 months by event type
- Deaths0
- Injuries4
- Malfunctions43
- Other or not stated0
Most reported device problems, last 12 months
- False Positive Result23
- False Negative Result12
- Low Test Results8
- Non Reproducible Results7
- High Test Results4
In plain English: what the newest reports describe Written by AI
The device reported CMV IgG antibody results as reactive, which normally signals past infection, in six reports. Repeat testing on the same or a new sample, or testing at another laboratory, gave negative results, and the reports suggest the first reactive results were false.
6 reports: 3008344661-2026-00136, 3008344661-2026-00139, 3008344661-2026-00104, 3008344661-2026-00086, 3008344661-2026-00070, 3008344661-2026-00068
In four reports involving pregnant patients, the device gave non-reactive or immunity-suggesting CMV results, while other tests were positive. Later, fetal or newborn CMV infection was reported, and one report alleges hearing loss. The reports say the device results may not have detected an earlier infection, but the cause was not confirmed.
4 reports: 1823260-2026-03625, 1823260-2026-01898, 1823260-2026-01549, 1823260-2026-03018
In three reports, the device gave negative or equivocal CMV results where another test method gave positive results on the same patient sample. In one report, the device read below its cutoff while a different analyzer gave about 10 AU/mL, which is positive.
3 reports: 1823260-2026-01389, 1823260-2026-03307, 1823260-2026-02884
In two reports, results from the same patient changed between samples or repeat tests. One report had reactive, then non-reactive, results on different samples. The other had indeterminate results on the device while other methods gave positive results.
2 reports: 1823260-2026-01596, 1823260-2026-01346
Some reports refer to attachments or separate reports for full patient data, so the details of these cases are not fully given here.
2 reports: 1823260-2026-03520, 1823260-2026-02737
Written by AI from the narratives of the 19 newest reports that have one, received Apr 8, 2026 to Sep 14, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 54in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 48registered establishments that list this product code
Other codes under 21 CFR 866.3175
Product codes classified under the same regulation.
Watch LFZ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get LFZ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LFZ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:LFZRetrieved Nov 29, 2026, 7:56 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:LFZRetrieved Nov 29, 2026, 7:56 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:LFZ · count decision_dateRetrieved Nov 29, 2026, 7:56 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:LFZRetrieved Nov 29, 2026, 7:56 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LFZ · count event_date_initiatedRetrieved Nov 29, 2026, 7:56 PM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:LFZ · count res_event_numberRetrieved Nov 29, 2026, 7:56 PM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-0758-2026" "Z-0342-2023" "Z-1200-2017" "Z-1689-2016" "Z-1047-2009")Retrieved Nov 29, 2026, 7:56 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LFZ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 7:56 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LFZ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 7:56 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:LFZ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 7:56 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:LFZRetrieved Nov 29, 2026, 7:56 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:LFZRetrieved Nov 29, 2026, 7:56 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:866.3175Retrieved Nov 29, 2026, 7:23 AM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil