Product codes

FDA product code

KYJ: Prosthesis, Finger, Constrained, Polymer (FDA product code)

Follow KYJ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 888.3230
Review panel
Orthopedic
Premarket route
510(k)
Implant
Yes
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
151 in the last 12 months
Median review time
158 daysthe latest 15 decisions
Recall records
40 in the last 12 months
Adverse event reports
37last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code KYJ per year, by decision date, 15 since 1978. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20221
20232
20240
20250
2026 (so far)1

Latest decisions

NumberDevice and applicantDecision and review time
K263582 KeriFlex® MCP and PIP Finger Joint ProsthesesKeri Medical SA Oct 29, 202629 days
K233670 Ascension Silicone MCP; Ascension Silicone PIPAscension Orthopedics, Inc. Dec 14, 202329 days
K220142 BRM Digitalis SpacerBrm Extremities Srl Apr 4, 2023441 days
K211385 KeriFlex® MCP and PIP Finger Joint ProsthesesKeri Medical SA Jul 27, 2022449 days
K140453 OSTEOTEC SILICONE FINGER IMPLANTOsteotec, Ltd. Apr 17, 2015417 days
K083107 MODIFICATION TO DEPUY NEUFLEX PIP FINGERDePuy Orthopaedics, Inc. Feb 9, 2009112 days
K082231 ASCENSION SILICONE PIPAscension Orthopedics, Inc. Jan 12, 2009158 days
K022892 ASCENSION SILLICONE MCPAscension Orthopedics, Inc. Nov 25, 200287 days
K013629 FINGER JOINT PROSTHESISAvanta Orthopaedics, Inc. Jan 25, 200281 days
K001922 DEPUY NEUFLEX PIP FINGERDepuy, Inc. Aug 29, 200067 days

Recalls

Recall records for product code KYJ per year, by the date the recall began. Each recalled product is one record; these 4 records belong to 3 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20180
20191
20200
20210
20220
20230
20240
20251
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-2222-2025Class II Osteotec LimitedThere is the potential that the silicone implant may contain foreign material Jun 18, 2025
Z-0691-2020Class II Integra Lifesciences Sales LlcSilicone PIP Implant mislabeled as a size 1 implant on the outer packaging when in fact the package contained a size 0 implant Nov 4, 2019
Z-0570-2020Class II Wright Medical Technology, Inc.The firm discovered a lack of adequate documentation confirming sterility on certain products. Nov 8, 2016
Z-0571-2020Class II Wright Medical Technology, Inc.The firm discovered a lack of adequate documentation confirming sterility on certain products. Nov 8, 2016

Adverse event reports (MAUDE)

Medical device reports received per year that name product code KYJ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
201714
201811
201926
202048
202145
202239
202324
202429
202533
2026 (so far)30

Last 12 months by event type

  • Deaths0
  • Injuries32
  • Malfunctions5
  • Other or not stated0

Most reported device problems, last 12 months

  1. Insufficient Device Problem Information10
  2. Fracture9
  3. Break8
  4. Patient Device Interaction Problem4
  5. Adverse Event Without Identified Device or Use Problem3
  6. Appropriate Device Problem Term/Code Not Available2

In plain English: what the newest reports describe Written by AI

  • In these reports the joint came out of place (dislocated) after surgery, and the patient needed another operation. Surgeons often swapped the neck piece, which fits into the bone, for one of a different size or shape. One report links a dislocation to a neck that the surgeon felt was too tight, and another notes a traumatic fall three days after surgery.

    9 reports: 3009540749-2026-00005, 3009540749-2026-00006, 3009540749-2026-00009, 3009540749-2026-00012, 3009540749-2026-00013, 3009540749-2026-00015, 3009540749-2026-00016, 3009540749-2026-00017, 3009540749-2026-00018

  • The implant broke into pieces, and the patient needed revision surgery to remove it. One report describes breakage about nine weeks after implantation. Another describes a fracture with inflammation of the joint lining caused by silicone wear debris (tiny particles shed from the implant), with no infection found.

    5 reports: 1644408-2026-02155, 1644408-2026-01276, 1644408-2026-01602, 3024706750-2026-00063, 3024706750-2026-00071

  • During surgery, the cup that holds the implant tilted and would not stay stable in any size tried, and the bone it sits in had a large cyst. The surgeon then removed the trapezium bone instead. In another report, the trapezium broke into pieces while the surgeon was checking the cup, and the surgeon performed a suspensionplasty (a tendon-based repair to replace the bone).

    2 reports: 3009540749-2026-00010, 3009540749-2026-00011

  • One report describes an infection that led to revision surgery. The surgeon washed out the joint and then exchanged the neck piece.

    Report: 3009540749-2026-00014

Written by AI from the narratives of the 19 newest reports that have one, received Jul 2, 2026 to Oct 22, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
176in the FDA’s Global Unique Device Identification Database
Establishment listings
35registered establishments that list this product code

Product codes classified under the same regulation.

Watch KYJ in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get KYJ every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in KYJ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.