Product codes

FDA product code

KMW: Handpiece, Rotary Bone Cutting (FDA product code)

Follow KMW: new clearances and recalls, emailed on Mondays

Free. We email a confirmation link first; every email has a one-click unsubscribe. Privacy

Device class
Class II
Regulation
21 CFR 872.4120
Review panel
Dental
Premarket route
510(k)
Implant
No
Life-sustaining or supporting
No
GMP exempt
No
Third-party review
Eligible
Summary malfunction reporting
Eligible
510(k) and De Novo decisions
90 in the last 12 months
Median review time
87 daysthe latest 9 decisions
Recall records
100 in the last 12 months
Adverse event reports
27last 12 months, 0 deaths

510(k) and De Novo decisions

Decisions for product code KMW per year, by decision date, 9 since 1986. Review time is the days from receipt to decision.

All of them in the FDA’s 510(k) database

510(k) and De Novo decisions per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearDecisions
20172
20180
20190
20200
20211
20220
20230
20240
20250
2026 (so far)0

Latest decisions

NumberDevice and applicantDecision and review time
K211584 Oral Surgery ContraNakanishi, Inc. Aug 19, 202187 days
K171155 Surgical Angle Handpiece, Surgical Straight HandpieceNakanishi, Inc. Nov 21, 2017215 days
K162544 OsteoMed PINNACLE DriverOsteomed Apr 4, 2017204 days
K080939 SURGICAL HANDPIECES, MIKRO SAW HANDPIECESW&H Dentalwerk Buermoos GmbH Jul 29, 2008117 days
K994298 FRIOS MICROSAWFriadent GmbH Feb 18, 200059 days
K971692 OSTEOPOWER SYSTEM AND ACCESSORIESOsteomed Corp. Jun 27, 199751 days
K970953 E-TYPE SPEED REDUCER CONTRA ANGLENsk Nakanishi, Inc. Apr 15, 199729 days
K932255 YOUNG D'GRANULATORYoung Dental Manufacturing Co. 1, LLC Apr 15, 1994340 days
K862110 OSSEODENT DRILLING SYSTEMOtc/America Aug 7, 198665 days

Recalls

Recall records for product code KMW per year, by the date the recall began. Each recalled product is one record; these 10 records belong to 3 recall events.

All of them in the FDA’s recall database

Recall records per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearRecall records
20170
20188
20190
20200
20211
20220
20230
20241
20250
2026 (so far)0

Latest recalls

Recall and classFirm and reasonBegan
Z-1168-2024Class II W&H DENTALWERK BUERMOOS GMBH1:2 speed increasing surgical handpieces have been incorrectly laser marked during production. The specific laser marking should read 1:2, but instead it reads… Jan 16, 2024
Z-0335-2022Class II Osteomed, LLCDue to consoles not in compliance with the latest electrical safety standards IEC 60601 and as a result shocked a patient. Oct 28, 2021
Z-0219-2019Class II Nakanishi Inc.Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave ster… Jun 1, 2018
Z-0220-2019Class II Nakanishi Inc.Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave ster… Jun 1, 2018
Z-0221-2019Class II Nakanishi Inc.Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave ster… Jun 1, 2018

Adverse event reports (MAUDE)

Medical device reports received per year that name product code KMW. Reports aren’t validated, not every event is reported, and a count of reports isn’t an event rate.

Where did MAUDE go?

Adverse event reports received per year, 2017 to 2026. 2026 is the year so far.
Show the numbers
YearReports
20174
201816
201913
202013
202132
202254
202325
202449
202524
2026 (so far)27

Last 12 months by event type

  • Deaths0
  • Injuries27
  • Malfunctions0
  • Other or not stated0

Most reported device problems, last 12 months

  1. Overheating of Device25
  2. Temperature Problem2

On the market

How widely the product code is used in FDA’s device identifier and establishment listing databases.

Device identifiers (UDI)
446in the FDA’s Global Unique Device Identification Database
Establishment listings
28registered establishments that list this product code

Watch KMW in DeviceVigil

DeviceVigil is a Chrome and Edge extension that shows this page’s numbers beside the FDA’s own 510(k), classification and recall pages, lists the individual adverse event reports, and tells you when something changes for the product codes you watch.

Add to Chrome — free See how it works

Get KMW every Monday

One short email after openFDA’s weekly update with that week’s 510(k), De Novo and PMA decisions and recalls in KMW. Free. Add up to two more codes, separated by commas.

Up to three codes. We email a confirmation link first; every brief has a one-click unsubscribe. Privacy

Sources

Every figure on this page comes from openFDA, the FDA’s public data service, which notes that its data is not validated. Check the FDA record before you rely on a figure.

DeviceVigil is an independent tool by Thump LLC and is not affiliated with or endorsed by the FDA. This page isn’t medical, legal or regulatory advice.