FDA product code
KPO: Dialysate Concentrate For Hemodialysis (Liquid Or Powder) (FDA product code)
Classification on the FDA’s site Total product life cycle report
- Device class
- Class II
- Regulation
- 21 CFR 876.5820
- Review panel
- Gastroenterology and Urology
- Premarket route
- 510(k)
- Implant
- No
- Life-sustaining or supporting
- Yes
- GMP exempt
- No
- Third-party review
- Not eligible
- Summary malfunction reporting
- Ineligible
- 510(k) and De Novo decisions
- 1242 in the last 12 months
- Median review time
- 231 daysthe latest 25 decisions
- Recall records
- 1139 in the last 12 months
- Adverse event reports
- 123last 12 months, 0 deaths
510(k) and De Novo decisions
Decisions for product code KPO per year, by decision date, 124 since 1978. Review time is the days from receipt to decision.
Show the numbers
| Year | Decisions |
|---|---|
| 2017 | 3 |
| 2018 | 2 |
| 2019 | 0 |
| 2020 | 5 |
| 2021 | 1 |
| 2022 | 2 |
| 2023 | 3 |
| 2024 | 3 |
| 2025 | 2 |
| 2026 (so far) | 2 |
Latest decisions
| Number | Device and applicant | Decision and review time |
|---|---|---|
| K260799 | Nipro Dry Acid Concentrate MixerNipro Technical Solutions, Incoporated | Aug 19, 2026161 days |
| K253462 | NaturaLyte® Dry Bicarbonate Concentrate (08-4112-2)Fresenius Medical Care Renal Therapies Group, LLC | Jan 29, 2026114 days |
| K252180 | Bicarby Dialysate RFP-404 (RFP-404-W); Bicarby Dialysate RFP-403 (RFP-403-W, RFP-403-G); Bicarby Dialysate RFP-453 (RFP-453-W, RFP-453-G); Bicarby Dialysate RFP-454 (RFP-454-W, RFP-454-G); Bicarby Dialysate RFP-456 (RFP-456-W)Fresenius Medical Care Renal Therapies Group, LLC | Sep 9, 202560 days |
| K243786 | Bicarby Dialysate RFP-402 (RFP-402-G); Bicarby Dialysate RFP-400 (RFP-400-G); Bicarby Dialysate RFP-407 (RFP-407-G); Bicarby Dialysate RFP-401 (RFP-401-G); Bicarby Dialysate RFP-404 (RFP-404-G); Bicarby Dialysate RFP-456 (RFP-456-G); Ci-Ca Dialysate 2K (RFP-457-G); Ci-Ca Dialysate 4K (RFP-458-G)Fresenius Medical Care Renal Therapies Group, LLC | Apr 4, 2025116 days |
| K240920 | HemoCare Bicarbonate Concentrate Set (BCS)Deka Research and Development | Jul 12, 202499 days |
| K233950 | pureFLOW 402 (F00012067); pureFLOW 406 (F00012068); pureFLOW 401 (F00012069); pureFLOW 400 (F00012070)Fresenius Medical Care Renal Therapies Group, LLC | May 10, 2024148 days |
| K233213 | NxStage PureFlow SolutionNxstage Medical, Inc. | Jan 18, 2024112 days |
| K233159 | pureFLOW 402, pureFLOW 400, pureFLOW 406, pureFLOW 401, pureFLOW 502, pureFLOW 504Fresenius Medical Care Renal Therapies Group, LLC | Dec 1, 202365 days |
| K223431 | Citric Complete Liquid Citric Acid ConcentrateNipro Renal Solutions USA, Corp. | Aug 4, 2023263 days |
| K221652 | NIKKICARTRenal Care Dialysis Solutions, S.A. DE C.V. | Mar 3, 2023269 days |
Recalls
Recall records for product code KPO per year, by the date the recall began. Each recalled product is one record; these 113 records belong to 45 recall events.
