Product codes

FDA product code

KWL: Prosthesis, Hip, Hemi-, Femoral, Metal (FDA product code)

Follow KWL: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 888.3360
Review panel
Orthopedic
Premarket route
510(k)
Implant
Yes
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
1083 in the last 12 months
Median review time
75 daysthe latest 25 decisions
Recall records
511 in the last 12 months
Adverse event reports
71last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code KWL per year, by decision date, 108 since 1979. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20182
20191
20201
20210
20221
20230
20241
20253
2026 (so far)2

Latest decisions

NumberDevice and applicantDecision and review time
K262290 REMEDY POLY+PLUS SNAP FIT Acetabular CupOsteoremedies, LLC Jul 24, 202618 days
K260536 Prostalac Hip System, SmartSet V+G Dual Antibiotic Bone CementDepuy Ireland UC May 15, 202687 days
K253675 OsteoRemedies Hip Spacer SystemOsteoremedies, LLC Dec 23, 202532 days
K252326 InterSpace GV Hip SpacerTecres S.P.A. Aug 6, 202512 days
K242865 Synicem Hip Spacer; Synicem Knee Spacer; Synicem Shoulder SpacerBiocomposites, Ltd. Jun 12, 2025265 days
K234028 COPAL® exchange G hip spacer; COPAL® G+VHeraeus Medical GmbH Feb 15, 202457 days
K222632 UHR Bipolar Implants, Restoration GAP II ImplantsHowmedica Osteonics Corp., Dba Stryker Orthopaedics Oct 21, 202251 days
K192995 OsteoRemedies Hip Spacer SystemOsteoremedies, LLC Jan 10, 202077 days
K191981 REMEDY Acetabular CupOsteoremedies, LLC Aug 23, 201930 days
K172906 REMEDY PLUS Hip Spacer, UNITE PLUS Bone CementOsteoremedies, LLC Aug 29, 2018341 days

Recalls

Recall records for product code KWL per year, by the date the recall began. Each recalled product is one record; these 51 records belong to 9 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20172
20181
20190
20200
20210
20220
20230
20243
20250
2026 (so far)1

Latest recalls

Recall and classFirm and reasonBegan
Z-3199-2026Class II Howmedica Osteonics Corp.Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential… Jul 30, 2026
Z-2370-2024Class II Howmedica Osteonics Corp.Mislabeled: the size on the package label of the UNITRAX Endoprosthesis Head Component potentially may not match the device within the packaging. This product… Jun 19, 2024
Z-2371-2024Class II Howmedica Osteonics Corp.Mislabeled: the size on the package label of the UNITRAX Endoprosthesis Head Component potentially may not match the device within the packaging. This product… Jun 19, 2024
Z-2372-2024Class II Howmedica Osteonics Corp.Mislabeled: the size on the package label of the UNITRAX Endoprosthesis Head Component potentially may not match the device within the packaging. This product… Jun 19, 2024
Z-2521-2018Class II DePuy Orthopaedics, Inc.Two lots of femoral stems were labeled as Size 11. However, they are actually Size 12. Potential adverse effects include poor joint mechanics, improper fit, lo… Jun 21, 2018

Adverse event reports (MAUDE)

Medical device reports received per year that name product code KWL. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017347
2018294
2019415
2020270
2021107
202299
202348
2024143
202529
2026 (so far)70

Last 12 months by event type

  • Deaths0
  • Injuries63
  • Malfunctions7
  • Other or not stated1

Most reported device problems, last 12 months

  1. Adverse Event Without Identified Device or Use Problem32
  2. Osseointegration Problem12
  3. Device Dislodged or Dislocated8
  4. Material Deformation4
  5. Detachment of Device or Device Component3
  6. Biocompatibility2

In plain English: what the newest reports describe Written by AI

  • Five reports describe thigh bone fractures: three around the implant, which led to revision surgery, and two after the patient fell. One of the fall reports says the device was removed and put in reclaim.

    5 reports: 1818910-2026-16796, 1818910-2026-17398, 1818910-2026-17036, 1818910-2026-18626, 1818910-2026-17588

  • Four reports describe infection around the hip implant, and three of them led to revision surgery in which parts were replaced. One report says a scan of the right thigh showed fluid collections, with abscesses (pus-filled pockets) suspected.

    4 reports: 0002249697-2026-00964, 1818910-2026-17895, 1818910-2026-17740, 1818910-2026-17815

  • In three reports, the inner head or locking mechanism separated from the cup when doctors tried to put a dislocated hip back in place. One surgeon asked the manufacturer whether this can happen by design or points to a problem with the part.

    3 reports: 0002249697-2026-00987, 0002249697-2026-00972, 0002249697-2026-00961

  • Three reports describe the stem coming loose from the bone, in some cases causing pain, and needing to be replaced. One of these also notes a femoral (thigh bone) fracture during the replacement.

    3 reports: 1818910-2026-17955, 1818910-2026-17012, 1818910-2026-17036

  • Two reports say the stem sat too high in the bone. One gives the stem as sitting 1 mm higher than the broach, a sizing tool used to prepare the bone, while the other gives no further detail.

    2 reports: 1818910-2026-18111, 1818910-2026-17376

Written by AI from the narratives of the 19 newest reports that have one, received Sep 29, 2026 to Oct 29, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
6,649in the FDA’s Global Unique Device Identification Database
Establishment listings
579registered establishments that list this product code

Watch KWL in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get KWL every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in KWL. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.