Product codes

FDA product code

LEG: Enzyme Immunoassay, Valproic Acid (FDA product code)

Follow LEG: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 862.3645
Review panel
Clinical Toxicology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
350 in the last 12 months
Median review time
72 daysthe latest 25 decisions
Recall records
70 in the last 12 months
Adverse event reports
7last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code LEG per year, by decision date, 35 since 1980. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K090358 ARCHITECT IVALPROIC ACID REAGENTS AND ARCHITECT IVALPROIC ACID CALIBRATORS, MODELS 1P35, 1P35Abbott Laboratories, Inc. Aug 6, 2009175 days
K060690 ONLINE VALPROIC ACIDRoche Diagnostics Corp. Aug 8, 2006146 days
K042807 ADVIA IMS VALPROIC ACID METHODBayer Healthcare, LLC Feb 8, 2005119 days
K042476 VITROS CHEMISTRY PRODUCTS VALP REAGENT; VITRO CHEMISTRY PRODUCTS CALIBRATOR KIT 12; VITROS CHEMISTRY PRODUCTS TDM PERFOSOrtho-Clinical Diagnostics, Inc. Dec 1, 200479 days
K032049 ROCHE ONLINE TDM VALPROIC ACIDRoche Diagnostics Corp. Aug 22, 200351 days
K031870 SYNCHRON SYSTEMS VALPROIC ACID (V-PA) REAGENTBeckman Coulter, Inc. Jul 15, 200328 days
K030405 QMS / MULTIGENT VALPROIC ACID REAGENT, PART #396075Seradyn Apr 28, 200381 days
K011649 RANDOX VALPROIC ACIDRandox Laboratories, Ltd. Jan 15, 2002231 days
K011947 EMIT 2000 VALPROIC ACID ASSAY, MODEL OSR4G229ULSyva Co. Jul 2, 200111 days
K002551 EMIT 2000 VALPROIC ACID ASSAY, EMIT 2000 VALPROIC ACID CALIBRATORS, MODEL 4G019UL, 4G109ULDade Behring, Inc. Oct 11, 200055 days

Recalls

Recall records for product code LEG per year, by the date the recall began. Each recalled product is one record; these 7 records belong to 5 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20181
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-0848-2019Class II Ortho-Clinical DiagnosticsNegative bias results when using VITROS VALP Reagent, GEN 25. Dec 11, 2018
Z-0984-2012 Ortho-Clinical DiagnosticsResults using VITROS Chemistry Products VALP Reagent were negatively biased compared to other manufacturers. OCD confirmed that patient results were also affec… May 10, 2011
Z-0920-2007 Roche Diagnostics Corp.Probe Wash: Reported results may have a positive bias because inadequate analyzer probe wash instructions may result in carry-over from the following assays: C… Apr 30, 2007
Z-0351-2007 Bayer Healthcare, LLC (Diagnostics Division)Bayer received a customer complaint where the Valproic Acid Quality control material recovered lower than target when it was run in a panel of other Therapeuti… Sep 8, 2006
Z-0352-2007 Bayer Healthcare, LLC (Diagnostics Division)Bayer received a customer complaint where the Valproic Acid Quality control material recovered lower than target when it was run in a panel of other Therapeuti… Sep 8, 2006

Adverse event reports (MAUDE)

Medical device reports received per year that name product code LEG. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20178
201824
201913
20209
202112
202219
20231
202414
20255
2026 (so far)7

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions7
  • Other or not stated0

Most reported device problems, last 12 months

  1. Low Test Results5
  2. High Test Results2

In plain English: what the newest reports describe Written by AI

  • Quality control fluids, which are lab materials with known values used to check the test, gave valproic acid results that were higher or lower than expected in five reports. The manufacturer's technical support center handled these calls, and no patient harm was reported.

    5 reports: 1319808-2026-00026, 1319808-2025-00029, 1319808-2025-00023, 1319808-2024-00016, 1319808-2024-00015

  • Five reports describe patient results that differed greatly on repeat testing, often after a data flag or a physician's question. Repeat values were often far from the first result, so the correct value was not always clear.

    5 reports: 1823260-2026-01989, 1823260-2025-04462, 1823260-2025-01388, 1823260-2025-01079, 1823260-2024-02023

  • Two reports describe a falsely low valproic acid result on one patient sample that was processed several times. The low result was not reported to the physician, and a higher result was reported instead. No adverse health effects were known.

    2 reports: 2517506-2024-00305, 2517506-2024-00306

  • Two reports describe higher-than-expected valproic acid results on patient samples. In one, a physician questioned a high result, and a retest on another analyzer gave a lower value that was issued as a corrected report. In the other, seven of eight discordant high results were repeated on the same instrument.

    2 reports: 2432235-2024-00295, 1219913-2024-00354

  • One report describes higher-than-expected valproic acid results on several patient samples run during a comparison of reagent batches. The manufacturer said these results were not sent out of the laboratory and that no erroneous patient results were reported.

    Report: 1319808-2024-00022

Written by AI from the narratives of the 15 newest reports that have one, received May 16, 2024 to May 28, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
23in the FDA’s Global Unique Device Identification Database
Establishment listings
42registered establishments that list this product code

Watch LEG in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get LEG every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in LEG. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.