FDA product code
KSZ: System, Test, Automated Blood Grouping And Antibody (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 864.9175
- Review panel
- Hematology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- No
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Eligible
- 510(k) and De Novo decisions
- 40 in the last 12 months
- Median review time
- 46 daysthe latest 4 decisions
- Recall records
- 00 in the last 12 months
- Adverse event reports
- 39last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code KSZ per year, by decision date, 4 since 1977. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 0 |
| 2018 | 0 |
| 2019 | 0 |
| 2020 | 0 |
| 2021 | 0 |
| 2022 | 0 |
| 2023 | 0 |
| 2024 | 0 |
| 2025 | 0 |
| 2026 (so far) | 0 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K810899 | TECHNICON AUTOGROUPER 16-C SYSTEMTechnicon Instruments Corp. | May 5, 198132 days |
| K790282 | ELDONCARDNordisk Infuser System | May 4, 197980 days |
| K771098 | BLOOD GROUPING SYS., BG-9Technicon Instruments Corp. | Aug 4, 197745 days |
| K770185 | RH CONTROL DILUENTBd Becton Dickinson Vacutainer Systems Preanalytic | Mar 14, 197746 days |
Recalls
Recall records for product code KSZ per year, by the date the recall began. Each recalled product is one record.
No recalls on record for this product code.
Adverse event reports (MAUDE)
Medical device reports received per year that name product code KSZ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 626 |
| 2018 | 686 |
| 2019 | 298 |
| 2020 | 75 |
| 2021 | 76 |
| 2022 | 78 |
| 2023 | 151 |
| 2024 | 66 |
| 2025 | 66 |
| 2026 (so far) | 35 |
Last 12 months by event type
- Deaths0
- Injuries0
- Malfunctions39
- Other or not stated0
Most reported device problems, last 12 months
- False Negative Result14
- False Positive Result9
- Incorrect, Inadequate or Imprecise Result or Readings9
- Imprecise or Erratic Results3
- Inaccurate Information2
- Low Readings2
In plain English: what the newest reports describe Written by AI
Several reports describe the analyzer's pipetting needle possibly carrying liquid between wells or samples, which can cause a false positive result. In one case a test on an anti-B well read positive and then negative on repeat, and the manufacturer said the cause could not be confirmed because card images were unavailable. Another report links a suspected carryover to a strongly reacting previous sample.
3 reports: 9610824-2026-00021, 9610824-2026-00012, 9610824-2026-00016
Some reports describe a false negative or weak result caused by too few red blood cells in the test. In one case a thin cell layer (pellet) did not reach the level that triggers the low pellet alarm, and the manufacturer attributed the result to a poorly mixed cell suspension prepared by the lab.
3 reports: 9610824-2026-00017, 2250051-2026-00014, 3002772505-2026-00002
A validation study found six samples that were positive on an older instrument but negative on the new analyzer, and the results were auto-verified without review. The investigation was still ongoing when reported.
2 reports: 9610824-2026-00006, 9610824-2026-00007
Several reports describe the analyzer giving a discrepant result that a lab detected before any transfusion. In one case a clot in a tube caused a false result, and the manufacturer said the test card's results should be checked against a previously typed sample.
2 reports: 3002772505-2026-00003, 2250051-2026-00026
A report describes a false positive result on a patient sample, which the manufacturer attributed to carryover after the needle accessed another patient's tube. The patient was receiving anti-CD38 therapy, which can interfere with antibody screening tests.
Report: 9610824-2026-00016
Written by AI from the narratives of the 19 newest reports that have one, received Feb 17, 2026 to Aug 5, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 102in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 92registered establishments that list this product code
Other codes under 21 CFR 864.9175
Product codes classified under the same regulation.
Watch KSZ in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get KSZ every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in KSZ. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:KSZRetrieved Nov 29, 2026, 11:13 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:KSZRetrieved Nov 29, 2026, 8:12 PM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:KSZ · count decision_dateRetrieved Nov 29, 2026, 11:13 PM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:KSZRetrieved Nov 29, 2026, 8:12 PM UTC.
- device/recall · search product_code:KSZ · count event_date_initiatedRetrieved Nov 29, 2026, 11:13 PM UTC.
- device/recall · search product_code:KSZ · count res_event_numberRetrieved Nov 29, 2026, 11:13 PM UTC.
- device/event · search device.device_report_product_code:KSZ AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 8:12 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:KSZ AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 8:12 PM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:KSZ AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 8:12 PM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:KSZRetrieved Nov 29, 2026, 11:13 PM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:KSZRetrieved Nov 29, 2026, 11:13 PM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:864.9175Retrieved Nov 29, 2026, 11:13 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.