Show the numbers
| Year | Recall records |
|---|---|
| 2017 | 2 |
| 2018 | 13 |
| 2019 | 3 |
| 2020 | 32 |
| 2021 | 4 |
| 2022 | 6 |
| 2023 | 0 |
| 2024 | 2 |
| 2025 | 8 |
| 2026 (so far) | 9 |
Latest recalls
| Recall and class | Firm and reason | Began |
|---|---|---|
| Z-3133-2026Class II | Nipro Renal Soultions USA, CorporationHemodialysis concentrates contain high endotoxin levels | Aug 5, 2026 |
| Z-3134-2026Class II | Nipro Renal Soultions USA, CorporationHemodialysis concentrates contain high endotoxin levels | Aug 5, 2026 |
| Z-3135-2026Class II | Nipro Renal Soultions USA, CorporationHemodialysis concentrates contain high endotoxin levels | Aug 5, 2026 |
| Z-3136-2026Class II | Nipro Renal Soultions USA, CorporationHemodialysis concentrates contain high endotoxin levels | Aug 5, 2026 |
| Z-3091-2026Class II | Fresenius Medical Care Renal Therapies Group, LLCComplaints of blue-green discoloration in product | Jul 14, 2026 |
Adverse event reports (MAUDE)
Medical device reports received per year that name product code KPO. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.
Show the numbers
| Year | Reports |
|---|---|
| 2017 | 80 |
| 2018 | 97 |
| 2019 | 79 |
| 2020 | 244 |
| 2021 | 63 |
| 2022 | 25 |
| 2023 | 59 |
| 2024 | 259 |
| 2025 | 98 |
| 2026 (so far) | 119 |
Last 12 months by event type
- Deaths0
- Injuries2
- Malfunctions119
- Other or not stated2
Most reported device problems, last 12 months
- Fluid/Blood Leak78
- Material Separation48
- Defective Component29
- Detachment of Device or Device Component29
- Thermal Decomposition of Device12
- Break6
On the market
How widely the product code is used in FDA’s device identifier and establishment listing databases.
- Device identifiers (UDI)
- 678in the FDA’s Global Unique Device Identification Database
- Establishment listings
- 101registered establishments that list this product code
Other codes under 21 CFR 876.5820
Product codes classified under the same regulation.
Watch KPO in DeviceVigil
DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.
Get KPO every Monday
One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in KPO. Free. Add up to two more codes, separated by commas.
Sources
Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.
- device/classification · search product_code:KPORetrieved Nov 29, 2026, 8:05 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:KPORetrieved Nov 29, 2026, 8:05 AM UTC. FDA data updated Nov 16, 2026.
- device/510k · search product_code:KPO · count decision_dateRetrieved Nov 29, 2026, 8:05 AM UTC. FDA data updated Nov 16, 2026.
- device/recall · search product_code:KPORetrieved Nov 29, 2026, 8:05 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:KPO · count event_date_initiatedRetrieved Nov 29, 2026, 8:05 AM UTC. FDA data updated Nov 26, 2026.
- device/recall · search product_code:KPO · count res_event_numberRetrieved Nov 29, 2026, 8:05 AM UTC. FDA data updated Nov 26, 2026.
- device/enforcement · search recall_number:("Z-3133-2026" "Z-3134-2026" "Z-3135-2026" "Z-3136-2026" "Z-3091-2026")Retrieved Nov 29, 2026, 8:05 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:KPO AND date_received:[20251129 TO 20261129] · count event_type.exactRetrieved Nov 29, 2026, 8:05 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:KPO AND date_received:[20251129 TO 20261129] · count product_problems.exactRetrieved Nov 29, 2026, 8:05 AM UTC. FDA data updated Nov 18, 2026.
- device/event · search device.device_report_product_code:KPO AND date_received:[20170101 TO 20261129] · count date_receivedRetrieved Nov 29, 2026, 8:05 AM UTC. FDA data updated Nov 18, 2026.
- device/udi · search product_codes.code:KPORetrieved Nov 29, 2026, 8:05 AM UTC. FDA data updated Nov 2, 2026.
- device/registrationlisting · search products.product_code:KPORetrieved Nov 29, 2026, 8:05 AM UTC. FDA data updated Nov 16, 2026.
- device/classification · search regulation_number:876.5820Retrieved Nov 28, 2026, 5:18 PM UTC. FDA data updated Nov 16, 2026.
DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.
DeviceVigil