Product codes

FDA product code

KSZ: System, Test, Automated Blood Grouping And Antibody (FDA product code)

Follow KSZ: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 864.9175
Review panel
Hematology
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Not eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
40 in the last 12 months
Median review time
46 daysthe latest 4 decisions
Recall records
00 in the last 12 months
Adverse event reports
39last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code KSZ per year, by decision date, 4 since 1977. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20170
20180
20190
20200
20210
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K810899 TECHNICON AUTOGROUPER 16-C SYSTEMTechnicon Instruments Corp. May 5, 198132 days
K790282 ELDONCARDNordisk Infuser System May 4, 197980 days
K771098 BLOOD GROUPING SYS., BG-9Technicon Instruments Corp. Aug 4, 197745 days
K770185 RH CONTROL DILUENTBd Becton Dickinson Vacutainer Systems Preanalytic Mar 14, 197746 days

Recalls

Recall records for product code KSZ per year, by the date the recall began. Each recalled product is one record.

All of them in the FDA’s recall database

No recalls on record for this product code.

Adverse event reports (MAUDE)

Medical device reports received per year that name product code KSZ. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
2017626
2018686
2019298
202075
202176
202278
2023151
202466
202566
2026 (so far)35

Last 12 months by event type

  • Deaths0
  • Injuries0
  • Malfunctions39
  • Other or not stated0

Most reported device problems, last 12 months

  1. False Negative Result14
  2. False Positive Result9
  3. Incorrect, Inadequate or Imprecise Result or Readings9
  4. Imprecise or Erratic Results3
  5. Inaccurate Information2
  6. Low Readings2

In plain English: what the newest reports describe Written by AI

  • Several reports describe the analyzer's pipetting needle possibly carrying liquid between wells or samples, which can cause a false positive result. In one case a test on an anti-B well read positive and then negative on repeat, and the manufacturer said the cause could not be confirmed because card images were unavailable. Another report links a suspected carryover to a strongly reacting previous sample.

    3 reports: 9610824-2026-00021, 9610824-2026-00012, 9610824-2026-00016

  • Some reports describe a false negative or weak result caused by too few red blood cells in the test. In one case a thin cell layer (pellet) did not reach the level that triggers the low pellet alarm, and the manufacturer attributed the result to a poorly mixed cell suspension prepared by the lab.

    3 reports: 9610824-2026-00017, 2250051-2026-00014, 3002772505-2026-00002

  • A validation study found six samples that were positive on an older instrument but negative on the new analyzer, and the results were auto-verified without review. The investigation was still ongoing when reported.

    2 reports: 9610824-2026-00006, 9610824-2026-00007

  • Several reports describe the analyzer giving a discrepant result that a lab detected before any transfusion. In one case a clot in a tube caused a false result, and the manufacturer said the test card's results should be checked against a previously typed sample.

    2 reports: 3002772505-2026-00003, 2250051-2026-00026

  • A report describes a false positive result on a patient sample, which the manufacturer attributed to carryover after the needle accessed another patient's tube. The patient was receiving anti-CD38 therapy, which can interfere with antibody screening tests.

    Report: 9610824-2026-00016

Written by AI from the narratives of the 19 newest reports that have one, received Feb 17, 2026 to Aug 5, 2026. Reports aren’t validated, not every event is reported, and a summary can be wrong: read the reports before you rely on it.

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
102in the FDA’s Global Unique Device Identification Database
Establishment listings
92registered establishments that list this product code

Product codes classified under the same regulation.

Watch KSZ in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get KSZ every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in KSZ. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.